Proton Calorimetry/Meetings/2022/06/15
Minutes for UCL Proton Calorimetry Meetings, 15th June
Present
Saad Shaikh, Simon Jolly
Saad Shaikh
- Tested crosstalk with mylar sheets in black scintillator stack enclosure – results significantly better than white sheets. Added results to crosstalk experiment page.
- Prepared Rev. B DDC232 prototype detector with 23 white painted and 25 clear scintillator sheets, latter using mylar sheets for optical isolation.
- Is necessary to take ground pin from DDC232 and connect to metal enclosure to minimise electronic noise pickup.
- Prepared Rev. C DDC232 prototype detector with 32 clear sheets, using mylar sheets for optical isolation.
- NUVIA sent 3.4 mm sheets instead of 3 mm sheets, therefore had to use 11 2 mm sheets and 21 3.4 mm sheets to make up the difference.
- Added option to DDC232 analysis code to reverse order of photodiodes.
- Will conduct beam test at UCLH tomorrow afternoon.