Proton Calorimetry/Experimental Runs/2018/Mar8-9
Beam tests of the UCL plastic scintillator range calorimeter.
Thursday 8th of March 2018
Pre-data Taking Information and Notes
- The tests were carried out with an energy of 36 MeV. A Bragg Peak measurement was carried out to confirm the exact beam energy.
- Collimators used: 2 mm and 20 mm diameter.
- PMMA sheets used: 3 mm, taped over the mylar window.
- Data was acquired with the DSLR camera and the ISDI CMOS sensor:
- The average background value per pixel for the sensor is ~ 2500. The sensor pixel dimensions are 1030 x 1536. The sensor frame rate was always set to 25 frames per second.
- The DSLR ISO was always set to 200.
- 6 x 3 mm scintillator sheets were used. The scintillator stack was set up in the following order, starting with the first sheet facing the beam:
- 31, 32, 33, 34, 35, 48
- The scintillator stack was coupled to the sensor using optical gel.
- The setup can be seen in the following pictures:
- The centre of the "bare" scintillator (where the vice is not covering any of the scintillator face) was marked on the Mylar window and lined up with the beam using a film:
- The dose received (as shown on the monitor in the control room) during the full day of measurements was:
- With a 2 mm collimator: 746.9 uC
- With a 20 mm collimator: 417.7 uC
- With a 2 mm collimator + 3 mm of PMMA: 251.5 uC
- The current was variable/unstable throughout the day.
- To calculate the actual dose delivered to the scintillator:
- / 160 -> to convert to proton current
- / (50/ collimator diameter)^2 -> to account for reduction in area due to the collimator
- The directory containing the sensor data is:
/unix/pbt/data/birmingham/mar_2018/EPIX/XCAP/data/20180308_bham
- The directory containing the DSLR data is:
/unix/pbt/data/birmingham/mar_2018/Nikon/20180308_bham
Data Taking
| Run number | DSLR Aperture | DSLR Exposure | Sensor Full Well Mode | Sensor Sequence | Beam Energy | Current on Monitor | Collimator Diameter | Comments | DSLR Thumbnail | Sensor Thumbnail |
|---|---|---|---|---|---|---|---|---|---|---|
| 1 | f/10 | 2 s | N/A | N/A | N/A | N/A | N/A | Beam off, lid open. Focus check for DSLR. Low light. | N/A | |
| 2 | f/10 | 5 s | N/A | N/A | N/A | N/A | N/A | Beam off, lid open. Focus check for DSLR. Low light. | N/A | |
| 3 | f/10 | 20 s | N/A | N/A | N/A | N/A | N/A | Beam off, lid open. Focus check for DSLR. Low light. | N/A | |
| 4 | f/10 | 20 s | N/A | N/A | N/A | N/A | N/A | Beam off, lid open. Focus check for DSLR. Focus - okay. | N/A | |
| 5 | f/5 | 5 s | N/A | N/A | N/A | N/A | N/A | Beam off, lid open. Focus check and background picture with white paper for DSLR. | N/A | |
| 6 | f/10 | 5 s | Low | 42 frames | N/A | N/A | N/A | Beam off, lid closed, lights off. Background run. | N/A | |
| 7 | f/3.2 | 2 s | Low | 21 frames | N/A | N/A | N/A | Beam off, lid closed, lights off. Background run. | N/A | |
| 8 | f/3.2 | 10 s | Low | 21 frames | 36 MeV | 1 uA | 2 mm | Beam on. | N/A | |
| 9 | N/A | N/A | Low | 21 frames | 36 MeV | 1 uA | 2 mm | Beam on. | N/A | N/A |
| 10 | f/3.2 | 10 s | Low | 1 frame | 36 MeV | 0.8 uA | 2 mm | Beam on. | ||
| 11 | f/3.2 | 10 s | Low | 21 frames | 36 MeV | 0.8 uA | 2 mm | Beam on. | N/A | |
| 12 | f/3.2 | 10 s | Low | 21 frames | 36 MeV | 10 nA | 20 mm | Beam on. | ||
| 13 | f/3.2 | 10 s | Low | 21 frames | 36 MeV | 20 nA | 20 mm | Beam on. At this current sensor is saturated on low full well mode. Unable to switch sensor into high full well mode so no sensor data taken until a return to low currents. | ||
| 14 | f/3.5 | 10 s | N/A | N/A | 36 MeV | 50 nA | 20 mm | Beam on. | N/A | |
| 15 | f/3.5 | 5 s | N/A | N/A | 36 MeV | 60 nA | 20 mm | Beam on. | N/A | |
| 16 | f/3.5 | 2 s | N/A | N/A | 36 MeV | 100 nA | 20 mm | Beam on. | N/A | |
| 17 | f/3.5 | 1 s | N/A | N/A | 36 MeV | 100 nA | 20 mm | Beam not caught on image. | N/A | |
| 18 | f/3.5 | 1 s | N/A | N/A | 36 MeV | 100 nA | 20 mm | Some beam caught on image. | N/A | |
| 19 | f/3.5 | 1 s | N/A | N/A | 36 MeV | 100 nA | 20 mm | Beam on. | N/A | |
| 20 | f/4 | 1 s | N/A | N/A | 36 MeV | 200 nA | 20 mm | Variable beam current. | N/A | |
| 21 | f/4 | 1 s | N/A | N/A | 36 MeV | 260 nA | 20 mm | Beam on. | N/A | |
| 22 | f/4 | 1 s | N/A | N/A | 36 MeV | 260 nA | 20 mm | Beam on. | N/A | |
| 23 | f/4.5 | 1 s | N/A | N/A | 36 MeV | 500 nA | 20 mm | Beam on. | N/A | |
| 24 | f/4.5 | 1/2 s | N/A | N/A | 36 MeV | 500 nA | 20 mm | Beam on. | N/A | |
| 25 | f/3.2 | 1/2 s | N/A | N/A | 36 MeV | 800 nA | 20 mm | Beam on. Saturation test for DSLR. | N/A | |
| 26 | f/6.3 | 1/2 s | N/A | N/A | 36 MeV | 800 nA | 20 mm | Some beam caught on image. | N/A | |
| 27 | f/6.3 | 1/2 s | N/A | N/A | 36 MeV | 850 nA | 20 mm | Beam on. | N/A | |
| 28 | f/7.1 | 1/2 s | N/A | N/A | 36 MeV | 1 uA | 20 mm | Beam not caught on image. | N/A | |
| 29 | f/7.1 | 1/2 s | N/A | N/A | 36 MeV | 1.1 uA | 20 mm | Beam on. | N/A | |
| 30 | f/7.1 | 1/2 s | N/A | N/A | 36 MeV | 1.1 uA | 20 mm | Some beam caught on image. | N/A | |
| 31 | f/7.1 | 1/2 s | N/A | N/A | 36 MeV | 950 nA | 20 mm | Beam on. | N/A | |
| 32 | f/8 | 1/2 s | N/A | N/A | 36 MeV | 1.1 uA | 20 mm | Beam on. | N/A | |
| 33 | f/10 | 1/2 s | N/A | N/A | 36 MeV | 1.2 uA | 20 mm | Beam on. | N/A | |
| 34 | f/3.2 | 10 s | N/A | N/A | 36 MeV | 1.1 uA | 2 mm | Radiation damage assessment. | N/A | |
| 35 | f/3.2 | 10 s | Low | 21 frames | 36 MeV | 1.1 uA | 2 mm | Beam on. | ||
| 36 | N/A | N/A | Low | 21 frames | 36 MeV - 3 mm PMMA | 1.1 uA | 2 mm | Beam not caught on DSLR image, okay for sensor. 3 mm of PMMA taped over Mylar window. |
N/A |
|
| 37 | f/3.2 | 10 s | Low | 21 frames | 36 MeV - 3 mm PMMA | 1 uA | 2 mm | Beam on. 3 mm of PMMA taped over Mylar window. |
Friday 9th of March 2018
Pre-data Taking Information and Notes
- The tests were carried out with an energy of 36 MeV and 28 MeV. A Bragg Peak measurement was carried out to confirm the exact beam energy.
- Collimators used: 20 mm diameter.
- PMMA sheets used: 3 mm, 2 mm and 1 mm, taped over the mylar window.
- Data was acquired with the DSLR camera and the ISDI CMOS sensor:
- The average background value per pixel for the sensor is ~ 2500. The sensor pixel dimensions are 1030 x 1536. The sensor frame rate was always set to 25 frames per second.
- The DSLR ISO was always set to 200.
- The light level in order to not saturate the sensor was reduced by placing green and blue see-through sheets in between the sensor and the scintillator stack for some of the tests.
- 8 x 32 mm scintillator sheets were used. The scintillator stack was set up in the following order, starting with the first sheet facing the beam:
- 7, 10, 13, 14, 15, 17, 18, 19
- Optical gel was not used for coupling, the sensor was placed directly on top of the scintillator stack.
- The dose received (as shown on the monitor in the control room) during the full day of measurements was:
- For the 36 MeV beam
- With a 20 mm collimator: 435.5 uC
- With a 20 mm collimator + PMMA sheets: 17.3 uC
- With a 20 mm collimator + films inserted between scintillator stack and sensor: 536.2 uC
- For the 28 MeV beam
- With a 20 mm collimator: 47.7 uC
- For the 36 MeV beam
- There current was variable/unstable throughout the day.
- To calculate the actual dose delivered to the scintillator:
- / 160 -> to convert to proton current
- / (50/ collimator diameter)^2 -> to account for reduction in area due to the collimator
- The directory containing the sensor data is:
/unix/pbt/data/birmingham/mar_2018/EPIX/XCAP/data/20180309_bham
- The directory containing the DSLR data is:
/unix/pbt/data/birmingham/mar_2018/Nikon/20180309_bham
Data Taking
| Run number | DSLR Aperture | DSLR Exposure | Sensor Full Well Mode | Sensor Sequence | Beam Energy | Current on Monitor | Collimator Diameter | Comments | DSLR Thumbnail | Sensor Thumbnail |
|---|---|---|---|---|---|---|---|---|---|---|
| 38 | f/3.2 | 2 s | N/A | N/A | N/A | N/A | N/A | Beam off, lid open. Focus check and background picture with white paper for DSLR. | N/A | |
| 39 | f/3.2 | 2 s | Low | 21 frames | N/A | N/A | N/A | Beam off, lid closed, lights off. Background run. | ||
| 40 | f/3.2 | 2 s | Low | 21 frames | 36 MeV | 10 nA | 20 mm | Beam on. | ||
| 41 | f/3.2 | 5 s | N/A | N/A | 36 MeV | 10 nA | 20 mm | Beam on. | N/A | |
| 42 | f/3.2 | 5 s | N/A | N/A | 36 MeV | 10 nA | 20 mm | Beam on. Sensor realigned after this run to be straighter with respect to scintillator stack. | N/A | |
| 43 | f/3.2 | 5 s | Low | 21 frames | 36 MeV | 10 nA | 20 mm | Beam on. Sensor realigned after this run to be straighter with respect to scintillator stack. | ||
| 44 | f/3.2 | 5 s | Low | 21 frames | 36 MeV | 10 nA | 20 mm | Beam on. Sensor realigned after this run to be straighter with respect to scintillator stack. | ||
| 45 | f/3.2 | 10 s | N/A | N/A | 36 MeV | 5 - 10 nA | 20 mm | Beam on. | N/A | |
| 46 | f/3.2 | 5 s | Low | 21 frames | 36 MeV | 20 nA | 20 mm | Beam on. | ||
| 47 | f/4 | 5 s | N/A | N/A | 36 MeV | 20 nA | 20 mm | Beam on. | N/A | |
| 48 | f/5 | 5 s | N/A | N/A | 36 MeV | 15 nA | 20 mm | Beam on. | N/A | |
| 49 | f/5 | 5 s | Low | 21 frames | 36 MeV | 50 nA | 20 mm | Beam on. Both camera and sensor saturated. | ||
| 50 | f/8 | 5 s | N/A | N/A | 36 MeV | 50 nA | 20 mm | Beam on. Variable beam during acquisition. | N/A | |
| 51 | f/8 | 5 s | N/A | N/A | 36 MeV | 50 nA | 20 mm | Beam on. Stable beam. | N/A | |
| 52 | f/8 | 5 s | High | 21 frames | 36 MeV | 50 nA | 20 mm | Beam on. | ||
| 53 | f/8 | 5 s | High | 21 frames | N/A | N/A | N/A | Beam off. Background run for sensor in high full well mode. | ||
| 54 | f/8 | 5 s | High | 21 frames | 36 MeV | 100 nA | 20 mm | Beam on. | ||
| 55 | f/8 | 2 s | N/A | N/A | 36 MeV | 90 nA | 20 mm | Beam on. | N/A | |
| 56 | f/8 | 2 s | High | 21 frames | 36 MeV | 200 nA | 20 mm | Beam on. | ||
| 57 | f/8 | 1 s | N/A | N/A | 36 MeV | 200 nA | 20 mm | Beam on. Sensor near saturation. | N/A | |
| 58 | f/8 | 1 s | N/A | N/A | 36 MeV | 300 nA | 20 mm | Beam on. Variable beam during acquisition. | N/A | |
| 59 | f/8 | 1 s | High | 21 frames | 36 MeV | 300 nA | 20 mm | Beam on. Stable beam. Sensor saturation occurs. | ||
| 60 | f/8 | 1 s | High | 21 frames | 36 MeV | 450 nA | 20 mm | Beam on. | ||
| 61 | f/8 | 1/2 s | N/A | N/A | 36 MeV | 450 nA | 20 mm | Beam on. | N/A | |
| 62 | f/8 | 1/2 s | N/A | N/A | 36 MeV | 470 nA | 20 mm | Beam on. | N/A | |
| 63 | f/13 | 1/2 s | N/A | N/A | 36 MeV | 600 nA | 20 mm | Beam on. | N/A | |
| 65 | f/13 | 1/2 s | Low | 21 frames | 36 MeV | 1.1 uA | 20 mm | Beam on. | ||
| 66 | f/13 | 1/2 s | N/A | N/A | 36 MeV | 1 uA | 20 mm | Beam on. Low beam during acquisition. | N/A | |
| 67 | f/13 | 1/2 s | N/A | N/A | 36 MeV | 1 uA | 20 mm | Beam on. Stable beam. | N/A | |
| 68 | f/8 | 1 s | High | 21 frames | 36 MeV | 200 nA | 20 mm | Beam on. | ||
| 69 | f/3.2 | 5 s | High | 21 frames | 36 MeV | 13 nA | 20 mm | Beam on. Some beam captured on DSLR image. | ||
| 70 | f/3.2 | 5 s | Low | 21 frames | 36 MeV | 15 nA | 20 mm | Beam on. | ||
| 71 | f/3.2 | 2 s | N/A | N/A | 36 MeV | 15 nA | 20 mm | Beam on. | N/A | |
| 72 | f/3.2 | 1 s | Low | 21 frames | 36 MeV - 3 mm PMMA | 30 nA | 20 mm | Beam on. 3 mm of PMMA taped over Mylar window. | ||
| 73 | f/3.2 | 1 s | Low | 21 frames | 36 MeV - (3 mm + 2 mm PMMA) | 30 nA | 20 mm | Beam on. 3 mm + 2 mm of PMMA taped over Mylar window. | ||
| 74 | f/3.2 | 1 s | N/A | N/A | 36 MeV - (3 mm + 2 mm PMMA) | 30 nA | 20 mm | Beam on. 3 mm + 2 mm of PMMA taped over Mylar window. | N/A | |
| 75 | f/3.2 | 1 s | Low | 21 frames | 36 MeV | 25 nA | 20 mm | Beam on. Green film inserted between scintillator stack and sensor. | ||
| 76 | f/8 | 1 s | Low | 21 frames | 36 MeV | 100 nA | 20 mm | Beam on. Green film inserted between scintillator stack and sensor. | ||
| 77 | f/8 | 1/2 s | Low | 21 frames | 36 MeV | 500 nA | 20 mm | Beam on. Green film inserted between scintillator stack and sensor. Sensor saturated on low full well mode. | ||
| 78 | f/8 | 1/2 s | High | 21 frames | 36 MeV | 480 nA | 20 mm | Beam on. Green film inserted between scintillator stack and sensor. Sensor not saturated on high full well mode. | ||
| 79 | f/13 | 1/2 s | High | 21 frames | 36 MeV | 900 nA | 20 mm | Beam on. Green film inserted between scintillator stack and sensor. Sensor shows some saturation on high full well mode. | ||
| 80 | f/13 | 1/2 s | High | 21 frames | 36 MeV | 700 nA | 20 mm | Beam on. Green film inserted between scintillator stack and sensor. Sensor saturation occurs. | ||
| 81 | f/13 | 1/2 s | High | 21 frames | 36 MeV | 1 uA | 20 mm | Beam on. Green + blue films inserted between scintillator stack and sensor. | ||
| 82 | f/13 | 1/2 s | N/A | N/A | 36 MeV | 1.1 uA | 20 mm | Beam on. Green + blue films inserted between scintillator stack and sensor. | N/A | |
| 83 | f/13 | 1/2 s | High | 21 frames | 36 MeV | 1.1 uA | 20 mm | Beam on. Green + blue films inserted between scintillator stack and sensor. Some beam captured on DSLR image. | ||
| 84 | f/3.2 | 1 s | Low | 21 frames | 36 MeV - 2 mm PMMA | 30 nA | 20 mm | Beam on. Green + blue films inserted between scintillator stack and sensor. 2 mm PMMA sheet taped in front of Mylar window. | ||
| 85 | f/3.2 | 1 s | Low | 21 frames | 36 MeV - 1 mm PMMA | 30 nA | 20 mm | Beam on. Green + blue films inserted between scintillator stack and sensor. 1 mm PMMA sheet taped in front of Mylar window. | ||
| 86 | f/3.2 | 5 s | Low | 21 frames | 36 MeV | 10 nA | 20 mm | Beam on. All films and PMMA sheets removed. Back to original settings to monitor scintillator damage. | ||
| 87 | f/3.2 | 5 s | Low | 21 frames | 36 MeV | 10 nA | 20 mm | Beam on. Back to original settings to monitor scintillator damage. | ||
| 88 | f/3.2 | 5 s | Low | 21 frames | 28 MeV | 10 nA | 20 mm | Beam on. Beam energy reduced to 28 MeV using RF frequency. | ||
| 89 | f/3.2 | 5 s | N/A | N/A | 28 MeV | 10 nA | 20 mm | Beam on. | N/A | |
| 90 | f/3.2 | 1 s | Low | 21 frames | 28 MeV | 30 nA | 20 mm | Beam on. | ||
| 91 | f/3.2 | 1 s | High | 21 frames | 28 MeV | 30 nA | 20 mm | Beam on. | ||
| 92 | f/8 | 1 s | High | 21 frames | 28 MeV | 300 nA | 20 mm | Beam on. | ||
| 93 | f/8 | 1 s | High | 21 frames | 28 MeV | 250 nA | 20 mm | Beam on. |