Proton Calorimetry/Experimental Runs/2018/Aug16
Beam tests of the UCL plastic scintillator range calorimeter.
Night of 15th to 16th of August 2018
Pre-data Taking Information and Notes
Sheet ordering from front to end (in beam direction): 31,25,24,23,20,16,12,11,9,8,6,5,4,3,2,1,30,29,28,27,26,22,19,18,17,15,14,13,10,7,21
100% beam intensity is about 10e6 particles per second
The MedAustron Ripple filter used in this experiment has a WET of 3mm.
The sensor pixel dimensions are 1030 x 1536 which has to be set manually
Serial Terminal instructions for ISDI sensor:
In Controls/Options: Make sure "Send string with CR" is selected. Should be default.
In Controls/Setup: Select Serial Port: Enabled. Baud rate has to be 115200
Set low full well mode using W000300000
Set high full well mode using W000300004
Read full well mode using R0003
Data Taking
1 hour of beam time: 30min setup, 30 min measurement time
Mistake during data taking: Full well mode should have been "high" in order to have a higher saturation threshold!
Run number | Full Well Mode | Beam Energy (MeV/u) | Intensity Degrader | Test type | Comments | Thumbnail |
---|---|---|---|---|---|---|
0 | low | 0 | 10% | background | two ripple filter in! have a WET of 3.0mm until stated otherwise | |
1 | low | 184.5 | 10% | energy scan | ||
2 | low | 184.5 | 10% | energy scan | repeat | |
3 | low | 184.5 | 10% | energy scan | repeat | |
4 | low | 400 | 10% | calibration RF | calibration shootthrough with ripple filters! | |
5 | low | 400 | 10% | calibration RF | repeat | |
6 | low | 154.5 | 10% | energy scan | ||
7 | low | 146.5 | 10% | energy scan | ||
8 | low | 138.3 | 10% | energy scan | ||
9 | low | 121.1 | 10% | energy scan | ||
10 | low | 120 | 10% | energy scan | ||
11 | low | 0 | 10% | background | ripple filter out! Background | |
12 | low | 400 | 10% | calibration | calibration shootthrough | |
13 | low | 184.5 | 10% | energy scan | ||
14 | low | 184.5 | 10% | energy scan | repeat | |
15 | low | 154.5 | 10% | energy scan | ||
16 | low | 146.5 | 10% | energy scan | ||
17 | low | 138.3 | 10% | energy scan | ||
18 | low | 121.1 | 10% | energy scan | ||
19 | low | 120 | 10% | energy scan | ||
20 | low | 184.5 | 10% | radiation damage | check for radiation damage | |
21 | low | 184.5 | 10% | radiation damage | repeat |