Proton Calorimetry/Experimental Runs/2017/Mar1-3
Include all details from experimental runs.
PAGE UNDER CONSTRUCTION
01-02/03/2017
Hamamatsu Base + Standard 45 cm Cuboid Tests
High Voltage (V) | Trigger | Rate + Additional Information | Electronics Setup | Collimator Size + Additional Information | Run Number |
---|---|---|---|---|---|
-900 | ST | I (Beam off): 150.15 μA |
|
|
|
-800 | ST |
|
|
5 second beam spill | 001 |
-800 | ST |
|
002 | ||
-800 | ST |
|
003 | ||
-800 | ST |
|
30 second beam spill | 004 | |
-800 | ST |
|
005 | ||
-800 | ST |
|
006 | ||
-800 | ST |
|
007 | ||
-800 | ST | Increase acquisition gate to see pile up effect |
|
008 | |
-800 | ST |
|
009 | ||
-900 | ST |
|
010 | ||
-1000 | ST |
|
|
011 | |
-1000 | ST |
|
012 | ||
-1000 | ST |
|
013 | ||
-1000 | ST |
|
014 | ||
-1000 | ST |
|
015 | ||
-1000 | ST |
|
016 | ||
-1000 | ST | 62 MeV beam (runs 017-019) |
|
62 MeV beam | 017 |
-1000 | ST |
|
018 | ||
-1200 | ST |
|
019 | ||
-1200 | ST | 100 MeV beam (run 020) |
|
100 MeV beam | 020 |
End of test beam |
02-03/03/2017
Continuation of Hamamatsu Tests
High Voltage (V) | Trigger | Rate + Additional Information | Electronics Setup | Collimator Size + Additional Information | Run Number |
---|---|---|---|---|---|
-1000 | ST |
|
|
||
|
|
021 | |||
-1000 | ST |
|
|
022 | |
-1000 | ST |
|
102.4 MeV beam | 023 | |
-1000 | ST | Beam shifted during run | 024 | ||
-1000 | ST | Beam shifted during run | 025 | ||
-1000 | ST | 62.4 MeV beam | 026 | ||
-1000 | ST |
|
027 | ||
-1000 | ST | 252.4 MeV beam | 028 | ||
-1000 | ST |
|
All electronics as above | 029 | |
-1000 | ST |
|
030 | ||
-1000 | ST |
|
031 | ||
-1000 | ST |
|
032 | ||
-1000 | ST |
|
033 | ||
-1000 | ST | 034 | |||
-1000 | ST |
|
035 | ||
-1000 | ST | 036 | |||
-1000 | ST | 037 | |||
-1000 | ST | 038 | |||
-1000 | ST | 232.4 MeV | 039 | ||
-1000 | ST | 040 | |||
-1000 | ST |
|
212.4 MeV | 041 | |
042 | |||||
043 | |||||
-1000 | ST |
|
192.4 MeV | 044 | |
-1000 | ST |
|
172.4 MeV | 045 | |
046 | |||||
-1000 | ST | 047 | |||
-1000 | ST | 152.4 MeV | 048 | ||
-1000 | ST | 049 | |||
-1000 | ST | 132.4 MeV | 050 | ||
-1000 | ST | 051 | |||
-1000 | ST | Strange oscillating behaviour between two points at this energy | 112.4 MeV | 052 | |
-1000 | ST | 053 | |||
-1000 | ST | 92.4 MeV | 054 | ||
-1000 | ST | 055 | |||
-1000 | ST | 72.4 MeV | 056 | ||
-1000 | ST | 057 | |||
-1000 | ST | 62.4 MeV | 058 | ||
-1000 | ST | 059 | |||
-1000 | ST | Back to 112.4 MeV to investigate strange energy. At the beginning of the spill the small higher ADC count peak starts growing, then moves onto "main" peak and grow until the end of the run | 112.4 MeV | 060 | |
-1000 | ST |
|
061 | ||
-1000 | ST |
|
64.2 MeV | 062 | |
-1000 | ST |
|
Peak shifts to lower ADC | 063 | |
-1000 | ST |
|
Peak shifts back up | 064 | |
-1000 | ST |
|
Peak shifts down again | 065 | |
-1000 | ST |
|
066 | ||
-1000 | ST | Same oscillating behaviour as for energy 112 MeV, but a lit less extreme | 067 | ||
-1000 | ST |
|
068 | ||
-1000 | ST |
|
069 | ||
-1000 | ST | 070 | |||
-1000 | ST |
|
5 second beam spill | 071 | |
-1000 | ST |
|
072 | ||
-1000 | ST | 073 | |||
-1000 | ST | Next step up in degrader, 50%. The data becomes invalid at this point. Flat top intensity: 8.8×109 | 074 | ||
-1000 | ST |
|
|
075 | |
-1000 | ST | Waveforms (~500 MB) | 076 | ||
-1000 | ST |
|
Waveforms | 077 | |
End of test beam |
Counts on integrated intensity monitor: 4,523,000 at 22:53 GMT Final count: 33,600,000 at 04:03 GMT 1×104 to 2×104 protons per count