Proton Calorimetry/Experimental Runs/2015/Jul8-9
Include all details of experimental runs from Anastasia's PDFs.
PAGE UNDER CONSTRUCTION
Day 1 (July 8th)
Notes
- Current at -800 V: 129.6 μA (beam off)
- Background rate (-800 V on PMT, beam off, no collimator): ~5 Hz - 4 Hz
- Threshold: -14 mV (for 10 segments)
- Rise time: 8 - 14 ns (<14 ns)
- Having the corridor lights on does NOT create enough light to trip PMT and is acceptable
- For treatment, 8 mm - 20 mm collimator used
Negative Base Tests
High Voltage (V) | Trigger | Rate + Additional Information | Electronics Setup | Collimator Size + Additional Information | Run Number |
---|---|---|---|---|---|
-800 | ST | Rate: 50 Hz |
|
||
Rate: 500 kHz |
|
||||
Rate: ~200 kHz |
|
|
|||
-800 | ST | Rate: 50 Hz |
|
|
001 (60,000 events) |
-800 | ST, list |
|
|
5 mm | 002 (71,000 events) |
-800 | ST, list |
|
|
8 mm | 003 (350,000 events) |
40 MeV protons, therefore trigger on 14 mV is ~5%, usually require 10%. Trigger on 30 mV threshold from now on. | |||||
Lunch break | |||||
-800 | ST, list |
|
|
16 mm | 005 (350,000 events) |
-800 | ST, list | As above, repeated to study shoulder at ~2500 ADC | 006 (300,000 events) | ||
-800 | ST, list |
|
|
20 mm | 007 (400,000 events) |
-800 | ST, list |
|
|
24 mm | 008 (400,000 events) |
-800 | ST, list |
|
|
30 mm | 009 (400,000 events) |
-800 | ST, list |
|
|
|
010 (1,500,000 events) |
-800 | ST, list |
|
|
|
012 (1,500,000 events |
-800 | ST, list | All as above, but create larger statistics | 014 (3,000,000 events) | ||
-800 | ST, list |
|
|
|
016 (1,500,000 events) |
-800 | ST, list |
|
|
4 mm | 017 (3,000,000 events) |
-800 | ST, list |
|
|
3 mm | 018 (3,000,000 events) |
-800 | ST, list |
|
|
5 mm | 019 (5,000,000 events) |
-800 | ST, list |
|
|
6 mm | 020 (list), 021 (eh) (5,000,000 events) |
-800 | ST, list |
|
|
Voltage changed on Ds: 47.4% | 025 (list), 027 (eh) (4,000,000 events) |
-800 | ST, list |
|
|
Voltage on Ds changed back | 028 (3,000,000 events) |
-800 | ST, list | All as above, take second for crosscheck | 029 (2,300,000 events) | ||
-800 | ST, list |
|
|
6.99 mm | 030 (3,000,000 events) |
-800 | ST, list |
|
039 (4,000,000 events) | ||
-800 | ST, list |
|
|
|
044 (3,500,000 events) |
-800 | ST, list |
|
|
|
045 (4,000,000 events) |
-800 | ST, list |
|
|
|
046 (4,000,000 events) |
-800 | ST, list |
|
|
|
049 (4,000,000 events) |
-800 | ST, list |
|
051 (4,000,000 events) | ||
-800 | ST, list |
|
053 (4,000,000 events) | ||
-800 | ST, list |
|
054 (4,000,000 events) | ||
-800 | ST, list |
|
|
056 (4,000,000 events) | |
-800 | ST, list |
|
057 (4,000,000 events) | ||
-800 | ST, list |
|
|
0.5 cc/minute | 065 (4,000,000 events) |
Day 2 (July 9th)
Notes
- From the morning of 9th July 2015 measurements taken without the black cloth covering the face of the scintillator.
- Dark rates have increased to ~6.5 Hz (for a -30 mV threshold)
- Current for dark rate: 129.66 μA
- Distance between beam and EJ-Block scintillator: ~40 mm
- No dark rate, background pickup up with beam off due to such a small scintillator size.
- At 1450 V, I: 324 μA, but still no visible signal. Wrong PMT connected to signal.
With correct PMT:
- At 800 V dark current + background rate: ~40 Hz at -30 mV threshold, I: 178.6 μA
- At 1000 V, dark current + background rates: 600 Hz at -30 mV threshold
Negative Base Tests
High Voltage (V) | Trigger | Rate + Additional Information | Electronics Setup | Collimator Size + Additional Information | Run Number |
---|---|---|---|---|---|
-800 | ST, list |
|
|
|
068 (2,000,000 events) |
-800 | ST, list |
|
Unable to take run, too much instability | ||
-800 | ST, list |
|
|
No run, unstable | |
-800 | ST, list | Rate: 100 kHz, unstable |
|
No run, unstable | |
-800 | ST, list |
|
|
|
075 (250,000 events) |
-800 | ST, list |
|
076 (250,000 events) | ||
-800 | ST, list |
|
077 (230,000 events) | ||
-800 | ST, list |
|
078 (230,000 events) | ||
-800 | ST, waves |
|
079 (Waveforms) | ||
-800 | ST, list |
|
|
|
081 (2,000,000 events) |
-800 | ST, list |
|
085 (2,000,000 events) | ||
-800 | ST, list |
|
087 (2,000,000 events) | ||
-800 | ST, list | Note: Run shifts towards the end |
|
089 (2,000,000 events) | |
-800 | ST, list | Note: Repeat of above run, looks cleaner |
|
092 (2,000,000 events) | |
-800 | ST, waves |
|
093 (Waveforms) | ||
-800 | ST, list |
|
|
|
095 (2,000,000 events) |
-800 | ST, list |
|
096 (2,000,000 events) | ||
-800 | ST, list |
|
097 (2,000,000 events) | ||
-800 | ST, list |
|
098 (2,000,000 events) | ||
-800 | ST, list |
|
|
|
099 (2,000,000 events) |
-800 | ST, list |
|
100 (2,000,000 events) | ||
-800 | ST, list |
|
101 (2,000,000 events) | ||
-800 | ST, list |
|
102 (2,000,000 events) | ||
-800 | ST, list |
|
|
3 mm | 105 (7,500,000 events) |
-800 | ST, list |
|
106 (ls), 107 (eh) (7,000,000 events) | ||
Beam tripped out | |||||
-800 | ST, list |
|
108 (9,000,000 events) | ||
-800 | ST, list |
|
109 (ls), 110 (eh) (6,500,000 events) |
3" PMT Tests
High Voltage (V) | Trigger | Rate + Additional Information | Electronics Setup | Collimator Size + Additional Information | Run Number |
---|---|---|---|---|---|
800 | ST, list |
|
Signal is off the ADC scale! Reduce the voltage! |
|
|
700 | ST, list |
|
|
113 (165,000 events) | |
700 | ST, list |
|
|
20 mm | 116 (460,000 events) |
700 | ST, list |
|
|
5.07 mm of PMMA (1 plate) | 118 (715,000 events) |
700 | ST, list |
|
9.8 mm of PMMA (1 plate) | 119 (370,000 events) | |
700 | ST, list | Repeat above measurement after beam reset |
|
120 (800,000 events) | |
700 | ST, list |
|
14.12 mm of PMMA (1 plate) | 121 (900,000 events) | |
700 | ST, list |
|
17.94 mm of PMMA (9.04 mm + 10.9 mm, 2 plates) | 122 (950,000 events) | |
700 | ST, list |
|
19.94 mm of PMMA (9.04 mm + 10.9 mm, 2 plates) | 123 (805,000 events) | |
700 | ST, list |
|
21.70 mm of PMMA (9.04 mm + 12.66 mm, 2 plates) | 124 (550,000 events) | |
700 | ST, list |
|
|
No PMMA | 125 (700,000 events) |