Proton Calorimetry/Experimental Runs/2018/Aug16
Beam tests of the UCL plastic scintillator range calorimeter.
Night of 15th to 16th of August 2018
Pre-data Taking Information and Notes
Sheet ordering from front to end (in beam direction): 31,25,24,23,20,16,12,11,9,8,6,5,4,3,2,1,30,29,28,27,26,22,19,18,17,15,14,13,10,7,21
100% beam intensity is about 10e6 particles per second
Set low full well mode using W000300000
Set high full well mode using W000300004
Baud rate has to be 115200. Also check that the serial port is enabled
Read full well mode using R0003
The sensor pixel dimensions are 1030 x 1536 which has to be set manually
The MedAustron Ripple filter used in this experiment has a WET of 3mm.
Data Taking
1 hour of beam time: 30min setup, 30 min measurement time
Mistake during data taking: Full well mode should have been "high" in order to have a higher saturation threshold!
Run number | Full Well Mode | Beam Energy (MeV/u) | Intensity Degrader | Test type | Comments | Thumbnail |
---|---|---|---|---|---|---|
0 | low | 0 | 10% | background | two ripple filter in! have a WET of 3.0mm until stated otherwise | |
1 | low | 184.5 | 10% | energy scan | ||
2 | low | 184.5 | 10% | energy scan | repeat | |
3 | low | 184.5 | 10% | energy scan | repeat | |
4 | low | 400 | 10% | calibration RF | calibration shootthrough with ripple filters! | |
5 | low | 400 | 10% | calibration RF | repeat | |
6 | low | 154.5 | 10% | energy scan | ||
7 | low | 146.5 | 10% | energy scan | ||
8 | low | 138.3 | 10% | energy scan | ||
9 | low | 121.1 | 10% | energy scan | ||
10 | low | 120 | 10% | energy scan | ||
11 | low | 0 | 10% | background | ripple filter out! Background | |
12 | low | 400 | 10% | calibration | calibration shootthrough | |
13 | low | 184.5 | 10% | energy scan | ||
14 | low | 184.5 | 10% | energy scan | repeat | |
15 | low | 154.5 | 10% | energy scan | ||
16 | low | 146.5 | 10% | energy scan | ||
17 | low | 138.3 | 10% | energy scan | ||
18 | low | 121.1 | 10% | energy scan | ||
19 | low | 120 | 10% | energy scan | ||
20 | low | 184.5 | 10% | radiation damage | check for radiation damage | |
21 | low | 184.5 | 10% | radiation damage | repeat |