Proton Calorimetry/Experimental Runs/2017/September7: Difference between revisions
No edit summary |
No edit summary |
||
Line 8: | Line 8: | ||
*From previous measurements, the average background value per pixel is ~ 700. | *From previous measurements, the average background value per pixel is ~ 700. | ||
*The data was recorded at '''45 frames/second'''. | *The data was recorded at '''45 frames/second'''. | ||
*The scintillator stack was set up perpendicular to the proton beam, with the exposed sides of the scintillator facing the sensors: | |||
<div class="image400px" style="text-align: center;"> | |||
[http://www.hep.ucl.ac.uk/pbt/wikiData/images/Birmingham_sept2017/setup_1.jpg http://www.hep.ucl.ac.uk/pbt/wikiData/images/Birmingham_sept2017/setup_1.jpg] | |||
[http://www.hep.ucl.ac.uk/pbt/wikiData/images/Birmingham_sept2017/setup_2.jpg http://www.hep.ucl.ac.uk/pbt/wikiData/images/Birmingham_sept2017/setup_2.jpg] | |||
<br /> | |||
</div> | |||
*The scintillator stack was quite well aligned with the horizontal centre of the beam. We did our best to align the vertical centre of the scintillator stack with the beam as well, however as can be seen from the image below, where the black "line" shows the divide between and hence the centre of the sensors, the beam hit just below centre: | |||
<div class="image400px" style="text-align: center;"> | |||
[http://www.hep.ucl.ac.uk/pbt/wikiData/images/Birmingham_sept2017/off_centre.jpg http://www.hep.ucl.ac.uk/pbt/wikiData/images/Birmingham_sept2017/off_centre.jpg] | |||
<br /> | |||
</div> | |||
*The working directory on the PRIAPUS computer where the data is stored is: | *The working directory on the PRIAPUS computer where the data is stored is: | ||
Line 53: | Line 65: | ||
<br /> | <br /> | ||
</div> | </div> | ||
|| | *This setup remains in place until mentioned otherwise | ||
*Lights were left on for the first run. Data not useable! | |||
|| 28MeV_3mmscint_ABC_spread_10pA_1cmC | |||
|- |
Revision as of 13:28, 19 September 2017
Beam tests of the Nuvia 10 mm, 4 mm and 3 mm scintillator sheets with PRaVDA CMOS pixel sensors (PRIAPUS) at the Birmingham Cyclotron with a 28 MeV beam.
Pre-data Taking Information and Notes
- The tests were carried out with an energy of 28 MeV (32 MeV was unavailable as the cyclotron had recently been through repairs). No Bragg Peak measurement was carried out to confirm the exact beam energy due to time restrictions.
- The beam was run without a scattering foil. A 1 cm diameter uniform beam was achieved by placing a 1 cm collimator in the beam, with a beam scan onto film confirming the size. An ionisation chamber was placed upstream of the collimator to accurately monitor the current.
- The dynamic range of the CMOS sensors is 14 bits. The electron conversion value from ADC is 50 electrons per ADC value. The measurements to establish this were carried out at the beginning of experimentation with the sensors and this value may have changed due to radiation damage etc.
- From previous measurements, the average background value per pixel is ~ 700.
- The data was recorded at 45 frames/second.
- The scintillator stack was set up perpendicular to the proton beam, with the exposed sides of the scintillator facing the sensors:
- The scintillator stack was quite well aligned with the horizontal centre of the beam. We did our best to align the vertical centre of the scintillator stack with the beam as well, however as can be seen from the image below, where the black "line" shows the divide between and hence the centre of the sensors, the beam hit just below centre:
- The working directory on the PRIAPUS computer where the data is stored is:
idMateData\Bham070917\
- The directory the data is stored at UCL in is:
/unix/pbt/data/birmingham/sept_2017
Data Taking
Type of run | Current | Scintillator sheets used | Scintillator arrangement | Notes | File name |
---|---|---|---|---|---|
Background | N/A | 3 x 3 mm | ABC |
|
Background_1 |
Beam | 1 nA | 3 x 3 mm | ABC | Saturation of sensor, ADC values at 14,000 | 28MeV_3mmscint_ABC_1nA_1cmC |
Beam | 0.1 nA | 3 x 3 mm | ABC |
|
28MeV_3mmscint_ABC_100pA_1cmC_1 |
Beam | 0.1 nA | 3 x 3 mm | ABC | Saturation of sensor | 28MeV_3mmscint_ABC_100pA_1cmC_2 |
Beam | 10 pA | 3 x 3 mm | ABC | Sensor no longer saturated, but no separation due to formation of tiles | 28MeV_3mmscint_ABC_10pA_1cmC |
Beam | 10 pA | 3 x 3 mm | ABC |
|
28MeV_3mmscint_ABC_spread_10pA_1cmC |