Proton Calorimetry/Experimental Runs/2025/Trento 2025-03: Difference between revisions
Line 36: | Line 36: | ||
# Set up mains cabling and networking in treatment and control rooms. | # Set up mains cabling and networking in treatment and control rooms. | ||
## no beam | |||
# Install Peli case on treatment couch. | # Install Peli case on treatment couch. | ||
# Calibration shoot-through measurements of 4 | ## no beam | ||
# Background | |||
## no beam | |||
# Calibration shoot-through measurements of 4 stack modules: either rotating case or modules for back shoot-through. | |||
## 8 beam deliveries (2 beam deliveries (back and forth) per each module): | |||
{| class="wikitable" | |||
!Run number | |||
!Detector config | |||
!Beam Energy (MeV) | |||
!Beam Current (nA) | |||
!Physical thickness (mm) | |||
!D80 measured (mm) | |||
!WET (mm) | |||
!RSP | |||
!Comments | |||
|- | |||
|33 || Stack 1 front || 245 || 18.8 (18.7) || 95.95 || 198.69 || 98.24 || 1.024 || | |||
|- | |||
|34 || Stack 1 back || 245 || 18.8 (18.7) || 95.95 || 198.74 || 98.19 || 1.023 || | |||
|- | |||
|36 || Stack 2 front || 245 || 18.8 (18.7) || 96.81 || 198.51 || 98.42 || 1.016 || | |||
|- | |||
|37 || Stack 3 front || 245 || 18.8 (18.7) || 95.48 || 199.78 || 97.15 || 1.017 || | |||
|- | |||
|40 || Stack 4 front || 245 || 18.8 (18.7) || 94.07 || 202.26 || 94.67 || 1.006 || | |||
|- | |||
|41 || Empty Box || 245 || 18.8 (18.7) || 0.012 || 296.93 || 0 || 0 || | |||
|- | |||
|42 || No Box || 245 || 18.8 (18.7) || 0 || 296.90 || 0 || 0 || | |||
|} | |||
# Pristine Bragg peak measurements at clinical current for full clinical range. | # Pristine Bragg peak measurements at clinical current for full clinical range. | ||
# FLASH measurements at 245 MeV. | # FLASH measurements at 245 MeV. |
Revision as of 10:20, 4 February 2025
Experiment Equipment
Item | Notes |
---|---|
Network Hub | Set in control room to take output from experimental room ethernet connection. Control laptops connected via 5GHz WiFi. |
Control Laptop x2 | 1 for remote control of FPGA, 1 for notes/web GUI. |
Ethernet Cable x 2 | To connect DAQ laptop to network in the experimental room, control laptop to network in control room. |
Portable Enclosure | |
Scintillator stacks | 4 Modules (to cover the full clinical range, from 60 to 250 MeV) |
Scintillating fibers module | 1 2D Module |
DAQ laptop x1 | Control photodiode acquisition. |
Nexys Video FPGA development board. | For interfacing between DDC232 and PC. |
Schematic
Experiment Plan
- Set up mains cabling and networking in treatment and control rooms.
- no beam
- Install Peli case on treatment couch.
- no beam
- Background
- no beam
- Calibration shoot-through measurements of 4 stack modules: either rotating case or modules for back shoot-through.
- 8 beam deliveries (2 beam deliveries (back and forth) per each module):
Run number | Detector config | Beam Energy (MeV) | Beam Current (nA) | Physical thickness (mm) | D80 measured (mm) | WET (mm) | RSP | Comments |
---|---|---|---|---|---|---|---|---|
33 | Stack 1 front | 245 | 18.8 (18.7) | 95.95 | 198.69 | 98.24 | 1.024 | |
34 | Stack 1 back | 245 | 18.8 (18.7) | 95.95 | 198.74 | 98.19 | 1.023 | |
36 | Stack 2 front | 245 | 18.8 (18.7) | 96.81 | 198.51 | 98.42 | 1.016 | |
37 | Stack 3 front | 245 | 18.8 (18.7) | 95.48 | 199.78 | 97.15 | 1.017 | |
40 | Stack 4 front | 245 | 18.8 (18.7) | 94.07 | 202.26 | 94.67 | 1.006 | |
41 | Empty Box | 245 | 18.8 (18.7) | 0.012 | 296.93 | 0 | 0 | |
42 | No Box | 245 | 18.8 (18.7) | 0 | 296.90 | 0 | 0 |
- Pristine Bragg peak measurements at clinical current for full clinical range.
- FLASH measurements at 245 MeV.