Proton Calorimetry/Meetings/2021/11/03
Minutes for UCL Proton Calorimetry Meetings, 3rd November
Present
Simon Jolly, Saad Shaikh, Raffaella Radogna
Saad Shaikh
- Completed beam test at UCLH and will begin data analysis.
- Overall resounding success:
- Electronic noise significantly reduced (factor 10 reduction)
- Live fitting worked reasonably well.
- Questions to be answered:
- Light output did not improve: optical gel likely not used properly and scintillator sheets could be damaged from removal of black paint and application of white paint. Light coupling investigations will be required to optimise this.
- Fits failed at times: significantly improved by using ROOT's "binned" fit, at the expense of frame rate. Need to investigate optimal fit range and if parameters (particularly sigma) need to be constrained.
- Appear to observe 2-3mm systematic offset between reconstructed Bragg curve and reference Bragg curve – need to investigate source of this.
- Overall resounding success:
- Matt has provided first draft for fix of FPGA code problem with signals crossing clock domains.
- Will test and debug.
- Simon to ask CosyLab for additional daughter board that connects with Zmod and to ask Matt to port design to USB104.