07 Oct 2020


LED injected light measurements

Setup: - scintillator sheets number from bottom to top in sensor image: 51, 31, 50, 52, 34, 44, 53, 47, 33, 54, 46, 42, 55, 45, 38, 36, 43, 32, 48, 41, 8, 39, 40, 37, 35, 49, 9 + other sheet as fillers - CMOS ISDI Image sensor. Sensor was allowed to heat up for an hour before measurements - 16 Photodiodes covering the sheets sequence 42, 55, 45, 38, 36, 43, 32, 48, 41, 8, 39, 40, 37, 35, 49, 9. - Optical fibre used with optical gel to couple LED to scintillator sheet n. 8

Aim: Compare light output from sensor and photodiodes. Plan is to insert light into sheet n. 8 and measure the light output with the CMOS sensor and different photodiodes arrays.

list of runs in this file unix/pbt/data/ucl/20201007/20201007_LEDandADCBoard.xlsx data saved in unix/pbt/data/ucl/20201007/

Run LED Voltage (V) LED Period (us) Width (ns) Sensor Full Well Mode Illuminated sheet DDC int time DDC Full scale DDC Board and photodiodes Notes
1 4.5 1 300 high 8 170us 350pC FSR Saad's (PDs: S1337-16BR) PD placed opposite to the LED fibre
2 4.5 1 300 high 8 170us 350pC FSR Saad's (PDs: S1337-16BR) Sensor removed and PD placed 90deg to the fibre, in the spot of highes intensity seen by the sensor
3 4.5 1 300 high 8 2000us 350pC FSR Saad's (PDs: S1337-16BR)
4 4.5 1 300 high 8 2000us 12.5pC FSR Saad's (PDs: S1337-16BR)
5 4.5 1 300 high 8 20000us 100pC FSR Saad's (PDs: S1337-16BR)
6 4.5 1 300 high 8 170us 12.5pC FSR Saad's (PDs: S1337-16BR)
7 4.5 1 300 high 8 170us 12.5pC FSR Raffy's (PDs: S12915-16R)
8 4.5 1 300 high 8 170us 12.5pC FSR Raffy's (PDs: S12915-16R) large statistics for noise analysis
9 4.5 1 300 high 8 170us 350pC FSR Saad's (PDs: S1337-16BR) large statistics for noise analysis
10 4.5 1 300 high 8 2000us 12.5pC FSR Raffy's (PDs: S12915-16R) large statistics for noise analysis
11 4.5 1 300 high 8 2000us 100pC FSR Raffy's (PDs: S12915-16R) large statistics for noise analysis