Simulation of the Clatterbridge beamline

This simulation is a model of the monoenergetic 62.5 MeV proton beam at the Clatterbridge Cancer Centre [1] as it traverses the components of the beamline and finally hits a volume of water. The simulation was built using the existing example "monoenergetic proton pencil beam" [2]. The physics list used is QGSP_BIC_EMY, the standard for simulating clinical proton beams.

Geometry

Treatment room

The schematic below shows the Clatterbridge treatment room, illustrating the layout of the geometries defined in the DetectorConstruction class. The whole beamline is contained within the "inner room" volume, which is placed in the "outer room" volume. As the outer volume is a solid concrete box and the inner volume is a box filled with air, the small overlap models the concrete walls of the treatment room. The proton source is placed against the wall in the inner room (-4200 mm from the origin), and all beamline components are placed relative to this reference point.



Beamline

The beamline is shown below. The top figure is a perspective view. The second figure is a top-down view which also shows the borated plastic shielding (f) that was previously hidden to allow the full beamline to be visible.


The beamline consists of the following components:

  • an aluminium tube (a) containing from left to right;
    • a brass collimator
    • a tungsten scattering foil
    • a brass beam stopper
    • another tungsten scattering foil
    • a mylar window
  • an empty aluminium box (b)
  • an iron block (c)
  • a second aluminium tube (d)
  • a second aluminium box (e) containing from left to right;
    • a brass collimator
    • two dose monitors
    • cross wires
  • a brass nozzle (g)

The water phantom (h) is also shown for completeness.


Dual scattering tube

The proton beam at Clatterbridge is delivered using passive spreading. That is, the beam is spread out using dual scattering where the proton beam is scattered through a first scattering foil followed by a beam stopper and a second scattering foil. The beam stopper is used to reduce the intensity of the beam at its centre and a high-Z material such as tungsten is chosen for the foils as it is highly scattering.

A visualisation of the first tube with 500 primary protons represented by blue tracks is shown below. The protons first travel through the collimator, followed by the first scattering foil which spreads out the beam. The brass stopper then blocks out approximately half of the remaining protons before the beam is again spread out through the second scattering foil. The intention is to produce a wide, homogeneous beam. The particles exit the first tube through a Kapton window which is used to keep the tube under vacuum to reduce random scattering off air molecules.

Dosimetry box

The aluminium box shown below contains the dosimetry equipment. The beam is first collimated using a brass collimator to remove any stray protons. The beam then travels through the dose monitors, the cross wires and finally exits the box through the brass nozzle.

The dose monitors are drift chambers used to measure the dose deposited by the proton beam. An exploded view of a dose monitor as used in the dosimetry box is shown below. It consists of a set of aluminised mylar foils wedged between layers of perspex to hold them in place. The guard ring is used to create a sealed volume of air between the foils. The aluminium layers face towards the centre of the dose monitor such that the assembly acts as a drift chamber when a potential difference is applied.


Visualisation

The geometry and particle tracks can be visualised using the DAWN visualiser [3]. Other, perhaps more suitable, visualisers are available (see [4]) but DAWN is useful in that it produces very high-resolution visualisations. The main drawback of DAWN is that it does not allow for click-and-drag functionality. Instead, parameters must be set before each visualisation is drawn which tends to require a fair amount of trial-and-error to obtain good images. All images on this page were created using DAWN and an example visualisation of the beamline with all particle tracks drawn can be seen below. The first tube and second box are visualised using the wireframe setting so that their inner components are visible. This example contains 500 primary protons. Protons are shown in blue, electrons are red, positrons are cyan, gamma rays are green, and neutrons are yellow.


Generation of primary protons

The protons are generated using the G4GeneralParticleSource (GPS) class which allows for a range of properties of the primary protons to be set in the proton.mac file (primaries are the initial particles generated by the GPS). First, the particle source is positioned against the wall on one side of the room which is at -4200 mm from the centre. In an attempt to achieve a more realistic beam, the primary protons are distributed normally in the x and y directions centred at 0 with standard deviations of 4.0 mm and 4.5 mm respectively. This gives the beam width and results in a nearly circular profile. The primaries are generated with initial energies following a Gaussian distribution with mean 62.5 MeV and standard deviation 0.082 MeV, and Gaussian radial distributions with respect to the z-axis with standard deviations of 2.3 mrad and 1.2 mrad in the x and y directions respectively.

The parameters are set in the following section of the proton.mac macro:

### particle gun settings
/gps/particle proton
/gps/number 1 			# 1 proton per event

# energy
/gps/ene/type Gauss
/gps/ene/mono 62.5 MeV
/gps/ene/sigma 0.082 MeV

# source position
/gps/position 0.0 0.0 -4200 mm

# beam width
/gps/pos/type Beam
/gps/pos/sigma_x 4.0 mm
/gps/pos/sigma_y 4.5 mm

# angular distribution of primaries for realistic emittance
/gps/ang/type beam2d
/gps/ang/sigma_x 2.3 mrad
/gps/ang/sigma_y 1.2 mrad
/gps/ang/rot1 -1 0 0		# aligns gps with positive z-axis

Gathering data

The energy of the beam is monitored at several stages in the beamline through the SteppingAction class. If the protons traverse a position in z defined in the vector trackingDistFiller in the SteppingAction header, their parent ID, xyz-coordinates, orientation with respect to the z-axis and energy are recorded at the given position in z. The "sensitive area" is restricted to a square of 100 mm in each direction in the xy-plane.

Running the simulation

Follow the instructions below to copy the simulation files to your directory:

[username@plus1 ~]$ mkdir Clatterbridge_sim
[username@plus1 ~]$ cd Clatterbridge_sim
[username@plus1 Clatterbridge_sim]$ /unix/pbt/software/scripts/pbt.sh
[username@plus1 Clatterbridge_sim]$ cp -r /unix/pbt/mhentz/Clatterbridge_sim/source .
[username@plus1 Clatterbridge_sim]$ mkdir build
[username@plus1 Clatterbridge_sim]$ cd build
[username@plus1 ProtonPB_build]$ make -DGeant4_DIR=/unix/pbt/software/dev /home/username/Clatterbridge_sim/source
[username@plus1 ProtonPB_build]$ make

Run macro proton.mac

This will run the simulation and produce the required output files.

[username@plus1 ProtonPB_build]$ ./protonPB proton.mac

Output files

The simulation code and proton.mac produce several output files:

protonKE.out

This text file contains output information from SteppingAction.cc , printed on a step-by-step basis for each proton (event). The first two columns contain the x and y coordinates (mm) of the proton. The third column contains the z position of the proton (mm), relative to the position of the source at the inner room boundary. The last column contains the energy (MeV) of the proton at this z position.

 6.66882  -11.4767  1770  59.8738
-7.60842  -11.4391  1770  60.186
-9.37841  -14.363   1770  43.1861
-13.2143   1.55943  1770  60.4776
 1.81337   11.558   1770  60.2989

secondariesKE.out

This text file contains information recorded in SteppingAction.cc about neutrons produced in the simulation. The values shown in the file are the kinetic energies [MeV] of the neutrons produced within the specified volume in the code.

5.12058
2.71641

sensitiveDetectors.out

This file contains the ID of the sensitive detector (first column) and the energy deposited in that detector (second column). The ID number "0" corresponds to the water volume sensitive detector, and ID number "1" corresponds to the user-defined sensitive detector.

1	62.2594
0	43.1061
1	0
0	0
1	62.2851

ProtonFluxBeamline.txt

This file is automatically written by proton.mac and contains the proton flux data from the longitudinal scorer. The columns represent the bin number in the x, y and z directions, and the number of protons per cm2:

# mesh name: waterMeshlongitudinal
# primitive scorer name: protonFlux
# iX, iY, iZ, value [percm2]
0,0,0,25.97935669321147
0,0,1,25.90015344712939
0,0,2,25.84461190407653

In simulation_analysis.C the data in ProtonFluxBeamline.txt is read in and written to another text file, ProtonFluxBeamline_Mod.txt to modify the standard formatting into a suitable format for further analysis:

0 0 0 25.97935669321147
0 0 1 25.90015344712939
0 0 2 25.84461190407653

A similar output file, NeutronFluxBeamline.txt is produced in the simulation, and contains information about the flux of neutrons along the entire length of the beamline. An additional two text files are produced from another longitudinal scoring mesh, located specifically along the length of the water volume. ProtonFluxWaterLongitudinal.txt contains information about the proton flux within the water volume, and ProtonEnergyWaterLongitudinal.txt contains information on the cumulative energy deposition of the proton beam within the water volume.

ProtonEnergyWaterLongitudinal.txt

This text file is produced by proton.mac in the same way as ProtonFluxBeamline.txt above, and contains information about the proton energy deposited within the water volume, divided into bins parallel to the beam direction:

# mesh name: waterMeshlongitudinal
# primitive scorer name: energyDeposit
# iX, iY, iZ, value [MeV]
0,0,0,14778.66421571247
0,0,1,14787.50686991539
0,0,2,14943.0041959171

EnergyLateralMesh.txt

This text file is produced by proton.mac in the same way as the other scorers above, and contains information about the proton energy deposition in bins perpendicular to the beam:

# mesh name: waterMeshlateral
# primitive scorer name: energyDeposit
# iX, iY, iZ, value [MeV]
0,0,0,0
1,0,0,0

...

16,0,0,0.002151585502782383
17,0,0,0.01565800625140965
18,0,0,9.09434626226223
19,0,0,0.01811169052807143

Data Analysis

Open ROOT and run analysis file

The simulation analysis file reads the data in the output files and produces the associated plots.

[username@plus1 ProtonPB_build]$ root -l

root [0] .x simulation_analysis.C

Proton energy deposition in water

On the left is a histogram showing the frequency of incident protons to the water box at different energy values. It is produced from data in stopper.txt which contains the hits data from the sensitive detectors in the simulation, including the water volume sensitive detector which is used here. The peak in proton energy deposited in the water volume is found to be at 60.08 MeV. The plot on the right shows the cumulative proton energy deposited within the water volume, plotting data collected in the longitudinal water volume scoring mesh, showing the Bragg peak at 31 mm.

Proton stopping distance in water

This plot shows the stopping distance of protons in the water volume, with the stopping distance defined as the point at which the velocity of the protons is zero.

Proton flux along beamline

This plot is produced from FluxLongitudinal_Mod.txt and shows the number of protons per cm2 passing through distinct points along the beamline (in the z direction). The decrease in proton flux is clearly shown at particular points along the beamline, for example at the location of the brass stopper (30cm).

Proton flux along water volume

Using a similar setup as the previous beamline flux plot, proton flux data from the scoring mesh within the water volume was plotted.

Kinetic energy of the proton beam

This left-hand plot is produced from kin.out and shows the kinetic energy distribution of the proton beam at a specified point along the beamline. This location is specified in SteppingAction.cc in which a particular position in z can be input (see Section 5.6 Kinetic Energy Readings).

The xy coordinates of each proton are also recorded, and are used to produce a plot which shows the spatial coordinates and the kinetic energy of each proton at the specified position in z, along the beamline.

The z position along the beamline chosen to produce the following plots was just after the nozzle.

Neutron production

Neutron flux along beamline

This was plotted from data contained in the file NeutronFluxBeamline.txt, and shows the increase in flux of neutrons at certain points (locations of beamline components) along the beamline.

Neutron kinetic energy

Data on the energy of neutrons produced within a particular volume along the beamline was collected, with the volume specified in SteppingAction.cc - the plot below shows the energy of neutrons produced within the water volume.

Changing parameters

Initial beam parameters

Initial parameters of the proton beam can be modified in proton.mac

Beam radius

/gps/pos/radius 3 mm

Beam energy

This simulation models the proton beam source with a Gaussian distribution.

/gps/ene/type Gauss
/gps/ene/mono 62.5 MeV
/gps/ene/sigma 0.082 MeV

Source position

The proton source is positioned at z = -420 cm relative to the centre of the inner room (the mother volume), which translates as the wall surface of the inner room.

/gps/pos/type Plane
/gps/pos/shape Circle
/gps/pos/centre 0.0 0.0 -420 cm

Scoring mesh

Longitudinal scoring mesh

A longitudinal scoring mesh extends along the length of the beamline from the source to the water volume. The mesh utilises a filter to detect the flux of protons per cm2 and writes the data to the text file FluxLongitudinal.txt . The location of the mesh centre can be changed in proton.mac , in addition to the dimensions of the mesh and the number of bins.

/score/create/boxMesh waterMeshlongitudinal
/score/mesh/boxSize 10. 10. 10. cm
/score/mesh/nBin 1 1 400
/score/mesh/translate/xyz 0. 0. -226 cm

The filter can also be changed to observe the flux of particles other than protons:

/score/quantity/cellFlux protonFlux
/score/filter/particle protonFilter proton

Lateral scoring mesh

A lateral scoring mesh is positioned at the end of the nozzle to record the energy distribution of the protons perpendicular to the direction of the beam. The position, size and bin number of this mesh can be modified in the same way as the longitudinal mesh example above.

Beamline components

Components of the beamline can be added/removed in DetectorConstruction.cc . If the dimensions of the water box are modified, the following lines in simulation_analysis.C will also need to be modified:

Float_t lengthBox = 200, widthBox = 200;

The depthFix variable to calculate the stopping distance of the protons within the analysis will also need to be adjusted if the water box dimensions or location are modified. depthFix is calculated by taking the z position of the centre of the water box (relative to the centre of the inner room), and subtracting the half length of the water box (calculated in mm):

//depth fix - water box centred at -2260 mm, half length = 100 mm. 
Double_t depthFix = 2360;

Sensitive detectors

In this simulation, the water volume is assigned as a sensitive detector in DetectorConstruction.cc :

G4SDManager* SDman = G4SDManager::GetSDMpointer();
G4String name="SD";
DetectorSD = new SensitiveDetector(name);
SDman->AddNewDetector(DetectorSD);
logicWater->SetSensitiveDetector(DetectorSD);

Another beamline component may be used by modifying the following line and setting its logical volume as a sensitive detector:

logicWater->SetSensitiveDetector(DetectorSD);

The component should be chosen such that a significant proportion of the proton beam deposits energy, such as in the brass stopper, in order to produce enough data for plots.

SensitiveDetector.cc is derived from the G4VSensitiveDetector.cc base class. On a step-by-step basis, the energy deposited by the proton is recorded as a "hit" and added to a HitsCollection object. Other parameters may be retrieved at each step in the method ProcessHits .

G4bool SensitiveDetector::ProcessHits(G4Step* aStep, G4TouchableHistory* ROhist)
{
  G4double edep = aStep->GetTotalEnergyDeposit();
  if(aStep->GetTrack()->GetDefinition()->GetParticleName() == "proton"){
    ::Hit* newHit = new ::Hit();
    newHit->SetEdep(edep);
    HitID = detectorCollection->insert(newHit);
    return true;
  }

Physics list

This simulation uses one of the standard physics lists: QGSP_BIC_HP. A user-defined physics list can be specified in the class PhysicsList.cc , and then selected in protonPB.cc

runManager->SetUserInitialization(new  PhysicsList);

Kinetic energy readings

The kinetic energy of the beam after passing through a particular component of the beamline is found in the class SteppingAction.cc . This class contains a map named componentMap with the names of the components as keys and the z coordinate of each component (taken as the furthest edge of the component from the source):

componentMap.insert( std::pair<std::string,double>("ScatteringFoil1",80.025));
componentMap.insert( std::pair<std::string,double>("BrassStopper",306.6));
componentMap.insert( std::pair<std::string,double>("ScatteringFoil2",306.625));

To find the kinetic energy distribution of the beam at a particular point, insert the component name from the map above into this line:

double cPos = (double)componentMap.find("ScatteringFoil2")->second;

The code then compares the z position of a proton with the position of the chosen component on a step-by-step basis, and if the proton is at the chosen position its kinetic energy is recorded in kin.out :

if(fabs(zpos-cPos) < 1.0e-8){
   std::ofstream hitsfile(filename, std::ios::app);
   if(hitsfile.is_open()){
      G4double ke =  step->GetPreStepPoint()->GetKineticEnergy();
      hitsfile << zpos << "\t"
	       << ke <<G4endl;
      hitsfile.close();
   }
}


After any modifications to the simulation files, the code will need to be compiled. In the build directory, write:

[username@plus1 ProtonPB_build]$ make 

After this proton.mac can be run:

[username@plus1 ProtonPB_build]$ ./protonPB proton.mac

Modifying Analysis Methods

The number of bins to store the data for each histogram can be changed, here the number of bins is 800:

TH1D* waterEnergy = new TH1D ("waterDep", "Energy deposition in water box", 800, 0., 65.);

The range of the axes for each plot can also be specified in order to zoom into the energy peak:

waterEnergy->GetXaxis()->SetRangeUser(0, 64);


Files

List of files for Clatterbridge simulation.