Proton Calorimetry/Experimental Runs/2021/Apr15
Afternoon test at UCLH with photodiodes and CMOS sesnor.
Experiment Equipment
Item | Notes |
---|---|
Network Hub | Set in control room to take output from experimental room ethernet connection. Control laptops connected via 5GHz WiFi. |
Control Laptop x3 | 2x remote control for sensor and FPGA, 1 for notes/web GUI. |
Ethernet Cable x 3 | To connect DAQ laptops to network in the experimental room, network hub to network in control room. |
Portable Enclosure | Modified Big Zarges Waterproof Wheeled Equipment Case.
Features mount for scintillator stack, front and back openings for beam, patch panel with ports for SHV, BNC, SMA, USB, Camera Link cables. |
Scintillator stack | X x 2 mm, Y x 2.6 mm and Z x 3 mm sheets in ascending order from the front/beam of the end of the scintillator. Sheet numbers from front to end (in beam direction): 25, 31, 50, 52, 34, 44, 53, 47, 33, 54, 46, 42, 55, 45, 38, 36, 43, 32, 48, 41, 8, 39, 40, 37, 35, 49, 9, 2, 1, 51, 11, 4, 3, 23, 6. Full stack thickness measurements (in mm): 98.10, 98.08, 97.93, 98.49. |
ISDI CMOS sensor | sensor pixel dimension: 1030 x 1536. NO optical grease between scintillator and sensor. Connected to DAQ PC via Camera Link cable. |
Nikon D7500 | Connected to DAQ Laptop via USB cable. |
DAQ laptop x2 | Control sensor and photodiode acquisition. |
Nexys Video FPGA development board. | For interfacing between DDC232 and PC. |
Rev. 2 Texas Instruments DDC232 custom circuit board (x2) | Each housing 16x Hamamatsu S12915-16R photodiodes, coupled directly to scintillator sheets. No optical gel and aligned by eye to scintillator sheets. Photodiodes do not cover sheets 25, 23 and 6. |
Gloves | For handling scintillator |
Schematic
ISDI Sensor Instructions
General Sensor Configuration
- Base Configuration is 14 x 1 Bit
- Resolution is 1030 x 1536 pixels
- For selecting pixel rows in centre: Use horizontal offset (ignore DVAL to activate field) and horizontal resolution
Increase number of frame buffers
- To increase the number of frame buffers to the allowed maximum, start XCAP as administrator.
- Select Capture/Sequence Capture/Video to Frame Buffers/Driver Assistant
- Tick box in left field, then set number of frame buffers here.
Serial Terminal instructions for ISDI sensor
- Find serial terminal in PIXCI(R) tab of sensor control software.
- In Controls/Options: Make sure "Send string with CR" is selected. Should be default.
- In Controls/Setup: Select Serial Port: Enabled. Baud rate has to be 115200
- Set low full well mode using W000300000
- Set high full well mode using W000300004
- Read full well mode using R0003
Continuous Field buffer setting
- This setting makes sure that the buffer is filled with a new frame as they come in. If this option is not selected the sensor and software will not be synchronised and we *will record a lot of pitch black images.
- Select Capture/Live Options//Live Mode/Live Video: Continuous All buffer capture
DDC232 Instructions
- Bash script
prog.sh
executesFTDI.cpp
andlivePlot.cpp
. Execute with. prog.sh 2 350 170 0 50 true Data/capture.txt
in the directoryPhotodiodes
.- First argument is the number of DDC232s.
- Second argument is the full-scale range (FSR) in pC. Choices are: 12.5, 50, 100, 150, 200, 250, 300, 350.
- Third argument is the integration time in us. Range is 170-1455. Set in multiples of 5 us.
- Fourth argument is the number of measurements to acquire. Typically acquire 5700 measurements per second. Set to 0 for infinite loop.
- Fifth argument is the rate in Hz to update the file
average.csv
for ROOT/web GUI plotting. - Sixth argument is a boolean to choose whether to display the ROOT plot.
- Seventh argument is the filename and destination for the output hexadecimal data.
- After acquisition, close ROOT by choosing File->Quit in the ROOT window and then stopping DAQ (if in infinite loop) with CTRL-C.
Experiment Plan
- Shoot-through measurements (back & front) for light calibration and level setting.
- Range measurements between 70-120 MeV with 1mm steps with CMOS sensor and photodiodes.
- High intensity range measurements between 70-120 MeV with 1mm steps with photodiodes at optimal light level.
- Live measurement of 3cm x 3cm x 3cm treatment plan.
Experiment Notes
Range calorimeter measurements with ISDI sensor + photodiodes.
Images from sensor: /unix/pbt/data/uclh/20210415/CMOS
Photodiode hexadecimal data: /unix/pbt/data/uclh/20210415/DDC232
Best photodiode QB fits for each energy can be found in this document.
Best CMOS QB fits for each energy can be found in this document.
Best PDLs for CMOS and photodiode for each energy can be found in this document.
Run number | Full Well Mode | Beam Energy (MeV) | Estimated Range (mm) | Current (nA) | Spot size (mm) | DDC232 FSR (pC) | DDC232 Integration Time (us) | Degrader WET (mm) | Comments | PDL Comparison | CMOS QB Fit | PD QB Fit |
---|---|---|---|---|---|---|---|---|---|---|---|---|
0 | high | Background | N/A | N/A | N/A | 350 | 170 | 0 | room light on | N/A | N/A | N/A |
1 | high | Background | N/A | N/A | N/A | 12.5 | 170 | 0 | repeat – room light on | N/A | N/A | N/A |
2 | high | Background | N/A | N/A | N/A | 12.5 | 170 | 0 | room lights off | N/A | N/A | N/A |
3 | high | 245 | 366.3 | 10 | 7.1 | 12.5 | 170 | 57.33 | shoot-through, gantry angle 270° | N/A | N/A | N/A |
4 | high | 245 | 366.3 | 20 | 7.1 | 12.5 | 170 | 57.33 | shoot-through, gantry angle 270°, CMOS sensor saturated | N/A | N/A | N/A |
5 | high | 245 | 366.3 | 10 | 7.1 | 12.5 | 170 | 57.33 | shoot-through, gantry angle 90° | N/A | N/A | N/A |
6 | high | 245 | 366.3 | 10 | 7.1 | 12.5 | 170 | 57.33 | shoot-through, gantry angle 90° | N/A | N/A | N/A |
7 | high | 105 | 300 | 7.1 | 12.5 | 170 | 0 | CMOS sensor saturated, BP seen | [1] | [2] | ||
8 | high | 105 | 150 | 7.1 | 12.5 | 170 | 0 | CMOS sensor saturated, BP seen | [4] | [5] | ||
9 | high | 70 | 40.8 | 700 | 7.1 | 12.5 | 170 | 0 | CMOS sensor saturated, BP seen | [7] | [8] | |
10 | high | 70 | 40.8 | 350 | 7.1 | 12.5 | 170 | 0 | [10] | [11] | ||
11 | high | 80 | 51.8 | 500 | 7.1 | 12.5 | 170 | 0 | Couple 1000 off saturation | [13] | [14] | |
12 | high | 90 | 63.9 | 400 | 7.1 | 12.5 | 170 | 0 | Saturation | [16] | [17] | |
13 | high | 100 | 77.1 | 350 | 7.1 | 12.5 | 170 | 0 | [19] | [20] | ||
14 | high | 110 | 91.3 | 300 | 7.1 | 12.5 | 170 | 0 | Saturation | [22] | [23] | |
15 | high | 120 | 106.5 | 270 | 7.1 | 12.5 | 170 | 0 | BP not captured in range | N/A | N/A | N/A |
16 | high | 110 | 91.3 | 150 | 7.1 | 12.5 | 170 | 0 | [25] | [26] | ||
17 | high | 105 | 350 | 7.1 | 12.5 | 170 | 0 | Saturated | [28] | [29] | ||
18 | high | 100 | 77.1 | 200 | 7.1 | 12.5 | 170 | 0 | Saturated | [31] | [32] | |
19 | high | 100 | 77.1 | 350 | 7.1 | 12.5 | 1000 | 0 | Saturated | [34] | [35] | |
20 | high | 95 | 400 | 7.1 | 12.5 | 170 | 0 | Saturated | [37] | [38] | ||
21 | high | 90 | 63.9 | 200 | 7.1 | 12.5 | 170 | 0 | [40] | [41] | ||
22 | high | 90 | 63.9 | 400 | 7.1 | 12.5 | 1000 | 0 | [43] | [44] | ||
23 | high | 85 | 500 | 7.1 | 12.5 | 170 | 0 | just missed saturation | [46] | [47] | ||
24 | high | 80 | 51.8 | 250 | 7.1 | 12.5 | 170 | 0 | [49] | [50] | ||
25 | high | 80 | 51.8 | 500 | 7.1 | 12.5 | 1000 | 0 | [52] | [53] | ||
26 | high | 75 | 600 | 7.1 | 12.5 | 170 | 0 | [55] | [56] | |||
27 | high | 70 | 40.8 | 350 | 7.1 | 12.5 | 170 | 0 | [58] | [59] | ||
28 | high | 70 | 40.8 | 700 | 7.1 | 12.5 | 1000 | 0 | [61] | [62] | ||
29 | N/A | Treatment Plan | N/A | N/A | N/A | 12.5 | 170 | 0 | clinical mode from now - scanned field | N/A | N/A | N/A |
30 | N/A | Treatment Plan | N/A | N/A | N/A | 12.5 | 170 | 0 | repeat | N/A | N/A | N/A |
31 | N/A | Treatment Plan | N/A | N/A | N/A | 50 | 170 | 0 | N/A | N/A | N/A |