Proton Calorimetry/Experimental Runs/2018/Mar22

From PBTWiki
Jump to navigation Jump to search

Beam tests of the UCL plastic scintillator range calorimeter.

Night of Thursday 22th to Friday 23rd of March 2018

Pre-data Taking Information and Notes

  • Data was acquired with the DSLR camera and the ISDI CMOS sensor.
  • The average background value per pixel for the sensor is ~ 2500. The sensor pixel dimensions are 1030 x 1536. The sensor frame rate was always set to 25 frames per second.
  • NO optical grease between scintillator and sensor.
  • Scintillator stack arrangement: 15 x 2 mm, 15 x 2.6 mm and 1 x 3 mm sheets used: In ascending order from the back of the scintillator to the front/beam. Except for sheet number 21 (2mm) which is placed at the end (away end from the beam) of the stack because of a crack in the sheet.
  • Folder for DSLR images: digiCamControl/20180322_medaustron
  • Note: When powering on PC, sensor has to be connected, but unpowered!
  • 16383 = saturation pixel number (16 bit x 1 tap)
  • Sensor images are always saved as a sequence of 21 Tiff images
  • Folder for sensor images: XCAP/data/20180322_medaustron


Data Taking

DSLR Sensor Beam Info Thumbnails
Run number Aperture Exposure (s) Full Well Mode Intensity (%) Number of particles/s Beam Energy (MeV) Test type Comments DSLR Sensor
1 f/5 1s Low N/A Background Background run with ND filter

DSC_0001.jpg

N/A
2 f/5 1s Low 100% 1.2 x 1e10/5s 80.5 Bragg peak ND filter in - not enough light passes through filter. Remove! DSLR slightly saturated. 10 spills used for this run.
3 f/8 1s High N/A Background No ND filter!
4 f/8 1s High 100% 1.2 x 1e10/5s 80.5 Bragg peak
5 f/7.1 1s N/A 100% 1.2 x 1e10/5s ? Bragg peak
6 f/7.1 1s N/A 100% 1.2 x 1e10/5s ? Bragg peak
7 f/7.1 1s N/A 100% 1.2 x 1e10/5s ? Bragg peak
8 f/5 1s High 50% 0.6 x 1e10/5s ? Bragg peak
9 f/5.6 1s N/A 50% 0.6 x 1e10/5s ? Bragg peak
10 f/5.6 1s High 50% 0.6 x 1e10/5s 85.6 Bragg peak
11 f/5.6 1s High 50% 0.6 x 1e10/5s 85.6 Bragg peak
12 f/5.6 1s High 50% 0.6 x 1e10/5s 62.4 Bragg peak
13 f/5.6 1s High 50% 0.6 x 1e10/5s 62.4 Bragg peak
14 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan
15 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan
16 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan
17 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan
18 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan
19 f/5.6 1s High 50% 0.6 x 1e10/5s Bragg peak
20 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan
21 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan
22 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan
23 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan
24 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan
25 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan
26 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan
27 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan
28 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan
29 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan
30 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Bragg peak
31 f/5.6 1s High 50% 0.6 x 1e10/5s 101.8 Energy Scan
32 f/5.6 1s High 50% 0.6 x 1e10/5s 99.6 Energy Scan
33 f/5.6 1s High 50% 0.6 x 1e10/5s 97.4 Energy Scan
34 f/5.6 1s High 50% 0.6 x 1e10/5s 95.1 Energy Scan
35 f/5.6 1s High 50% 0.6 x 1e10/5s 92.8 Energy Scan
36 f/5.6 1s High 50% 0.6 x 1e10/5s 90.4 Energy Scan
37 f/5.6 1s High 50% 0.6 x 1e10/5s 86.4 Energy Scan
38 f/5.6 1s High 50% 0.6 x 1e10/5s 85.6 Energy Scan
39 f/5.6 1s High 50% 0.6 x 1e10/5s 84.7 Energy Scan
40 f/5.6 1s High 50% 0.6 x 1e10/5s 83.9 Energy Scan
41 f/5.6 1s High 50% 0.6 x 1e10/5s 83.0 Energy Scan
42 f/5.6 1s High 50% 0.6 x 1e10/5s 82.3 Energy Scan
43 f/5.6 1s High 50% 0.6 x 1e10/5s 81.3 Energy Scan
44 f/5.6 1s High 50% 0.6 x 1e10/5s 80.5 Energy Scan
45 f/5.6 1s High 50% 0.6 x 1e10/5s 79.6 Energy Scan
46 N/A High 50% 0.6 x 1e10/5s 78.7 Energy Scan
47 f/5.6 1s High 20% 2.4 x 1e9/5s 78.7 Energy Scan
48 f/5.6 1s High 20% 2.4 x 1e9/5s 77.8 Energy Scan
49 f/5 1s High 20% 2.4 x 1e9/5s 76.9 Energy Scan
50 f/5 1s High 20% 2.4 x 1e9/5s 76.0 Energy Scan
51 f/5 1s High 20% 2.4 x 1e9/5s 75.1 Energy Scan
52 f/5 1s High 20% 2.4 x 1e9/5s 74.2 Energy Scan
53 f/5 1s High 20% 2.4 x 1e9/5s 72.4 Energy Scan
54 f/5 1s High 20% 2.4 x 1e9/5s 70.5 Energy Scan
55 f/5 1s High 20% 2.4 x 1e9/5s 68.5 Energy Scan
56 f/5 1s High 20% 2.4 x 1e9/5s 66.5 Energy Scan
57 f/5 1s High 20% 2.4 x 1e9/5s 64.5 Energy Scan
58 f/5 1s High 20% 2.4 x 1e9/5s 62.4 Energy Scan
59 f/7.1 1s High 100% 1.2 x 1e10/5s 85.6 Intensity Scan
60 f/5.6 1s High 50% 0.6 x 1e10/5s 85.6 Intensity Scank
61 f/4.5 1s High 20% 2.4 x 1e9/5s 85.6 Intensity Scan
62 f/3.2 1s High 10% 1.2 x 1e9/5s 85.6 Intensity Scan
63 f/3.2 1s Low 10% 1.2 x 1e9/5s 85.6 Intensity Scan
64 f/3.2 5s High 2.22% 2.66 x 1e8/5s 85.6 Intensity Scan
65 f/3.2 5s Low 2.22% 2.66 x 1e8/5s 85.6 Intensity Scan
66 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan
67 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan
68 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan
69 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan
70 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan
71 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan
72 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan
73 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan
74 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan
75 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan
76 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan
77 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan
78 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan
79 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan
80 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan
81 f/3.2 1s High 50% 0.6 x 1e10/5s 99.6 Radiation damage assessment
82 f/3.2 1s High 50% 0.6 x 1e10/5s 99.6 Radiation damage assessment
83 f/3.2 1s High 50% 0.6 x 1e10/5s 99.6 Radiation damage assessment



End of range telescope tests.