Line 15: | Line 15: | ||
**There is some interplay with Birks' constant and range – worse values of Birks' constant give worse range results. | **There is some interplay with Birks' constant and range – worse values of Birks' constant give worse range results. | ||
**Possibly have 0.8mm systematic offset from reference data, but need more reference data to confirm. | **Possibly have 0.8mm systematic offset from reference data, but need more reference data to confirm. | ||
*Programmed flash memory on USB104 board so that device is programmed with DDC232 code on power-up. | |||
**Updated Raspberry Pi and loaded latest versions of DAQ code: set up works with Sonia's GUI. | |||
*Organised first meeting of UCL HEP Climate Change projects and drafted a call for proposals for 4 small funded projects for PhD students. | *Organised first meeting of UCL HEP Climate Change projects and drafted a call for proposals for 4 small funded projects for PhD students. |
Revision as of 09:57, 16 March 2023
Minutes for UCL Proton Calorimetry Meetings, 15th March
Present
Saad Shaikh, Simon Jolly, Sonia Escribano, Maya Maciurzynska
Saad Shaikh
- Delivered update on Geant4 simulation development, slides can be found here.
- Found a list of values for Gaussian energy straggling to produce curves similar to those measured at UCLH, across clinical energy range.
- Added detector messenger to allow Birks' constant to be set at runtime.
- Gave presentation on analysis on PARTREC beam test, slides can be found here.
- Beam quality worsens at lower energies, possibly due to aluminium beam degrader.
- Appear to have linear scaling of average charge measured with beam current.
- Fit results do vary with different beam current: effect is more pronounced at lower (worse) energies, best results still have 0.1-0.2 mm difference in range.
- There is some interplay with Birks' constant and range – worse values of Birks' constant give worse range results.
- Possibly have 0.8mm systematic offset from reference data, but need more reference data to confirm.
- Programmed flash memory on USB104 board so that device is programmed with DDC232 code on power-up.
- Updated Raspberry Pi and loaded latest versions of DAQ code: set up works with Sonia's GUI.
- Organised first meeting of UCL HEP Climate Change projects and drafted a call for proposals for 4 small funded projects for PhD students.