Proton Calorimetry/Experimental Runs/2022/ManchesterFLASH: Difference between revisions

From PBTWiki
Jump to navigation Jump to search
Line 116: Line 116:
|7 || Full Stack  || Background || || - || - ||  12.5 || 170 ||  ||  || N/A ||  
|7 || Full Stack  || Background || || - || - ||  12.5 || 170 ||  ||  || N/A ||  
|-
|-
|8 || Full Stack  || 245 MeV || - || || N/A || 12.5 || 170 ||  ||  || N/A ||  
|8 || Full Stack  || 245 MeV || - || 18.8 || N/A || 12.5 || 170 ||  ||  || N/A ||  
|-
|-
|9 || Full Stack  || 245 MeV || || 800 || 4.0 ||  350 || 170 ||  ||    || N/A ||  
|9 || Full Stack  || 245 MeV || - || 800 || N/A ||  350 || 170 ||  ||    || N/A ||  
|-
|-
|10 || Full Stack  || Background || || - || N/A ||  350 || 170 ||  ||  || N/A ||  
|10 || Full Stack  || Background || - || - || N/A ||  350 || 170 ||  ||  || N/A ||  
|-
|-
|11 || Full Stack || 245 MeV || || 800 || 4.0 || 350 || 170 ||  ||  || N/A ||  
|11 || Full Stack || 245 MeV || - || 800 || N/A || 350 || 170 ||  ||  || N/A ||  
|-
|-
|12 || Full Stack  || 245 MeV || || 800 || 4.0 ||  350 || 170 ||  ||  || N/A ||  
|12 || Full Stack  || 245 MeV || - || 800 || N/A ||  350 || 170 ||  ||  || N/A ||  
|-
|-
|13 || Full Stack  ||  245 MeV || || 800 || 4.0 ||  350 || 170 ||  ||    || N/A ||
|13 || Full Stack  ||  245 MeV || - || 800 || N/A ||  350 || 170 ||  ||    || N/A ||
|-
|-
|14 || Full Stack  ||  245 MeV || || 800 || 4.0 ||  350 || 170 ||  ||  || N/A ||
|14 || Full Stack  ||  245 MeV || - || 800 || N/A ||  350 || 170 ||  ||  || N/A ||
|-
|-
|}
|}

Revision as of 21:22, 26 October 2022

Experiment Equipment

Item Notes
Network Hub Set in control room to take output from experimental room ethernet connection. Control laptops connected via 5GHz WiFi.
Control Laptop x2 1 for remote control of FPGA, 1 for notes/web GUI.
Ethernet Cable x 2 To connect DAQ laptop to network in the experimental room, control laptop to network in control room.
Portable Enclosure Modified Large Peli Case Waterproof Wheeled Equipment Case. Features mount for scintillator stacks, front and back openings for beam, patch panel with ports for USB and power connections. Mylar windows with 3D printed alignment plates: approximately light-tight.
Rev. D Scintillator stack 1 32 X x 105 mm, Y x 105 mm and Z x (approx) 3 mm sheets. Original injection moulded batch with polished edges and sanded faces. Containing sheets (ascending photodiode order, board upside down): 137, 138, 139, 140, 141, 142, 143, 144, 145, 146, 147, 148, 149, 150, 151, 152, 153, 154, 155, 156, 157, 158, 159, 160, 161, 162, 163, 164, 165, 166, 167, 168. Full stack thickness measurements (in mm):
Rev. D Scintillator stack 2 32 X x 105 mm, Y x 105 mm and Z x (approx) 3 mm sheets. Machined block with polished edges and sanded faces. Containing sheets (ascending photodiode order, board upside down): 289, 268, 291, 277, 300, 264, 282, 276, 271, 285, 262, 280, 295, 281, 297, 290, 270, 284, 294, 299, 269, 302, 301, 296, 287, 263, 288, 293, 285, 265, 275, 292. Full stack thickness measurements (in mm):
Rev. D Scintillator stack 3 32 X x 105 mm, Y x 105 mm and Z x (approx) 3 mm sheets. Injection moulded with polished edges and sanded faces. Containing sheets (ascending photodiode order, board upside down): 232, 208, 242, 206, 243, 223, 254, 210, 229, 217, 235, 255, 246, 230, 213, 241, 231, 257, 256, 236, 258, 215, 219, 249, 247, 260, 251, 259, 226, 239, 237, 245 Full stack thickness measurements (in mm):
Rev. D Scintillator stack 4 32 X x 105 mm, Y x 105 mm and Z x (approx) 3 mm sheets. Mix of injection moulded and machined block with polished edges and sanded faces. Containing sheets (ascending photodiode order, board upside down): 228, 220, 225, 227, 218, 214, 238, 248, 209, 253, 224, 222, 211, 261, 250, 234, 212, 233, 252, 240, 221, 207, 244, 283, 272, 279, 298, 274, 273, 267, 278, 266. Full stack thickness measurements (in mm):
DAQ laptop x1 Control photodiode acquisition.
Nexys Video FPGA development board. For interfacing between DDC232 and PC.
Rev. D Texas Instruments DDC232 custom circuit board (x4) Housing 32x Hamamatsu S12915-16R photodiodes, coupled directly to scintillator sheets. Aligned by eye to scintillator sheets.
Gloves For handling scintillator

Schematic

UCL_Setup.png

Experiment Plan

Day 1 (25/10/22) – estimated first beam: 23.00.

  1. Set up mains cabling and networking in treatment and control rooms.
  2. Install Peli case on treatment couch.
  3. Calibration shoot-through measurements of 4 rev. D modules: either rotating case or modules for back shoot-through.
  4. Pristine Bragg peak measurements at clinical current for full clinical range.
  5. FLASH measurements at 245 MeV.

Day 2 (26/10/22) – estimated first beam: 23.00.

  1. Set up PeakFinder
  2. WET Measurements for each of the 4 rev. D modules.
  3. Single scintillator/mylar foil removal experiment with Peakfinder.

Experiment Notes

Range calorimeter measurements with photodiodes and PTW Peakfinder.

Photodiode hexadecimal data: /unix/pbt/data/manchester/20221025/DDC232

Peakfinder data: /unix/pbt/data/manchester/20220901/Peakfinder

PTW PeakFinder Range Pull Back Measurements

Run number Detector config Beam Energy (MeV) Beam Current (nA) Physical thickness (mm) D80 measured (mm) Comments
0 Clear stack + white stack 210 40 N/A 139.1 Nominal: 282.1 mm
1 Clear stack 210 40 N/A 206.6 Nominal: 282.1 mm

DDC232 Calibration and Bragg Peak Measurements 26/10/2022

Run number Detector Beam Energy (MeV) Estimated Range (mm) Current (nA) Spot size (mm, FWHM) DDC232 FSR (pC) DDC232 Integration Time (us) Degrader WET (mm) Comments Photodiode QB Fit Replay Fit
0 Full Stack (1-2-3-4) Background - - N/A 350 170 (constant) N/A
1 Full Stack Background - - 300 170 N/A
2 Full Stack Background - - 250 170 N/A
3 Full Stack Background - - 200 170 N/A
4 Full Stack Background - - 150 170 N/A
5 Full Stack Background - - 100 170 N/A
6 Full Stack Background - - 50 170 N/A
7 Full Stack Background - - 12.5 170 N/A
8 Full Stack 245 MeV - 18.8 N/A 12.5 170 N/A
9 Full Stack 245 MeV - 800 N/A 350 170 N/A
10 Full Stack Background - - N/A 350 170 N/A
11 Full Stack 245 MeV - 800 N/A 350 170 N/A
12 Full Stack 245 MeV - 800 N/A 350 170 N/A
13 Full Stack 245 MeV - 800 N/A 350 170 N/A
14 Full Stack 245 MeV - 800 N/A 350 170 N/A

DDC232 Calibration and Bragg Peak Measurements 25/10/2022

Run number Detector Beam Energy (MeV) Estimated Range (mm) Current (nA) Spot size (mm, FWHM) DDC232 FSR (pC) DDC232 Integration Time (us) Degrader WET (mm) Comments Photodiode QB Fit Replay Fit
0 Stack 1 Background - - N/A 12.5 170 N/A
1 Stack 1 245 MeV 800 4.0 12.5 170 Shoot through front (saturated) N/A
2 Stack 1 245 MeV 800 4.0 50 170 Shoot through front (saturated) N/A
3 Stack 1 245 MeV 800 4.0 350 170 Shoot through front N/A
4 Stack 2 Background - N/A 350 170 N/A
5 Stack 2 245 MeV 800 4.0 350 170 Shoot through front N/A
6 Stack 3 Background - N/A 350 170 N/A
7 Stack 3 245 MeV 800 4.0 350 170 Shoot through front N/A
8 Stack 4 Background - N/A 350 170 N/A
9 Stack 4 245 MeV 800 4.0 350 170 Shoot through front N/A
10 Stack 1 Background - N/A 350 170 N/A
11 Stack 1 245 MeV 800 4.0 350 170 Shoot through back N/A
12 Stack 2 245 MeV 800 4.0 350 170 Shoot through back N/A
13 Stack 3 245 MeV 800 4.0 350 170 Shoot through back N/A
14 Stack 4 245 MeV 800 4.0 350 170 Shoot through back N/A