== Data Taking == | == Data Taking == | ||
* All images taken with | * All images taken with | ||
* Used camera | ** Canon EOS 1100D | ||
** 18-55mm lens | |||
** f/5 | |||
** 200ISO | |||
** 30s exposure | |||
* Beam turned on for a given amount of time within the 30s exposure | |||
* This is noted in monitor units, with the beam having constant 1000MU/minute | |||
* Used EOS utility app for controlling camera | |||
* Image of ruler taken at sheet for pixels to mm conversion | * Image of ruler taken at sheet for pixels to mm conversion | ||
* | * 3 images taken at each collimator size with 1 background image each. | ||
<!-- ------------------------------------table3------------------------------------------ --> | <!-- ------------------------------------table3------------------------------------------ --> |
Revision as of 14:51, 24 August 2022
Second experimental measurements with Cyberknife at Birmingham. Green scintillator thin sheet measurements only.
Setup
Schematic
Photos of setup
(Inside pelicase, outside pelicase, collimators, through pelicase window)
Pre-data Taking Information and Notes
- Used 200mmx200mmx2mm green scintillator sheet polystyrene base sandwiched by two 200mmx200mmx50mm polystyrene clear plastic blocks
- 53cm from lens to sheet
- Check camera in focus with ruler on the sheet
- Near/Far limit of focus 49.4cm 57.1cm calculated using https://dofmaster.com/dofjs.html
- Pelicase on couch
- Scintillator blocks aligned to laser cross hairs
- 785 mm source to surface distance suitable for output factor/PDD/profile measurements.
- Room lights out and pelicase window open.
- Check optical axis aligned with top surface using crosshair in the image analysis.
- Arrived at hospital at 11am, finished data taking at 5pm
Data Taking
- All images taken with
- Canon EOS 1100D
- 18-55mm lens
- f/5
- 200ISO
- 30s exposure
- Beam turned on for a given amount of time within the 30s exposure
- This is noted in monitor units, with the beam having constant 1000MU/minute
- Used EOS utility app for controlling camera
- Image of ruler taken at sheet for pixels to mm conversion
- 3 images taken at each collimator size with 1 background image each.
With scintillator sheet | |||
---|---|---|---|
Collimator Size (mm) | Monitor Units | Image | Comments |
5 | 400 | ||
7.5 | 400 | ||
10 | 400 | ||
12.5 | 400 | ||
15 | 400 | ||
20 | 400 | ||
25 | 300 | ||
30 | 300 | ||
35 | 250 | ||
40 | 250 | ||
50 | 150 | ||
60 | 150 |
No scintillator sheet | |||
---|---|---|---|
Collimator Size (mm) | Monitor Units | Image | Comments |
5 | 400 | ||
7.5 | 400 | ||
10 | 400 | ||
12.5 | 400 | ||
15 | 400 | ||
20 | 400 | ||
25 | 300 | ||
30 | 300 | ||
35 | 250 | ||
40 | 250 | ||
50 | 150 | ||
60 | 150 |