Proton Calorimetry/Experimental Runs/2018/Mar23: Difference between revisions

From PBTWiki
Jump to navigation Jump to search
Line 26: Line 26:
!PMT Current, uA
!PMT Current, uA
!Beam Intensity
!Beam Intensity
!Rate (Measured on Digitiser)
!Rate (Measured on Digitiser) and Notes
!Digitiser Settings
!Digitiser Settings
|-
|-
Line 34: Line 34:


| 001 || -900 || 150.1 || Lowest || 0.3 - 5 kHz || DC Offset: -30, Threshold: 40 LSB, Gate Offset: 20 ns, Long Gate: 150 ns  
| 001 || -900 || 150.1 || Lowest || 0.3 - 5 kHz || DC Offset: -30, Threshold: 40 LSB, Gate Offset: 20 ns, Long Gate: 150 ns  
|-
| 002 || -800 || 133.2 || Lowest || 0.3 - 5 kHz || As above
|-
| 003 || -700 || 116.35 || Lowest || 0.3 - 5 kHz || As above
|-
| 004 || -600 || 99.45 || Lowest || 0.3 - 5 kHz || Beam dropped during running!
|-
| 005 || -600 || 99.45 || Lowest || 0.3 - 5 kHz || Repeat run with beam back.
|-
| 006 || -500 || 82.6 || Lowest || 0.3 - 5 kHz || Threshold: 10 LSB. Too close to the threshold, but cannot lower more due to noise floor.
|-
|-

Revision as of 10:42, 28 March 2018

Beam tests of the UCL single module at medAustron.

Friday 23rd - Saturday 24th of March 2018

Pre-data Taking Information and Notes

  • The tests were carried out with an energy of 252.7 MeV with a 5 second beam spill and a range of intensities.
  • No collimators were used at any point.
  • The setup consisted of:
    • 2" Hamamatsu R13089-100-11 PMT with negative HV active divider base (made by Hamamatsu)
    • coupled with Eljen optical gel to a
    • 3 cm x 3 cm x 45 cm cuboid ENVINET standard scintillator wrapped in two layers 6 um (each) of Mylar
  • The "pensioner" 5 dynode stage PMT was also briefly tested, but no sensible data was able to be acquired.
  • The digitiser always operated in Self Trigger mode.
  • Differences since last medAustron test beam:
    • Short cables used for signals and SMA instead of BNC, controlling the acquisition remotely.
    • Tests carried out at a range of HVs.
  • Working directory where digitiser data is stored:

C:\Users\abf\Desktop\digitiser_data\medAustron\mar18\ham

Data Taking

Run number High Voltage, V PMT Current, uA Beam Intensity Rate (Measured on Digitiser) and Notes Digitiser Settings
N/A -900 150.1 (beam off) N/A N/A DC Offset: 0, Rec. Length: 396 ns, Threshold: 40 LSB, Gate Offset: 1 ns, Long Gate: 396 ns, BL Mean: 8 ns, Charge. Sensitivty: 20 fC/LSB, Pre-trigger: 9 ns
001 -900 150.1 Lowest 0.3 - 5 kHz DC Offset: -30, Threshold: 40 LSB, Gate Offset: 20 ns, Long Gate: 150 ns
002 -800 133.2 Lowest 0.3 - 5 kHz As above
003 -700 116.35 Lowest 0.3 - 5 kHz As above
004 -600 99.45 Lowest 0.3 - 5 kHz Beam dropped during running!
005 -600 99.45 Lowest 0.3 - 5 kHz Repeat run with beam back.
006 -500 82.6 Lowest 0.3 - 5 kHz Threshold: 10 LSB. Too close to the threshold, but cannot lower more due to noise floor.