Radiotherapy/Experimental Runs/2022-07-26: Difference between revisions
Jump to navigation
Jump to search
Second run
Replaced plastic blocks from first run with new un-irradiated plastic blocks, but using same green scintillator sheet.
JeremyOcampo (talk | contribs) |
JeremyOcampo (talk | contribs) No edit summary |
||
Line 11: | Line 11: | ||
=== Photos of setup === | === Photos of setup === | ||
<div class="image300px" style="text-align: center;"> | |||
[https://www.hep.ucl.ac.uk/pbt/wikiData/images/Cyberknife_Birmingham_Jul2022/_DSC4718.JPG https://www.hep.ucl.ac.uk/pbt/wikiData/images/Cyberknife_Birmingham_Jul2022/_DSC4718.JPG] | |||
[https://www.hep.ucl.ac.uk/pbt/wikiData/images/Cyberknife_Birmingham_Jul2022/_DSC4715.JPG https://www.hep.ucl.ac.uk/pbt/wikiData/images/Cyberknife_Birmingham_Jul2022/_DSC4715.JPG] | |||
[https://www.hep.ucl.ac.uk/pbt/wikiData/images/Cyberknife_Birmingham_Jul2022/_DSC4713.JPG https://www.hep.ucl.ac.uk/pbt/wikiData/images/Cyberknife_Birmingham_Jul2022/_DSC4713.JPG] | |||
[https://www.hep.ucl.ac.uk/pbt/wikiData/images/Cyberknife_Birmingham_Jul2022/_DSC4716.JPG https://www.hep.ucl.ac.uk/pbt/wikiData/images/Cyberknife_Birmingham_Jul2022/_DSC4716.JPG] | |||
</div> | |||
(Inside pelicase, outside pelicase, collimators, through pelicase window) | |||
=== Pre-data Taking Information and Notes === | === Pre-data Taking Information and Notes === | ||
Revision as of 16:56, 19 August 2022
Second experimental measurements with Cyberknife at Birmingham. Green scintillator thin sheet measurements only.
Setup
Schematic
Photos of setup
(Inside pelicase, outside pelicase, collimators, through pelicase window)
Pre-data Taking Information and Notes
- 53cm from lens to sheet
- Check camera in focus with ruler on the sheet
- Near/Far limit of focus 49.4cm 57.1cm calculated using https://dofmaster.com/dofjs.html
- Pelicase on couch
- Scintillator blocks aligned to laser cross hairs
- 785 mm source to surface distance suitable for output factor/PDD/profile measurements.
- Room lights out and pelicase window open.
- Check optical axis aligned with top surface using crosshair in the image analysis.
- Arrived at hospital at 11am, finished data taking at 5pm
Data Taking
- All images taken with Nikon D7000, 20mm lens, f/2.8, 400ISO
- Used camera control pro 2 app
- Image of ruler taken at sheet for pixels to mm conversion
- 5 images taken at each collimator size with 1 background image each.
- Change exposure time until near max saturation - get's best bit depth
- View data analysis here
First run
With scintillator sheet | |||
---|---|---|---|
Collimator Size (mm) | Exposure (s) | Image | Comments |
5 | 30 | ||
7.5 | 20 | ||
10 | 15 | ||
12.5 | 10 | ||
15 | 8 | ||
20 | 8 | ||
25 | 6 | ||
30 | 6 | ||
35 | 5 | ||
40 | 4 | ||
50 | 3 | ||
60 | 3 |
No scintillator sheet | |||
---|---|---|---|
Collimator Size (mm) | Exposure (s) | Image | Comments |
60 | 3 | Can see green discolouration caused by previous irradiations. Blocks must have been rotated by 90 when re-assembling |
Second run
Replaced plastic blocks from first run with new un-irradiated plastic blocks, but using same green scintillator sheet.
With scintillator sheet | |||
---|---|---|---|
Collimator Size (mm) | Exposure (s) | Image | Comments |
5 | 30 | ||
7.5 | 20 | ||
10 | 15 | ||
12.5 | 10 | ||
15 | 8 | ||
20 | 8 | ||
25 | 6 | ||
30 | 6 | ||
35 | 5 | ||
40 | 4 | ||
50 | 3 | ||
60 | 3 |
Green discolouration probably from scintillator sheet. |
No scintillator sheet | |||
---|---|---|---|
Collimator Size (mm) | Exposure (s) | Image | Comments |
5 | 30 | ||
7.5 | 20 | ||
10 | 15 | ||
12.5 | 10 | ||
15 | 8 | ||
20 | 8 | ||
25 | 6 | ||
30 | 6 | ||
35 | 5 | ||
40 | 4 | ||
50 | 3 | ||
60 | 3 |