Proton Calorimetry/Experimental Runs/2022/ManchesterFLASH: Difference between revisions
SaadShaikh (talk | contribs) No edit summary |
SaadShaikh (talk | contribs) No edit summary |
||
Line 1: | Line 1: | ||
== | == Experiment Equipment == | ||
{| class="wikitable" | |||
! style="text-align: center;" | Item | |||
! style="text-align: center;" | Notes | |||
|- | |||
| style="text-align: center;" | Network Hub | |||
| style="text-align: center;" | Set in control room to take output from experimental room ethernet connection. Control laptops connected via 5GHz WiFi. | |||
|- | |||
| style="text-align: center;" | Control Laptop x31 | |||
| style="text-align: center;" | 1 for remote control of Nikon camera and FPGA, 1 for notes/web GUI. | |||
|- | |||
| style="text-align: center;" | Ethernet Cable x 2 | |||
| style="text-align: center;" | To connect DAQ laptop to network in the experimental room, control laptop to network in control room. | |||
|- | |||
| style="text-align: center;" | Portable Enclosures | |||
| style="text-align: center;" | Modified Big Zarges Waterproof Wheeled Equipment Case. | |||
Features mount for scintillator stack, front and back openings for beam, patch panel with ports for SHV, BNC, SMA, USB, Camera Link cables. <br/> | |||
Hole for sensor powering. <br/> | |||
Modified Small Peli Case Waterproof Wheeled Equipment Case. | |||
Features mount for scintillator stack, front and back openings for beam, patch panel with ports for USB and power connections. | |||
Mylar windows with Al support mounted to internal face: approximately light-tight. | |||
|- | |||
| style="text-align: center;" | Clear Scintillator stack for WET measurement. | |||
| style="text-align: center;" | 25 clear X x 100 mm, Y x 100 mm and Z x (approx) 2.8 mm sheets. Containing sheets (not in order): 79, 68, 73, 62, 81, 72, 64, 67, 65, 71, 78, 77, 74, 75, 85, 82, 83, 76, 70, 80, 69, 84, 63, 66, 61. Full stack thickness measurements (in mm): 71.00, 71.00, 71.01 = 71.00. | |||
|- | |||
| style="text-align: center;" | White Painted Scintillator stack for WET measurement. | |||
| style="text-align: center;" | 23 white painted X x 100 mm, Y x 100 mm and Z x (approx) 2.8 mm sheets. Containing sheets (not in order): 34, 43, 32, 8, 40, 37, 38, 47, 46, 33, 42, 9, 39, 44, 49, 50, 31, 1, 45, 3, 54, 5, 51. Full stack thickness measurements (in mm): 65.51, 65.53, 65.45 = 65.50. | |||
|- | |||
| style="text-align: center;" | Clatterbridge Prototype A | |||
| style="text-align: center;" | 14 clear X x 100 mm, Y x 100 mm and Z x (approx) 2.8 mm sheets. Containing sheets (ascending photodiode order, board upside down): 81, 83, 65, 74, 69, 78, 61, 77, 63, 71, 70, 73, 64, 85. Full stack thickness measurements (in mm): 39.89. | |||
|- | |||
| style="text-align: center;" | Clatterbridge Prototype B | |||
| style="text-align: center;" | 11 clear and 3 white X x 100 mm, Y x 100 mm and Z x (approx) 2.8 mm sheets. Containing sheets (ascending photodiode order, board upside down): 52, 79, 68, 66, 75, 82, 84, 62, 80, 72, 67, 76, 51, 54. Full stack thickness measurements (in mm): 39.92. | |||
|- | |||
| style="text-align: center;" | Rev. D Scintillator stack | |||
| style="text-align: center;" | 32 X x 105 mm, Y x 105 mm and Z x (approx) 3 mm sheets. Containing sheets (ascending photodiode order, board upside down): 137, 138, 139, 140, 141, 142, 143, 144, 145, 146, 147, 148, 149, 150, 151, 152, 153, 154, 155, 156, 157, 158, 159, 160, 161, 162, 163, 164, 165, 166, 167, 168. Full stack thickness measurements (in mm): 95.84, 95.72, 95.78 = 95.78. | |||
|- | |||
| style="text-align: center;" | Large Single Scintillator Block | |||
| style="text-align: center;" | X x 100 mm, Y x 100 mm and Z x 178.57 mm. | |||
|- | |||
| style="text-align: center;" | DAQ laptop x1 | |||
| style="text-align: center;" | Control Nikon camera and photodiode acquisition. | |||
|- | |||
| style="text-align: center;" | Nikon D7500 | |||
| style="text-align: center;" | Connected to DAQ Laptop via USB cable. | |||
|- | |||
| style="text-align: center;" | Nexys Video FPGA development board. | |||
| style="text-align: center;" | For interfacing between DDC232 and PC. | |||
|- | |||
| style="text-align: center;" | Rev. B Texas Instruments DDC232 custom circuit board (x2) | |||
| style="text-align: center;" | Each housing 16x Hamamatsu S12915-16R photodiodes, coupled directly to scintillator sheets. Aligned by eye to scintillator sheets. | |||
|- | |||
| style="text-align: center;" | Rev. D Texas Instruments DDC232 custom circuit board (x1) | |||
| style="text-align: center;" | Housing 32x Hamamatsu S12915-16R photodiodes, coupled directly to scintillator sheets. Aligned by eye to scintillator sheets. | |||
|- | |||
| style="text-align: center;" | Gloves | |||
| style="text-align: center;" | For handling scintillator | |||
|} | |||
==== Schematic ==== | |||
<div class="image600px" style="text-align: center;"> | |||
[http://www.hep.ucl.ac.uk/pbt/wikiData/images/UCLH_Nov2021/UCL_Setup.png http://www.hep.ucl.ac.uk/pbt/wikiData/images/UCLH_Jun2022/UCL_Setup.png] | |||
</div> | |||
== Experiment Plan == | == Experiment Plan == | ||
Estimated first beam: 16.00. | |||
# Set up mains cabling and networking in treatment and control rooms. | |||
# Install Peli case and Giraffe on treatment couch: | |||
## Set up 2 x 2.86mm stacks for shoot through measurement of range pullback at ~245MeV (no DAQ needed). | |||
## Align Peli case and Giraffe using in-room lasers. | |||
## Measure range pullback of combined stack, each individual stack and empty case. | |||
# Set up Peli case for range pullback measurement of single block: | |||
## Remove long rail from Peli case and install single scintillator block. | |||
## Simultaneously dismantle 2 x 2.86mm stacks and assemble Clatterbridge stacks with rev.B DDC board. | |||
## Check alignment of Peli case and Giraffe. | |||
## Measure range pullback of single block. | |||
# Set up Peli case for range pullback measurement of 32 x 3mm stack: | |||
## Remove single scintillator block. | |||
## Install 32 x 3mm stack with rotated rail; reinstall long rail. | |||
## Simultaneously install single scintillator block in Zarges case with camera. | |||
## Check alignment of Peli case and Giraffe. | |||
## Measure range pullback of 32 x 3mm stack. | |||
# Set up Peli case for range pullback measurement of Clatterbridge stacks: | |||
## Remove 32 x 3mm stack; install Clatterbridge stacks on long rail and clamp in place. | |||
## Check alignment of Peli case and Giraffe. | |||
## Measure range pullback of Clatterbridge stacks together and individually. | |||
# Set up Zarges case for single block quenching measurements: | |||
## Remove Peli case and Giraffe from treatment couch. | |||
## Install Zarges case on treatment couch and align with room lasers. | |||
## Connect DAQ and check camera functionality. | |||
## Simultaneously set up Peli case with Clatterbridge stacks and FPGAs ready for shoot throughs. | |||
## Measure light output from shoot through and set exposure time. | |||
## Record photos of single block light output for energies from 70 MeV to ~150MeV (maximum energy with beam still in block). | |||
# Set up Peli case for Clatterbridge DAQ measurements: | |||
## Remove Zarges case from treatment couch; install Peli case with FPGA, cabling and Clatterbridge stacks. | |||
## Check alignment of Peli case and DAQ hardware. | |||
## Record DDC calibration data for shoot through and background. | |||
# Set up Peli case for 32 x 3mm stack measurements: | |||
## Remove Clatterbridge stacks and store safely. | |||
## Install 32 x 3mm stack in Peli case with rev.C/D DDC board. | |||
## Check alignment of Peli case and DAQ hardware. | |||
## Measure background and shoot through of 32 x 3mm stack. | |||
## Measure proton energies in range 70–110 MeV. | |||
# Dismantle equipment, pack up and fuck off. | |||
== Experiment Notes == | |||
Range calorimeter measurements with photodiodes, camera and IBA Giraffe. | |||
Images from Camera: '''[http://www.hep.ucl.ac.uk/pbt/wikiData/images/UCLH_Sep2022/Nikon /unix/www/html/pbt/wikiData/images/UCLH_Sep2022/Nikon]''' | |||
Photodiode hexadecimal data: '''/unix/pbt/data/uclh/20220901/DDC232''' | |||
Giraffe data: '''/unix/pbt/data/uclh/20220901/Giraffe''' | |||
=== PTW PeakFinder Range Pull Back Measurements === | |||
{| class="wikitable" | |||
!Run number | |||
!Detector config | |||
!Beam Energy (MeV) | |||
!Beam Current (nA) | |||
!Physical thickness (mm) | |||
!D80 measured (mm) | |||
!Comments | |||
|- | |||
|0 || Clear stack + white stack || 210 || 40 || N/A || 139.1 || Nominal: 282.1 mm | |||
|- | |||
|1 || Clear stack || 210 || 40 || N/A || 206.6 || Nominal: 282.1 mm | |||
|} | |||
=== DDC232 Calibration and Bragg Peak Measurements === | |||
{| class="wikitable" | |||
!Run number | |||
!Detector | |||
!Beam Energy (MeV) | |||
!Estimated Range (mm) | |||
!Current (nA) | |||
!Spot size (mm, FWHM) | |||
!DDC232 FSR (pC) | |||
!DDC232 Integration Time (us) | |||
!Degrader WET (mm) | |||
!Comments | |||
!Photodiode QB Fit | |||
!Replay Fit | |||
|- | |||
|0 || Clatter. A || Background || || - || N/A || 12.5 (constant) || 170 || || || N/A || | |||
|- | |||
|1 || Clatter. A || 245 MeV || || 10 nA || N/A || 12.5 || 170 || || Gantry angle 270, shoot through || N/A || | |||
|} |
Revision as of 16:14, 25 October 2022
Experiment Equipment
Item | Notes |
---|---|
Network Hub | Set in control room to take output from experimental room ethernet connection. Control laptops connected via 5GHz WiFi. |
Control Laptop x31 | 1 for remote control of Nikon camera and FPGA, 1 for notes/web GUI. |
Ethernet Cable x 2 | To connect DAQ laptop to network in the experimental room, control laptop to network in control room. |
Portable Enclosures | Modified Big Zarges Waterproof Wheeled Equipment Case.
Features mount for scintillator stack, front and back openings for beam, patch panel with ports for SHV, BNC, SMA, USB, Camera Link cables. |
Clear Scintillator stack for WET measurement. | 25 clear X x 100 mm, Y x 100 mm and Z x (approx) 2.8 mm sheets. Containing sheets (not in order): 79, 68, 73, 62, 81, 72, 64, 67, 65, 71, 78, 77, 74, 75, 85, 82, 83, 76, 70, 80, 69, 84, 63, 66, 61. Full stack thickness measurements (in mm): 71.00, 71.00, 71.01 = 71.00. |
White Painted Scintillator stack for WET measurement. | 23 white painted X x 100 mm, Y x 100 mm and Z x (approx) 2.8 mm sheets. Containing sheets (not in order): 34, 43, 32, 8, 40, 37, 38, 47, 46, 33, 42, 9, 39, 44, 49, 50, 31, 1, 45, 3, 54, 5, 51. Full stack thickness measurements (in mm): 65.51, 65.53, 65.45 = 65.50. |
Clatterbridge Prototype A | 14 clear X x 100 mm, Y x 100 mm and Z x (approx) 2.8 mm sheets. Containing sheets (ascending photodiode order, board upside down): 81, 83, 65, 74, 69, 78, 61, 77, 63, 71, 70, 73, 64, 85. Full stack thickness measurements (in mm): 39.89. |
Clatterbridge Prototype B | 11 clear and 3 white X x 100 mm, Y x 100 mm and Z x (approx) 2.8 mm sheets. Containing sheets (ascending photodiode order, board upside down): 52, 79, 68, 66, 75, 82, 84, 62, 80, 72, 67, 76, 51, 54. Full stack thickness measurements (in mm): 39.92. |
Rev. D Scintillator stack | 32 X x 105 mm, Y x 105 mm and Z x (approx) 3 mm sheets. Containing sheets (ascending photodiode order, board upside down): 137, 138, 139, 140, 141, 142, 143, 144, 145, 146, 147, 148, 149, 150, 151, 152, 153, 154, 155, 156, 157, 158, 159, 160, 161, 162, 163, 164, 165, 166, 167, 168. Full stack thickness measurements (in mm): 95.84, 95.72, 95.78 = 95.78. |
Large Single Scintillator Block | X x 100 mm, Y x 100 mm and Z x 178.57 mm. |
DAQ laptop x1 | Control Nikon camera and photodiode acquisition. |
Nikon D7500 | Connected to DAQ Laptop via USB cable. |
Nexys Video FPGA development board. | For interfacing between DDC232 and PC. |
Rev. B Texas Instruments DDC232 custom circuit board (x2) | Each housing 16x Hamamatsu S12915-16R photodiodes, coupled directly to scintillator sheets. Aligned by eye to scintillator sheets. |
Rev. D Texas Instruments DDC232 custom circuit board (x1) | Housing 32x Hamamatsu S12915-16R photodiodes, coupled directly to scintillator sheets. Aligned by eye to scintillator sheets. |
Gloves | For handling scintillator |
Schematic
Experiment Plan
Estimated first beam: 16.00.
- Set up mains cabling and networking in treatment and control rooms.
- Install Peli case and Giraffe on treatment couch:
- Set up 2 x 2.86mm stacks for shoot through measurement of range pullback at ~245MeV (no DAQ needed).
- Align Peli case and Giraffe using in-room lasers.
- Measure range pullback of combined stack, each individual stack and empty case.
- Set up Peli case for range pullback measurement of single block:
- Remove long rail from Peli case and install single scintillator block.
- Simultaneously dismantle 2 x 2.86mm stacks and assemble Clatterbridge stacks with rev.B DDC board.
- Check alignment of Peli case and Giraffe.
- Measure range pullback of single block.
- Set up Peli case for range pullback measurement of 32 x 3mm stack:
- Remove single scintillator block.
- Install 32 x 3mm stack with rotated rail; reinstall long rail.
- Simultaneously install single scintillator block in Zarges case with camera.
- Check alignment of Peli case and Giraffe.
- Measure range pullback of 32 x 3mm stack.
- Set up Peli case for range pullback measurement of Clatterbridge stacks:
- Remove 32 x 3mm stack; install Clatterbridge stacks on long rail and clamp in place.
- Check alignment of Peli case and Giraffe.
- Measure range pullback of Clatterbridge stacks together and individually.
- Set up Zarges case for single block quenching measurements:
- Remove Peli case and Giraffe from treatment couch.
- Install Zarges case on treatment couch and align with room lasers.
- Connect DAQ and check camera functionality.
- Simultaneously set up Peli case with Clatterbridge stacks and FPGAs ready for shoot throughs.
- Measure light output from shoot through and set exposure time.
- Record photos of single block light output for energies from 70 MeV to ~150MeV (maximum energy with beam still in block).
- Set up Peli case for Clatterbridge DAQ measurements:
- Remove Zarges case from treatment couch; install Peli case with FPGA, cabling and Clatterbridge stacks.
- Check alignment of Peli case and DAQ hardware.
- Record DDC calibration data for shoot through and background.
- Set up Peli case for 32 x 3mm stack measurements:
- Remove Clatterbridge stacks and store safely.
- Install 32 x 3mm stack in Peli case with rev.C/D DDC board.
- Check alignment of Peli case and DAQ hardware.
- Measure background and shoot through of 32 x 3mm stack.
- Measure proton energies in range 70–110 MeV.
- Dismantle equipment, pack up and fuck off.
Experiment Notes
Range calorimeter measurements with photodiodes, camera and IBA Giraffe.
Images from Camera: /unix/www/html/pbt/wikiData/images/UCLH_Sep2022/Nikon
Photodiode hexadecimal data: /unix/pbt/data/uclh/20220901/DDC232
Giraffe data: /unix/pbt/data/uclh/20220901/Giraffe
PTW PeakFinder Range Pull Back Measurements
Run number | Detector config | Beam Energy (MeV) | Beam Current (nA) | Physical thickness (mm) | D80 measured (mm) | Comments |
---|---|---|---|---|---|---|
0 | Clear stack + white stack | 210 | 40 | N/A | 139.1 | Nominal: 282.1 mm |
1 | Clear stack | 210 | 40 | N/A | 206.6 | Nominal: 282.1 mm |
DDC232 Calibration and Bragg Peak Measurements
Run number | Detector | Beam Energy (MeV) | Estimated Range (mm) | Current (nA) | Spot size (mm, FWHM) | DDC232 FSR (pC) | DDC232 Integration Time (us) | Degrader WET (mm) | Comments | Photodiode QB Fit | Replay Fit |
---|---|---|---|---|---|---|---|---|---|---|---|
0 | Clatter. A | Background | - | N/A | 12.5 (constant) | 170 | N/A | ||||
1 | Clatter. A | 245 MeV | 10 nA | N/A | 12.5 | 170 | Gantry angle 270, shoot through | N/A |