Proton Calorimetry/Experimental Runs/2021/Apr15: Difference between revisions
SaadShaikh (talk | contribs) |
SaadShaikh (talk | contribs) |
||
Line 121: | Line 121: | ||
|0 || high || Background || N/A || N/A || N/A || 350 || 170 || 0 || room light on || | |0 || high || Background || N/A || N/A || N/A || 350 || 170 || 0 || room light on || | ||
|- | |- | ||
|1 || high || Background || N/A || N/A || N/A || 12.5 || 170 || 0 || room light on || | |1 || high || Background || N/A || N/A || N/A || 12.5 || 170 || 0 || repeat – room light on || | ||
|- | |- | ||
|2 || high || Background || N/A || N/A || N/A || 12.5 || 170 || 0 || room lights off || | |2 || high || Background || N/A || N/A || N/A || 12.5 || 170 || 0 || room lights off || | ||
Line 127: | Line 127: | ||
|3 || high || 245 || 366.3 || 10 || 7.1 || 12.5 || 170 || 57.33 || "ST, gantry angle 270" || | |3 || high || 245 || 366.3 || 10 || 7.1 || 12.5 || 170 || 57.33 || "ST, gantry angle 270" || | ||
|- | |- | ||
|4 || high || 245 || 20 || 7.1 || 12.5 || 170 || 57.33 || "ST, gantry angle 270, CMOS sensor saturated" || | |4 || high || 245 || || 20 || 7.1 || 12.5 || 170 || 57.33 || "ST, gantry angle 270, CMOS sensor saturated" || | ||
|- | |- | ||
|5 || high || 245 || 366.3 || 10 || 7.1 || 12.5 || 170 || 57.33 || "ST, gantry angle 90" || | |5 || high || 245 || 366.3 || 10 || 7.1 || 12.5 || 170 || 57.33 || "ST, gantry angle 90" || | ||
Line 135: | Line 135: | ||
|7 || high || 105 || || 300 || 7.1 || 12.5 || 170 || 0 || "CMOS sensor saturated, BP seen" || | |7 || high || 105 || || 300 || 7.1 || 12.5 || 170 || 0 || "CMOS sensor saturated, BP seen" || | ||
|- | |- | ||
|8 high 105 | |8 || high || 105 || || 150 || 7.1 || 12.5 || 170 || 0 || "CMOS sensor saturated, BP seen" || | ||
|- | |- | ||
|9 high 70 | |9 || high || 70 || || 700 || 7.1 || 12.5 || 170 || 0 || "CMOS sensor saturated, BP seen" || | ||
|- | |- | ||
|10 high 70 | |10 || high || 70 || || 350 7.1 || 12.5 || 170 || 0 || Having issues with beam requests (sharing between 2 gantries has commenced) || | ||
|- | |- | ||
|11 high 80 | |11 || high || 80 || || 500 || 7.1 || 12.5 || 170 || 0 || Couple 1000 off saturation || | ||
|- | |- | ||
|12 high 90 | |12 || high || 90 || || 400 || 7.1 || 12.5 || 170 || 0 || Saturation || | ||
|- | |- | ||
|13 high 100 | |13 || high || 100 || || 350 || 7.1 || 12.5 || 170 || 0 || || | ||
|- | |- | ||
|14 high 110 | |14 || high || 110 || || 300 || 7.1 || 12.5 || 170 || 0 || Saturation|| | ||
|- | |- | ||
|15 high 120 | |15 || high || 120 || || 270 || 7.1 || 12.5 || 170 || 0 || BP not captured in range || | ||
|- | |- | ||
|16 high 110 | |16 || high || 110 || || 150 || 7.1 || 12.5 || 170 || 0 || || | ||
|- | |- | ||
|17 high 105 | |17 || high || 105 || || 350 || 7.1 || 12.5 || 170 || 0 || Saturated || | ||
|- | |- | ||
|18 high 100 | |18 || high || 100 || || 200 || 7.1 || 12.5 || 170 || 0 || Saturated || | ||
|- | |- | ||
|19 high 100 | |19 || high || 100 || || 350 || 7.1 || 12.5 || 1000 || 0 || Saturated || | ||
|- | |- | ||
|20 high 95 | |20 || high || 95 || || 400 || 7.1 || 12.5 || 170 || 0 || Saturated || | ||
|- | |- | ||
|21 high 90 | |21 || high || 90 || || 200 || 7.1 || 12.5 || 170 || 0 || || | ||
|- | |- | ||
|22 high 90 | |22 || high || 90 || || 400 || 7.1 || 12.5 || 1000 || 0 || || | ||
|- | |- | ||
|23 high 85 | |23 || high || 85 || || 500 || 7.1 || 12.5 || 170 || 0 || just missed saturation || | ||
|- | |- | ||
|24 high 80 | |24 || high || 80 || || 250 || 7.1 || 12.5 || 170 || 0 || || | ||
|- | |- | ||
|25 high 80 | |25 || high || 80 || || 500 || 7.1 || 12.5 || 1000 || 0 || || | ||
|- | |- | ||
|26 high 75 | |26 || high || 75 || || 600 || 7.1 || 12.5 || 170 || 0 || || | ||
|- | |- | ||
|27 high 70 | |27 || high || 70 || || 350 || 7.1 || 12.5 || 170 || 0 || || | ||
|- | |- | ||
|28 high 70 | |28 || high || 70 || || 700 || 7.1 || 12.5 || 1000 || 0 || || | ||
|- | |- | ||
|29 | |29 || N/A || Treatment Plan || N/A || N/A || N/A || 12.5 || 170 || 0 || clinical mode from now - scanned field || | ||
|- | |- | ||
|30 | |30 || N/A || Treatment Plan || N/A || N/A || N/A || 12.5 || 170 || 0 || repeat || | ||
|- | |- | ||
|31 | |31 || N/A || Treatment Plan || N/A || N/A || N/A || 50 || 170 || 0 || || | ||
|- | |- |
Revision as of 14:34, 19 April 2021
Afternoon test at UCLH with photodiodes and CMOS sesnor.
Experiment Equipment
Item | Notes |
---|---|
Network Hub | Set in control room to take output from experimental room ethernet connection. Control laptops connected via 5GHz WiFi. |
Control Laptop x3 | 2x remote control for sensor and FPGA, 1 for notes/web GUI. |
Ethernet Cable x 3 | To connect DAQ laptops to network in the experimental room, network hub to network in control room. |
Portable Enclosure | Modified Big Zarges Waterproof Wheeled Equipment Case.
Features mount for scintillator stack, front and back openings for beam, patch panel with ports for SHV, BNC, SMA, USB, Camera Link cables. |
Scintillator stack | X x 2 mm, Y x 2.6 mm and Z x 3 mm sheets in ascending order from the front/beam of the end of the scintillator. Sheet numbers from front to end (in beam direction): UNKNOWN. |
ISDI CMOS sensor | sensor pixel dimension: 1030 x 1536. NO optical grease between scintillator and sensor. Connected to DAQ PC via Camera Link cable. |
Nikon D7500 | Connected to DAQ Laptop via USB cable. |
DAQ laptop x2 | Control sensor and photodiode acquisition. |
Nexys Video FPGA development board. | For interfacing between DDC232 and PC. |
Rev. 2 Texas Instruments DDC232 custom circuit board (x2) | Each housing 16x Hamamatsu S12915-16R photodiodes, coupled directly to scintillator sheets. Aligned by eye to scintillator sheets. |
Gloves | For handling scintillator |
Digital Caliper | For measuring PMMA air gaps |
Schematic
ISDI Sensor Instructions
General Sensor Configuration
- Base Configuration is 14 x 1 Bit
- Resolution is 1030 x 1536 pixels
- For selecting pixel rows in centre: Use horizontal offset (ignore DVAL to activate field) and horizontal resolution
Increase number of frame buffers
- To increase the number of frame buffers to the allowed maximum, start XCAP as administrator.
- Select Capture/Sequence Capture/Video to Frame Buffers/Driver Assistant
- Tick box in left field, then set number of frame buffers here.
Serial Terminal instructions for ISDI sensor
- Find serial terminal in PIXCI(R) tab of sensor control software.
- In Controls/Options: Make sure "Send string with CR" is selected. Should be default.
- In Controls/Setup: Select Serial Port: Enabled. Baud rate has to be 115200
- Set low full well mode using W000300000
- Set high full well mode using W000300004
- Read full well mode using R0003
Continuous Field buffer setting
- This setting makes sure that the buffer is filled with a new frame as they come in. If this option is not selected the sensor and software will not be synchronised and we *will record a lot of pitch black images.
- Select Capture/Live Options//Live Mode/Live Video: Continuous All buffer capture
DDC232 Instructions
- Bash script
prog.sh
executesFTDI.cpp
andlivePlot.cpp
. Execute with. prog.sh 2 350 170 0 50 true Data/capture.txt
in the directoryPhotodiodes
.- First argument is the number of DDC232s.
- Second argument is the full-scale range (FSR) in pC. Choices are: 12.5, 50, 100, 150, 200, 250, 300, 350.
- Third argument is the integration time in us. Range is 170-1455. Set in multiples of 5 us.
- Fourth argument is the number of measurements to acquire. Typically acquire 5700 measurements per second. Set to 0 for infinite loop.
- Fifth argument is the rate in Hz to update the file
average.csv
for ROOT/web GUI plotting. - Sixth argument is a boolean to choose whether to display the ROOT plot.
- Seventh argument is the filename and destination for the output hexadecimal data.
- After acquisition, close ROOT by choosing File->Quit in the ROOT window and then stopping DAQ (if in infinite loop) with CTRL-C.
Experiment Plan
- Shoot-through measurements (back & front) for light calibration and level setting.
- Range measurements between 70-120 MeV with 1mm steps with CMOS sensor and photodiodes.
- High intensity range measurements between 70-120 MeV with 1mm steps with photodiodes at optimal light level.
- Live measurement of 3cm x 3cm x 3cm treatment plan.
Experiment Notes
Range calorimeter measurements with ISDI sensor + photodiodes.
Images from sensor: TBD
Photodiode hexadecimal data: TBD
- full sensor area
- shooting on BACK WINDOW
Run number | Full Well Mode | Beam Energy (MeV/u) | Range (estimated) | Current (nA) | Spot size (mm) | DDC232 FSR (pC) | DDC232 Integration Time (us) | Degrader WET (mm) | Comments | Thumbnail |
---|---|---|---|---|---|---|---|---|---|---|
Run | Full-well | Energy | Current | Range | Spot Size | FSR | t_INT | Degrader | Notes | |
00 | high | N/A | - | N/A | None | 350 | 170 | 0 | background | |
0 | high | Background | N/A | N/A | N/A | 350 | 170 | 0 | room light on | |
1 | high | Background | N/A | N/A | N/A | 12.5 | 170 | 0 | repeat – room light on | |
2 | high | Background | N/A | N/A | N/A | 12.5 | 170 | 0 | room lights off | |
3 | high | 245 | 366.3 | 10 | 7.1 | 12.5 | 170 | 57.33 | "ST, gantry angle 270" | |
4 | high | 245 | 20 | 7.1 | 12.5 | 170 | 57.33 | "ST, gantry angle 270, CMOS sensor saturated" | ||
5 | high | 245 | 366.3 | 10 | 7.1 | 12.5 | 170 | 57.33 | "ST, gantry angle 90" | |
6 | high | 245 | 366.3 | 10 | 7.1 | 12.5 | 170 | 57.33 | "ST, gantry angle 90" | |
7 | high | 105 | 300 | 7.1 | 12.5 | 170 | 0 | "CMOS sensor saturated, BP seen" | ||
8 | high | 105 | 150 | 7.1 | 12.5 | 170 | 0 | "CMOS sensor saturated, BP seen" | ||
9 | high | 70 | 700 | 7.1 | 12.5 | 170 | 0 | "CMOS sensor saturated, BP seen" | ||
10 | high | 70 | 350 7.1 | 12.5 | 170 | 0 | Having issues with beam requests (sharing between 2 gantries has commenced) | |||
11 | high | 80 | 500 | 7.1 | 12.5 | 170 | 0 | Couple 1000 off saturation | ||
12 | high | 90 | 400 | 7.1 | 12.5 | 170 | 0 | Saturation | ||
13 | high | 100 | 350 | 7.1 | 12.5 | 170 | 0 | |||
14 | high | 110 | 300 | 7.1 | 12.5 | 170 | 0 | Saturation | ||
15 | high | 120 | 270 | 7.1 | 12.5 | 170 | 0 | BP not captured in range | ||
16 | high | 110 | 150 | 7.1 | 12.5 | 170 | 0 | |||
17 | high | 105 | 350 | 7.1 | 12.5 | 170 | 0 | Saturated | ||
18 | high | 100 | 200 | 7.1 | 12.5 | 170 | 0 | Saturated | ||
19 | high | 100 | 350 | 7.1 | 12.5 | 1000 | 0 | Saturated | ||
20 | high | 95 | 400 | 7.1 | 12.5 | 170 | 0 | Saturated | ||
21 | high | 90 | 200 | 7.1 | 12.5 | 170 | 0 | |||
22 | high | 90 | 400 | 7.1 | 12.5 | 1000 | 0 | |||
23 | high | 85 | 500 | 7.1 | 12.5 | 170 | 0 | just missed saturation | ||
24 | high | 80 | 250 | 7.1 | 12.5 | 170 | 0 | |||
25 | high | 80 | 500 | 7.1 | 12.5 | 1000 | 0 | |||
26 | high | 75 | 600 | 7.1 | 12.5 | 170 | 0 | |||
27 | high | 70 | 350 | 7.1 | 12.5 | 170 | 0 | |||
28 | high | 70 | 700 | 7.1 | 12.5 | 1000 | 0 | |||
29 | N/A | Treatment Plan | N/A | N/A | N/A | 12.5 | 170 | 0 | clinical mode from now - scanned field | |
30 | N/A | Treatment Plan | N/A | N/A | N/A | 12.5 | 170 | 0 | repeat | |
31 | N/A | Treatment Plan | N/A | N/A | N/A | 50 | 170 | 0 |