Line 107: | Line 107: | ||
!Beam Energy (MeV/u) | !Beam Energy (MeV/u) | ||
!Range (estimated) | !Range (estimated) | ||
! | !Current (nA) | ||
!Spot size (mm) | !Spot size (mm) | ||
!DDC232 FSR (pC) | !DDC232 FSR (pC) | ||
Line 115: | Line 115: | ||
!Thumbnail | !Thumbnail | ||
|- | |- | ||
|00 || high || | |Run || Full-well || Energy || Current || Range || Spot Size || FSR || t_INT || Degrader || Notes || | ||
|- | |||
|00 || high || N/A || - || N/A || None || 350 || 170 || 0 || background || | |||
|- | |||
|0 || high || Background || N/A || N/A || N/A || 350 || 170 || 0 || room light on || | |||
|- | |||
|1 || high || Background || N/A || N/A || N/A || 12.5 || 170 || 0 || room light on || | |||
|- | |||
|2 || high || Background || N/A || N/A || N/A || 12.5 || 170 || 0 || room lights off || | |||
|- | |||
|3 || high || 245 || 366.3 || 10 || 7.1 || 12.5 || 170 || 57.33 || "ST, gantry angle 270" || | |||
|- | |||
|4 || high || 245 || 20 || 7.1 || 12.5 || 170 || 57.33 || "ST, gantry angle 270, CMOS sensor saturated" || | |||
|- | |||
|5 || high || 245 || 366.3 || 10 || 7.1 || 12.5 || 170 || 57.33 || "ST, gantry angle 90" || | |||
|- | |||
|6 || high || 245 || 366.3 || 10 || 7.1 || 12.5 || 170 || 57.33 || "ST, gantry angle 90" || | |||
|- | |||
|7 || high || 105 || || 300 || 7.1 || 12.5 || 170 || 0 || "CMOS sensor saturated, BP seen" || | |||
|- | |||
|8 high 105 150 7.1 12.5 170 0 "CMOS sensor saturated, BP seen" | |||
|- | |||
|9 high 70 700 7.1 12.5 170 0 "CMOS sensor saturated, BP seen" | |||
|- | |||
|10 high 70 350 7.1 12.5 170 0 Having issues with beam requests (sharing between 2 gantries has commenced) | |||
|- | |||
|11 high 80 500 7.1 12.5 170 0 Couple 1000 off saturation | |||
|- | |||
|12 high 90 400 7.1 12.5 170 0 Saturation | |||
|- | |||
|13 high 100 350 7.1 12.5 170 0 | |||
|- | |||
|14 high 110 300 7.1 12.5 170 0 Saturation | |||
|- | |||
|15 high 120 270 7.1 12.5 170 0 BP not captured in range | |||
|- | |||
|16 high 110 150 7.1 12.5 170 0 | |||
|- | |||
|17 high 105 350 7.1 12.5 170 0 Saturated | |||
|- | |||
|18 high 100 200 7.1 12.5 170 0 Saturated | |||
|- | |||
|19 high 100 350 7.1 12.5 1000 0 Saturated | |||
|- | |||
|20 high 95 400 7.1 12.5 170 0 Saturated | |||
|- | |||
|21 high 90 200 7.1 12.5 170 0 | |||
|- | |||
|22 high 90 400 7.1 12.5 1000 0 | |||
|- | |||
|23 high 85 500 7.1 12.5 170 0 just missed saturation | |||
|- | |||
|24 high 80 250 7.1 12.5 170 0 | |||
|- | |||
|25 high 80 500 7.1 12.5 1000 0 | |||
|- | |||
|26 high 75 600 7.1 12.5 170 0 | |||
|- | |||
|27 high 70 350 7.1 12.5 170 0 | |||
|- | |||
|28 high 70 700 7.1 12.5 1000 0 pretty | |||
|- | |||
|29 clinical mode from now - scanned field | |||
|- | |||
|30 same field | |||
|- | |||
|31 | |||
|- | |- | ||
Revision as of 14:15, 19 April 2021
Afternoon test at UCLH with photodiodes and CMOS sesnor.
Experiment Equipment
Item | Notes |
---|---|
Network Hub | Set in control room to take output from experimental room ethernet connection. Control laptops connected via 5GHz WiFi. |
Control Laptop x3 | 2x remote control for sensor and FPGA, 1 for notes/web GUI. |
Ethernet Cable x 3 | To connect DAQ laptops to network in the experimental room, network hub to network in control room. |
Portable Enclosure | Modified Big Zarges Waterproof Wheeled Equipment Case.
Features mount for scintillator stack, front and back openings for beam, patch panel with ports for SHV, BNC, SMA, USB, Camera Link cables. |
Scintillator stack | X x 2 mm, Y x 2.6 mm and Z x 3 mm sheets in ascending order from the front/beam of the end of the scintillator. Sheet numbers from front to end (in beam direction): UNKNOWN. |
ISDI CMOS sensor | sensor pixel dimension: 1030 x 1536. NO optical grease between scintillator and sensor. Connected to DAQ PC via Camera Link cable. |
Nikon D7500 | Connected to DAQ Laptop via USB cable. |
DAQ laptop x2 | Control sensor and photodiode acquisition. |
Nexys Video FPGA development board. | For interfacing between DDC232 and PC. |
Rev. 2 Texas Instruments DDC232 custom circuit board (x2) | Each housing 16x Hamamatsu S12915-16R photodiodes, coupled directly to scintillator sheets. Aligned by eye to scintillator sheets. |
Gloves | For handling scintillator |
Digital Caliper | For measuring PMMA air gaps |
Schematic
ISDI Sensor Instructions
General Sensor Configuration
- Base Configuration is 14 x 1 Bit
- Resolution is 1030 x 1536 pixels
- For selecting pixel rows in centre: Use horizontal offset (ignore DVAL to activate field) and horizontal resolution
Increase number of frame buffers
- To increase the number of frame buffers to the allowed maximum, start XCAP as administrator.
- Select Capture/Sequence Capture/Video to Frame Buffers/Driver Assistant
- Tick box in left field, then set number of frame buffers here.
Serial Terminal instructions for ISDI sensor
- Find serial terminal in PIXCI(R) tab of sensor control software.
- In Controls/Options: Make sure "Send string with CR" is selected. Should be default.
- In Controls/Setup: Select Serial Port: Enabled. Baud rate has to be 115200
- Set low full well mode using W000300000
- Set high full well mode using W000300004
- Read full well mode using R0003
Continuous Field buffer setting
- This setting makes sure that the buffer is filled with a new frame as they come in. If this option is not selected the sensor and software will not be synchronised and we *will record a lot of pitch black images.
- Select Capture/Live Options//Live Mode/Live Video: Continuous All buffer capture
DDC232 Instructions
- Bash script
prog.sh
executesFTDI.cpp
andlivePlot.cpp
. Execute with. prog.sh 2 350 170 0 50 true Data/capture.txt
in the directoryPhotodiodes
.- First argument is the number of DDC232s.
- Second argument is the full-scale range (FSR) in pC. Choices are: 12.5, 50, 100, 150, 200, 250, 300, 350.
- Third argument is the integration time in us. Range is 170-1455. Set in multiples of 5 us.
- Fourth argument is the number of measurements to acquire. Typically acquire 5700 measurements per second. Set to 0 for infinite loop.
- Fifth argument is the rate in Hz to update the file
average.csv
for ROOT/web GUI plotting. - Sixth argument is a boolean to choose whether to display the ROOT plot.
- Seventh argument is the filename and destination for the output hexadecimal data.
- After acquisition, close ROOT by choosing File->Quit in the ROOT window and then stopping DAQ (if in infinite loop) with CTRL-C.
Experiment Plan
- Shoot-through measurements (back & front) for light calibration and level setting.
- Range measurements between 70-120 MeV with 1mm steps with CMOS sensor and photodiodes.
- High intensity range measurements between 70-120 MeV with 1mm steps with photodiodes at optimal light level.
- Live measurement of 3cm x 3cm x 3cm treatment plan.
Experiment Notes
Range calorimeter measurements with ISDI sensor + photodiodes.
Images from sensor: TBD
Photodiode hexadecimal data: TBD
- full sensor area
- shooting on BACK WINDOW
Run number | Full Well Mode | Beam Energy (MeV/u) | Range (estimated) | Current (nA) | Spot size (mm) | DDC232 FSR (pC) | DDC232 Integration Time (us) | Degrader WET (mm) | Comments | Thumbnail |
---|---|---|---|---|---|---|---|---|---|---|
Run | Full-well | Energy | Current | Range | Spot Size | FSR | t_INT | Degrader | Notes | |
00 | high | N/A | - | N/A | None | 350 | 170 | 0 | background | |
0 | high | Background | N/A | N/A | N/A | 350 | 170 | 0 | room light on | |
1 | high | Background | N/A | N/A | N/A | 12.5 | 170 | 0 | room light on | |
2 | high | Background | N/A | N/A | N/A | 12.5 | 170 | 0 | room lights off | |
3 | high | 245 | 366.3 | 10 | 7.1 | 12.5 | 170 | 57.33 | "ST, gantry angle 270" | |
4 | high | 245 | 20 | 7.1 | 12.5 | 170 | 57.33 | "ST, gantry angle 270, CMOS sensor saturated" | ||
5 | high | 245 | 366.3 | 10 | 7.1 | 12.5 | 170 | 57.33 | "ST, gantry angle 90" | |
6 | high | 245 | 366.3 | 10 | 7.1 | 12.5 | 170 | 57.33 | "ST, gantry angle 90" | |
7 | high | 105 | 300 | 7.1 | 12.5 | 170 | 0 | "CMOS sensor saturated, BP seen" | ||
8 high 105 150 7.1 12.5 170 0 "CMOS sensor saturated, BP seen" | ||||||||||
9 high 70 700 7.1 12.5 170 0 "CMOS sensor saturated, BP seen" | ||||||||||
10 high 70 350 7.1 12.5 170 0 Having issues with beam requests (sharing between 2 gantries has commenced) | ||||||||||
11 high 80 500 7.1 12.5 170 0 Couple 1000 off saturation | ||||||||||
12 high 90 400 7.1 12.5 170 0 Saturation | ||||||||||
13 high 100 350 7.1 12.5 170 0 | ||||||||||
14 high 110 300 7.1 12.5 170 0 Saturation | ||||||||||
15 high 120 270 7.1 12.5 170 0 BP not captured in range | ||||||||||
16 high 110 150 7.1 12.5 170 0 | ||||||||||
17 high 105 350 7.1 12.5 170 0 Saturated | ||||||||||
18 high 100 200 7.1 12.5 170 0 Saturated | ||||||||||
19 high 100 350 7.1 12.5 1000 0 Saturated | ||||||||||
20 high 95 400 7.1 12.5 170 0 Saturated | ||||||||||
21 high 90 200 7.1 12.5 170 0 | ||||||||||
22 high 90 400 7.1 12.5 1000 0 | ||||||||||
23 high 85 500 7.1 12.5 170 0 just missed saturation | ||||||||||
24 high 80 250 7.1 12.5 170 0 | ||||||||||
25 high 80 500 7.1 12.5 1000 0 | ||||||||||
26 high 75 600 7.1 12.5 170 0 | ||||||||||
27 high 70 350 7.1 12.5 170 0 | ||||||||||
28 high 70 700 7.1 12.5 1000 0 pretty | ||||||||||
29 clinical mode from now - scanned field | ||||||||||
30 same field | ||||||||||
31 |