*This setting makes sure that the buffer is filled with a new frame as they come in. If this option is not selected the sensor and software will not be synchronised and we *will record a lot of pitch black images.
*This setting makes sure that the buffer is filled with a new frame as they come in. If this option is not selected the sensor and software will not be synchronised and we *will record a lot of pitch black images.
*Select Capture/Live Options//Live Mode/Live Video: Continuous All buffer capture
*Select Capture/Live Options//Live Mode/Live Video: Continuous All buffer capture
=== DDC232 Instructions ===
*Bash script <code>prog.sh</code>


== Experiment Plan ==
== Experiment Plan ==

Revision as of 15:52, 14 April 2021

Afternoon test at UCLH with photodiodes and CMOS sesnor.

Experiment Equipment

Item Notes
Network Hub Set in control room to take output from experimental room ethernet connection. Control laptops connected via 5GHz WiFi.
Control Laptop x3 2x remote control for sensor and FPGA, 1 for notes/web GUI.
Ethernet Cable x 3 To connect DAQ laptops to network in the experimental room, network hub to network in control room.
Portable Enclosure Modified Big Zarges Waterproof Wheeled Equipment Case.

Features mount for scintillator stack, front and back openings for beam, patch panel with ports for SHV, BNC, SMA, USB, Camera Link cables.
Hole for sensor powering.
Mylar windows with Al support mounted to internal face: approximately light-tight.

Scintillator stack X x 2 mm, Y x 2.6 mm and Z x 3 mm sheets in ascending order from the front/beam of the end of the scintillator. Sheet numbers from front to end (in beam direction): UNKNOWN.
ISDI CMOS sensor sensor pixel dimension: 1030 x 1536. NO optical grease between scintillator and sensor. Connected to DAQ PC via Camera Link cable.
Nikon D7500 Connected to DAQ Laptop via USB cable.
DAQ laptop x2 Control sensor and photodiode acquisition.
Nexys Video FPGA development board. For interfacing between DDC232 and PC.
Rev. 2 Texas Instruments DDC232 custom circuit board (x2) Each housing 16x Hamamatsu S12915-16R photodiodes, coupled directly to scintillator sheets.
Gloves For handling scintillator
Digital Caliper For measuring PMMA air gaps

Range Detector Experiment

HITApr2019_setup.png

ISDI Sensor Instructions

General Sensor Configuration

  • Base Configuration is 14 x 1 Bit
  • Resolution is 1030 x 1536 pixels
  • For selecting pixel rows in centre: Use horizontal offset (ignore DVAL to activate field) and horizontal resolution

Increase number of frame buffers

  • To increase the number of frame buffers to the allowed maximum, start XCAP as administrator.
  • Select Capture/Sequence Capture/Video to Frame Buffers/Driver Assistant
  • Tick box in left field, then set number of frame buffers here.

Serial Terminal instructions for ISDI sensor

  • Find serial terminal in PIXCI(R) tab of sensor control software.
  • In Controls/Options: Make sure "Send string with CR" is selected. Should be default.
  • In Controls/Setup: Select Serial Port: Enabled. Baud rate has to be 115200
  • Set low full well mode using W000300000
  • Set high full well mode using W000300004
  • Read full well mode using R0003

Continuous Field buffer setting

  • This setting makes sure that the buffer is filled with a new frame as they come in. If this option is not selected the sensor and software will not be synchronised and we *will record a lot of pitch black images.
  • Select Capture/Live Options//Live Mode/Live Video: Continuous All buffer capture

DDC232 Instructions

  • Bash script prog.sh

Experiment Plan

  • Shoot-through measurements (back & front) for light calibration and level setting.
  • Range measurements between 70-120 MeV with 1mm steps with CMOS sensor and photodiodes.
  • High intensity range measurements between 70-120 MeV with 1mm steps with photodiodes at optimal light level.
  • Live measurement of 3cm x 3cm x 3cm treatment plan.

Experiment Notes

Range calorimeter measurements with ISDI sensor + photodiodes.

Images from sensor: TBD

Photodiode hexadecimal data: TBD

  • full sensor area
  • shooting on BACK WINDOW
Run number Full Well Mode Beam Energy (MeV/u) Range (estimated) N particles (Hz) Spot size (mm) DDC232 FSR (pC) DDC232 Integration Time (us) Degrader WET (mm) Comments Thumbnail
00 high None - None None 350 170 0 background