Proton Calorimetry/Experimental Runs/2018/Nov16-18: Difference between revisions

From PBTWiki
Jump to navigation Jump to search
Line 12: Line 12:
! style="text-align: center;" | Notes
! style="text-align: center;" | Notes
|-
|-
| style="text-align: center;" | Portable Enclosure
| style="text-align: center;" | Modified Big Zarges Waterproof Wheeled Equipment Case.
Features mount for scintillator and PMT, opening for beam, and ports for SHV, BNC, SMA, Camera Link cables. <br/>
Hole for sensor powering. <br/>
Mylar window with Al support mounted to internal face: approximately light-tight.
|-
| style="text-align: center;" | Scintillator stack
| style="text-align: center;" | 15 x 2 mm, 15 x 2.6 mm and 1 x 3 mm sheets in ascending order from the back of the scintillator to the front/beam. Except for sheet number 21 (2mm) which is placed at the end (away end from the beam) of the stack because of a crack in the sheet. Sheet numbers from front to end (in beam direction): 31,25,24,23,20,16,12,11,9,8,6,5,4,3,2,1,30,29,28,27,26,22,19,18,17,15,14,13,10,7,21.
|-
| style="text-align: center;" | ISDI CMOS sensor
| style="text-align: center;" | sensor pixel dimension: 1030 x 1536. NO optical grease between scintillator and sensor. Connected to DAQ PC via Camera Link cable.
|-
| style="text-align: center;" | DAQ desktop PC
| style="text-align: center;" | Controls sensor aquisition.
|-
| style="text-align: center;" | Gloves
| style="text-align: center;" | For handling scintillator
|}
|}


Line 30: Line 47:
|-
|-
| style="text-align: center;" | Portable Enclosure
| style="text-align: center;" | Portable Enclosure
| style="text-align: center;" | Modified Zarges 1510 Waterproof Wheeled Equipment Case. <br/>
| style="text-align: center;" | Modified Big Zarges Waterproof Wheeled Equipment Case.  
230 x 555 x 350mm <br/>
Features mount for scintillator and PMT, opening for beam, and ports for SHV, BNC, SMA, Camera Link cables. <br/>
Features mount for scintillator and PMT, opening for beam, and ports for SHV, BNC, and SMA cables. <br/>
Hole for sensor powering. <br/>
Mylar window with Al support mounted to internal face: approximately light-tight. <br/>
Mylar window with Al support mounted to internal face: approximately light-tight. <br/>
BNC output from case connected to SMA splitter.
BNC output from case connected to BNC to SMA adaptor.
|-
|-
| style="text-align: center;" | Caen DT5751 Digitiser
| style="text-align: center;" | Caen DT5751 Digitiser
| style="text-align: center;" | Records PMT output, connected via SMA-MCX conversion. Used channel 3.
| style="text-align: center;" | Records PMT output, connected via SMA-MCX conversion. Used channel 3.
|-
| style="text-align: center;" | DAQ desktop PC
| style="text-align: center;" | Controls Caen HV supply; records data from Caen digitiser.
|-
|-
| style="text-align: center;" | Caen NDT1470 HV Supply
| style="text-align: center;" | Caen NDT1470 HV Supply
Line 48: Line 62:
| style="text-align: center;" | Caen units to DAQ PC
| style="text-align: center;" | Caen units to DAQ PC
|-
|-
| style="text-align: center;" | Male-to-Male SMA Cable x2
| style="text-align: center;" | DAQ desktop PC
| style="text-align: center;" | Signal from enclosure port to digitiser. Short/Light and Long cables connected trough SMA to SMA connector
| style="text-align: center;" | Controls Caen HV supply; records data from Caen digitiser.
|-
| style="text-align: center;" | Male-to-Male SMA Cable x3
| style="text-align: center;" | Signal from enclosure port to digitiser. Short/Light and Long cables connected via SMA to SMA connector.<br/>
If scope used: SMA splitter + SMA cable + SMA to BNC adaptor used to split the input signal and send it both to scope and digitiser.
| style="text-align: center;" | DC-4200 MHz Signal Splitter
| style="text-align: center;" |Splits SMA signal to two SMA signals – sent to Oscilloscope and Digitiser.
|-
|-
| style="text-align: center;" | Female-to-Female SMA Connector
| style="text-align: center;" | Female-to-Female SMA Connector
| style="text-align: center;" | Connects Long SMA cable to Short/Light SMA cable.
| style="text-align: center;" | Connects Long SMA cable to Short/Light SMA cable.
|-
| style="text-align: center;" | Female-to-Female BNC to SMA Connector x2
| style="text-align: center;" | From case to SMA cable (or splitter); From SMA cable to scope.
|-
|-
| style="text-align: center;" | Male SMA to Male MCX Adaptor
| style="text-align: center;" | Male SMA to Male MCX Adaptor
Line 72: Line 95:
! style="text-align: center;" | Notes
! style="text-align: center;" | Notes
|-
|-
| style="text-align: center;" | Portable Enclosure
| style="text-align: center;" | Network Hub
| style="text-align: center;" | Modified Zarges 1510 Waterproof Wheeled Equipment Case. <br/>
| style="text-align: center;" | Set in control room to take output from experimental room ethernet switcher. Control laptops connected via ethernet or 5GHz WiFi.
230 x 555 x 350mm <br/>
Features mount for scintillator and PMT, opening for beam, and ports for SHV, BNC, and SMA cables. <br/>
Mylar window with Al support mounted to internal face: approximately light-tight. <br/>
BNC output from case connected to SMA splitter.
|-
|-
| style="text-align: center;" | DAQ desktop PC
| style="text-align: center;" | Control Laptop x2
| style="text-align: center;" |  
| style="text-align: center;" |  
|-
|-
| style="text-align: center;" | Control Laptop
| style="text-align: center;" | DAQ desktop PC
| style="text-align: center;" |  
| style="text-align: center;" |  
|-
|-
Line 88: Line 107:
| style="text-align: center;" | Set in experiment room and connected DAQ desktop PC. Output sent to control room Network Hub.
| style="text-align: center;" | Set in experiment room and connected DAQ desktop PC. Output sent to control room Network Hub.
|-
|-
| style="text-align: center;" | Network Hub
| style="text-align: center;" | Ethernet Cable x 5
| style="text-align: center;" | Set in control room to take output from ethernet switcher. Control laptops connected via ethernet or 5GHz WiFi.
| style="text-align: center;" | To connect DAQ PC to switcher, switcher to Network Hub (long cable), 2 laptops to Network Hub, if scope: scope to switcher.
|-
| style="text-align: center;" | Ethernet Cable x 2
| style="text-align: center;" | To connect DAQ PC to switcher, switcher to Network Hub.
|}
|}



Revision as of 12:16, 14 November 2018

2 night shifts with range calorimeter and single PMT module.

Equipment List

Range Detector Experiment

RangeExperiment.png

Item Notes
Portable Enclosure Modified Big Zarges Waterproof Wheeled Equipment Case.

Features mount for scintillator and PMT, opening for beam, and ports for SHV, BNC, SMA, Camera Link cables.
Hole for sensor powering.
Mylar window with Al support mounted to internal face: approximately light-tight.

Scintillator stack 15 x 2 mm, 15 x 2.6 mm and 1 x 3 mm sheets in ascending order from the back of the scintillator to the front/beam. Except for sheet number 21 (2mm) which is placed at the end (away end from the beam) of the stack because of a crack in the sheet. Sheet numbers from front to end (in beam direction): 31,25,24,23,20,16,12,11,9,8,6,5,4,3,2,1,30,29,28,27,26,22,19,18,17,15,14,13,10,7,21.
ISDI CMOS sensor sensor pixel dimension: 1030 x 1536. NO optical grease between scintillator and sensor. Connected to DAQ PC via Camera Link cable.
DAQ desktop PC Controls sensor aquisition.
Gloves For handling scintillator

Single Module Experiment

SingleModuleExperiment.png

Item Notes
Single Module Scintillator Blocks x5 1cm/3cm/5cm x 10cm x 10cm standard scintillator block. Optical gel required to couple Scintillator to PMT. 5cm and 3cm with black wrapping. 5cm, 3cm, 1cm mylar wrapping.
Photomultiplier Tube Hamamatsu R13089 2"
Portable Enclosure Modified Big Zarges Waterproof Wheeled Equipment Case.

Features mount for scintillator and PMT, opening for beam, and ports for SHV, BNC, SMA, Camera Link cables.
Hole for sensor powering.
Mylar window with Al support mounted to internal face: approximately light-tight.
BNC output from case connected to BNC to SMA adaptor.

Caen DT5751 Digitiser Records PMT output, connected via SMA-MCX conversion. Used channel 3.
Caen NDT1470 HV Supply Supplies HV to PMT
USB Cable Caen units to DAQ PC
DAQ desktop PC Controls Caen HV supply; records data from Caen digitiser.
Male-to-Male SMA Cable x3 Signal from enclosure port to digitiser. Short/Light and Long cables connected via SMA to SMA connector.

If scope used: SMA splitter + SMA cable + SMA to BNC adaptor used to split the input signal and send it both to scope and digitiser.

DC-4200 MHz Signal Splitter Splits SMA signal to two SMA signals – sent to Oscilloscope and Digitiser.
Female-to-Female SMA Connector Connects Long SMA cable to Short/Light SMA cable.
Female-to-Female BNC to SMA Connector x2 From case to SMA cable (or splitter); From SMA cable to scope.
Male SMA to Male MCX Adaptor Sends output from SMA cable to Digitiser
Gloves For handling scintillator
Optical gel For coupling scintillator to PMT
Wipes For removing optical gel

Common Items

Item Notes
Network Hub Set in control room to take output from experimental room ethernet switcher. Control laptops connected via ethernet or 5GHz WiFi.
Control Laptop x2
DAQ desktop PC
Ethernet Switcher Set in experiment room and connected DAQ desktop PC. Output sent to control room Network Hub.
Ethernet Cable x 5 To connect DAQ PC to switcher, switcher to Network Hub (long cable), 2 laptops to Network Hub, if scope: scope to switcher.

Experiment List

16–17th November

ISDI sensor + scintillator stack

17–18th November

Single module

Experiment Log

Beam Tests

16–17th November

Run Number High Voltage (V) Trigger Level (mV) Trigger Source Beam Current / Rate Collimator Filename Notes