Proton Calorimetry/Experimental Runs/2018/Mar23: Difference between revisions
Line 26: | Line 26: | ||
!PMT Current, uA | !PMT Current, uA | ||
!Beam Intensity | !Beam Intensity | ||
!Rate (Measured on Digitiser) | !Rate (Measured on Digitiser) and Notes | ||
!Digitiser Settings | !Digitiser Settings | ||
|- | |- | ||
Line 34: | Line 34: | ||
| 001 || -900 || 150.1 || Lowest || 0.3 - 5 kHz || DC Offset: -30, Threshold: 40 LSB, Gate Offset: 20 ns, Long Gate: 150 ns | | 001 || -900 || 150.1 || Lowest || 0.3 - 5 kHz || DC Offset: -30, Threshold: 40 LSB, Gate Offset: 20 ns, Long Gate: 150 ns | ||
|- | |||
| 002 || -800 || 133.2 || Lowest || 0.3 - 5 kHz || As above | |||
|- | |||
| 003 || -700 || 116.35 || Lowest || 0.3 - 5 kHz || As above | |||
|- | |||
| 004 || -600 || 99.45 || Lowest || 0.3 - 5 kHz || Beam dropped during running! | |||
|- | |||
| 005 || -600 || 99.45 || Lowest || 0.3 - 5 kHz || Repeat run with beam back. | |||
|- | |||
| 006 || -500 || 82.6 || Lowest || 0.3 - 5 kHz || Threshold: 10 LSB. Too close to the threshold, but cannot lower more due to noise floor. | |||
|- | |- |
Revision as of 10:42, 28 March 2018
Beam tests of the UCL single module at medAustron.
Friday 23rd - Saturday 24th of March 2018
Pre-data Taking Information and Notes
- The tests were carried out with an energy of 252.7 MeV with a 5 second beam spill and a range of intensities.
- No collimators were used at any point.
- The setup consisted of:
- 2" Hamamatsu R13089-100-11 PMT with negative HV active divider base (made by Hamamatsu)
- coupled with Eljen optical gel to a
- 3 cm x 3 cm x 45 cm cuboid ENVINET standard scintillator wrapped in two layers 6 um (each) of Mylar
- The "pensioner" 5 dynode stage PMT was also briefly tested, but no sensible data was able to be acquired.
- The digitiser always operated in Self Trigger mode.
- Differences since last medAustron test beam:
- Short cables used for signals and SMA instead of BNC, controlling the acquisition remotely.
- Tests carried out at a range of HVs.
- Working directory where digitiser data is stored:
C:\Users\abf\Desktop\digitiser_data\medAustron\mar18\ham
Data Taking
Run number | High Voltage, V | PMT Current, uA | Beam Intensity | Rate (Measured on Digitiser) and Notes | Digitiser Settings |
---|---|---|---|---|---|
N/A | -900 | 150.1 (beam off) | N/A | N/A | DC Offset: 0, Rec. Length: 396 ns, Threshold: 40 LSB, Gate Offset: 1 ns, Long Gate: 396 ns, BL Mean: 8 ns, Charge. Sensitivty: 20 fC/LSB, Pre-trigger: 9 ns |
001 | -900 | 150.1 | Lowest | 0.3 - 5 kHz | DC Offset: -30, Threshold: 40 LSB, Gate Offset: 20 ns, Long Gate: 150 ns |
002 | -800 | 133.2 | Lowest | 0.3 - 5 kHz | As above |
003 | -700 | 116.35 | Lowest | 0.3 - 5 kHz | As above |
004 | -600 | 99.45 | Lowest | 0.3 - 5 kHz | Beam dropped during running! |
005 | -600 | 99.45 | Lowest | 0.3 - 5 kHz | Repeat run with beam back. |
006 | -500 | 82.6 | Lowest | 0.3 - 5 kHz | Threshold: 10 LSB. Too close to the threshold, but cannot lower more due to noise floor. |