Radiotherapy/Experimental Runs/2022-07-26: Difference between revisions
Jump to navigation
Jump to search
Second run
Replaced plastic blocks from first run with new un-irradiated plastic blocks, but using same green scintillator sheet.
JeremyOcampo (talk | contribs) No edit summary |
JeremyOcampo (talk | contribs) |
||
(One intermediate revision by the same user not shown) | |||
Line 20: | Line 20: | ||
=== Pre-data Taking Information and Notes === | === Pre-data Taking Information and Notes === | ||
* Used 200mmx200mmx2mm green scintillator sheet polystyrene base sandwiched by two 200mmx200mmx50mm polystyrene clear plastic blocks | |||
* 53cm from lens to sheet | * 53cm from lens to sheet | ||
* Check camera in focus with ruler on the sheet | * Check camera in focus with ruler on the sheet |
Latest revision as of 16:58, 19 August 2022
Second experimental measurements with Cyberknife at Birmingham. Green scintillator thin sheet measurements only.
Setup
Schematic
Photos of setup
(Inside pelicase, outside pelicase, collimators, through pelicase window)
Pre-data Taking Information and Notes
- Used 200mmx200mmx2mm green scintillator sheet polystyrene base sandwiched by two 200mmx200mmx50mm polystyrene clear plastic blocks
- 53cm from lens to sheet
- Check camera in focus with ruler on the sheet
- Near/Far limit of focus 49.4cm 57.1cm calculated using https://dofmaster.com/dofjs.html
- Pelicase on couch
- Scintillator blocks aligned to laser cross hairs
- 785 mm source to surface distance suitable for output factor/PDD/profile measurements.
- Room lights out and pelicase window open.
- Check optical axis aligned with top surface using crosshair in the image analysis.
- Arrived at hospital at 11am, finished data taking at 5pm
Data Taking
- All images taken with Nikon D7000, 20mm lens, f/2.8, 400ISO
- Used camera control pro 2 app
- Image of ruler taken at sheet for pixels to mm conversion
- 5 images taken at each collimator size with 1 background image each.
- Change exposure time until near max saturation - get's best bit depth
- View data analysis here
First run
With scintillator sheet | |||
---|---|---|---|
Collimator Size (mm) | Exposure (s) | Image | Comments |
5 | 30 | ||
7.5 | 20 | ||
10 | 15 | ||
12.5 | 10 | ||
15 | 8 | ||
20 | 8 | ||
25 | 6 | ||
30 | 6 | ||
35 | 5 | ||
40 | 4 | ||
50 | 3 | ||
60 | 3 |
No scintillator sheet | |||
---|---|---|---|
Collimator Size (mm) | Exposure (s) | Image | Comments |
60 | 3 | Can see green discolouration caused by previous irradiations. Blocks must have been rotated by 90 when re-assembling |
Second run
Replaced plastic blocks from first run with new un-irradiated plastic blocks, but using same green scintillator sheet.
With scintillator sheet | |||
---|---|---|---|
Collimator Size (mm) | Exposure (s) | Image | Comments |
5 | 30 | ||
7.5 | 20 | ||
10 | 15 | ||
12.5 | 10 | ||
15 | 8 | ||
20 | 8 | ||
25 | 6 | ||
30 | 6 | ||
35 | 5 | ||
40 | 4 | ||
50 | 3 | ||
60 | 3 |
Green discolouration probably from scintillator sheet. |
No scintillator sheet | |||
---|---|---|---|
Collimator Size (mm) | Exposure (s) | Image | Comments |
5 | 30 | ||
7.5 | 20 | ||
10 | 15 | ||
12.5 | 10 | ||
15 | 8 | ||
20 | 8 | ||
25 | 6 | ||
30 | 6 | ||
35 | 5 | ||
40 | 4 | ||
50 | 3 | ||
60 | 3 |