=Day 1 (December 21st)=
====Notes====
* Current at -1200 V: 200.9 μA (beam off)
*Some double peaks and very large signals present - noise of some sort?
*Background rate (-1200 V, beam off, 3 mm collimator): ~4 kHz
**Threshold: -19.5 mV
**Rise time: 2.5 ns (without averaging), 2.9 ns over 10 sweeps
**Amplitude: ~75 mV over 10 sweeps
*Having the corridor lights on does NOT create enough light to trip the PMT and is acceptable! (Current goes up by 0.1 μA)
*Peak current estimate at -1100 V: 44 mA
**Amplitude: 2.2 V
**Width: ~20 ns
===Hamamatsu Base + Enhanced Cuboid Tests===
{|class="wikitable"
!High Voltage (V)
!Trigger
!Rate + Additional Information
!Electronics Setup
!Collimator Size + Additional Information
!Run Number
|-
| -1200 || ST
|
*Rate: ~4 kH
|
*Rec. Length: 456 ns
*BL Mean: 32
*Charge sens.: 20 pC/LSB
*Threshold: 40 LSB
*Gate offset: 20 ns
*Short gate: 50 ns
*Long gate: 100 ns
*DC Offset: 0
*Pre-trigger: 48 ns
| 3 mm
*Background measurement (beam off)
| 001 (4,462 events)
|-
| -1200 || ST || I: Tripped || || 3mm ||
|-
| -800 || ST
|
*Amplitude: 300 mV (-194 mV threshold)
*I: 152.5 μA
*Rate: ~500 kHz
| || ||
|-
| -800 || ST
|
*I: 133.4 μA
*Rate: ~4 kHz
*Unstable beam
| Long gate: 100 ns ||
*Gas off (leakage)
*0 arc current (±0.1 μA)
*0.1 dose
| 002 (845,794 events)
|-
| -900 || ST || I: 150.3 μA || || Edge on peak || 003-004
|-
| -1000 || ST || I: 167.3 μA || || || 005
|-
| -1100 || ST || I: 184.1 μA || || Peak shifted during running || 006
|-
| -1100 || ST ||
*Rate: ~2 kHz - 25 kHz
*Unstable
| Long gate: 200 ns || 2 mm
*gas off
*0 arc current
| 007 (~3,000 events in peaks)
|-
| -1100 || ST ||
*Rate ~19 kHz
*I: 184.1 μA
*Stable rate
| Long gate: 200 ns
|
*0.2 cc/minute
*Arc current ~0
| 008 (6,000 events in peak)
|-
| -1100 || ST || || Long gate: 250 ns || || 009
|-
| -1100 || ST || || Long gate: 150 ns || || 010
|-
| -1100 || ST || || Long gate: 100 ns || || 011
|-
| -1100 || ST || || Long gate: 200 ns (irrelevant) || Waveforms || 012-14
|-
| -1100 || ST ||
*Rate: 14 kHz
*I: 184 μA
*Beam dropped due to preset time running out
|
*PMMA is in range shifter box, upstream
*Long gate: 200 ns
| 2 mm
*+ 5.07 mm of PMMA (1 plate)
| 015 (600,000 events)
|-
| -1100 || ST || || Long gate: 200 ns || Repeat above with larger stats || 016 (1,000,000 events)
|-
| -1100 || ST || || Long gate: 100 ns || || 017 (1,000,000 events)
|-
| -1100 || ST || || Long gate: 100 ns || Waveforms (~2 GB) || 019
|-
| -1100 || ST ||
*Rate: 1.3 kHz
*I: 183.99 μA
| Long gate: 200 ns || 9.8 mm of PMMA (1 plate) || 020
|-
| -1100 || ST || || Long gate: 100 ns || || 021
|-
| -1100 || ST || || Long gate: 100 ns || Waveforms (~2 GB) || 023
|-
| -1100 || ST || Rate: 30 kHz || Long gate: 200 ns || Upped current at 0.1 mA || 025 (300,000 events)
|-
| -1100 || ST || || Long gate: 100 ns || || 026 (300,000 events)
|-
| -1100 || ST || || Long gate: 100 ns || Waveforms (~2 GB) || 027
|-
| -1100 || ST || || Long gate: 200 ns || 17.94 mm of PMMA
*9.04 mm + 8.9 mm, 2 plates
| 028 (400,000 events)
|-
| -1100 || ST || || Long gate: 100 ns || Waveforms (~2 GB) || 031
|-
| -1100 || ST || || Long gate: 200 ns || 19.94 mm of PMMA
*9.04 mm + 10.9 mm, 2 plates
| 032 (400,000 events)
|-
| -1100 || ST || || Long gate: 100 ns || || 033 (300,000 events)
|-
| -1100 || ST || I: 184 μA || Long gate: 100 ns || || 035 (400,000 events)
|-
| -1100 || ST ||
*Rate: 150 kHz
*I: 184.9 μA
| Long gate: 200 ns || No PMMA || 036 (2,000,000 events)
|-
| -1100 || ST || || Long gate: 100 ns || || 037 (2,000,000 events)
|-
| -1100 || ST || || Long gate: 100 ns || Waveforms (~2 GB) || 039
|-
| -1100 || ST || || Long gate: 50 ns || || 040 (2,000,000 events)
|-
| -1100 || ST || || Long gate: 60 ns || || 041 (1,500,000 events)
|-
| -1100 || ST || || Long gate: 70 ns || || 042 (1,500,000 events)
|-
| -1100 || ST || || Long gate: 80 ns || || 043 (1,500,000 events)
|-
| -1100 || ST || || Long gate: 90 ns || || 044 (1,500,000 events)
|-
|colspan="6"|Break for lunch
|-
| -1100 || ST ||
*Rate: 200 kHz
*I: 185.3 μA
| Long gate: 200 ns || || 045 (2,000,000 events)
|-
| -1100 || ST || || Long gate: 100 ns || || 046 (2,000,000 events)
|-
| -1100 || ST || || Long gate: 50 ns || || 047 (2,000,000 events)
|-
| -1100 || ST || || Long gate: 50 ns || Waveforms (~2 GB) || 049
|-
| -1100 || ST ||
*Rate: 400 kHz
*I: 209 μA
| Long gate: 200 ns || || 050 (2,000,000 events)
|-
| -1100 || ST || || Long gate: 100 ns || || 051 (2,000,000 evvents)
|-
| -1100 || ST || || Long gate: 50 ns || || 052 (2,000,000 events)
|-
| -1100 || ST || || Long gate: 30 ns || || 053 (3,000,000 events)
|-
| -1100 || ST || ||
*Long gate: 20 ns
*Short gate: 19 ns
| || 054 (2,000,000 events)
|-
| -1100 || ST || Note: Intrinsic spread in energy during running for ~1 s ||
*Long gate: 20 ns
*Short gate: 19 ns
| Waveforms (~2 GB) || 056
|-
| -1100 || ST || ||
*Gate offset: 1 ns
*Long gate: 30 ns
| || 057 (2,500,000 events)
|-
| -1100 || ST || || Long gate: 20 ns || || 059 (2,000,000 events)
|-
| -1100 || ST ||
*Rate: 450-550 kHz
*I: 211.6 μA
|
*Gate offset: 18 ns
*Long gate: 100 ns
| || 060 (4,000,000 events)
|-
| -1100 || ST || ||
*Gate offset: 10 ns
*Long gate: 50 ns
| || 061 (4,000,000 events)
|-
| -1100 || ST || ||
*Gate offste: 1 ns
*Long gate: 30 ns
| || 062 (3,000,000 events)
|-
| -1100 || ST || ||
*Gate offset: 1 ns
*Long gate: 20 ns
| || 063 (3,000,000 events)
|-
| -1100 || ST ||
*I: 211.9 μA
*Rate: 450-500 kHz
|
*Gate offset: 18 ns
*Long gate: 100 ns
| Waveforms (~3 GB) || 066
|}
=Day 2 (December 22nd)=
====Notes====
*Current at -900 V: 150.2 μA (beam off)
*Rate at -900 V: 400 kHz (threshold: -194 mV)
===Hamamatsu Base + Enhanced Cuboid Tests===
{|class="wikitable"
!High Voltage (V)
!Trigger
!Rate + Additional Information
!Electronics Setup
!Collimator Size + Additional Information
!Run Number
|-
| -900 || ST ||
*Rate: 30 kHz
*Peak ~7000 ADC
*I: 150.2 μA
|
*Long gate: 100 ns
*Gate offset: 18 ns
| || 002 (1,000,000 events)
|-
| -950 || ST ||
*I: 158.8 μA
*Rate: 30 kHz
*Peak ~9000 ADC
| Long gate: 100 ns || || 003 (3,000,000 events)
|-
| -950 || ST || || Long gate: 200 ns || || 004 (2,000,000 events)
|-
| -950 || ST || || Long gate: 50 ns || || 005 (1,000,000 events)
|-
| -950 || ST || || Long gate: 200 ns || Waveforms (~2 GB) || 007
|-
| -950 || ST ||
*Rate: 60-50 kHz
| Long gate: 100 ns || || 008 (2,000,000 events)
|-
| -950 || ST ||
*Rate: 50-40 kHz
*Current on beam dropped during run, creating shoulder
| Long gate: 200 ns || 5.07 mm of PMMA (1 plate)
*wd: scan
| 009 (2,000,000 events)
|-
| -950 || ST || Rate: 70 kHz || Long gate: 200 ns || || 010 (2,000,000 events)
|-
| -950 || ST || || Long gate: 100 ns || || 011 (2,000,000 events)
|-
| -950 || ST || || Long gate: 200 ns || 9.8 mm of PMMA (1 plate) || 012 (2,000,000 events)
|-
| -950 || ST || || Long gate: 100 ns || || 013 (2,000,000 events)
|-
| -950 || ST || || Long gate: 200 ns || 14.12 mm of PMMA (1 plate) || 014 (2,000,000 events)
|-
| -950 || ST || || Long gate: 100 ns || || 015 (2,000,000 events)
|-
| -950 || ST || || Long gate: 200 ns || 17.94 mm of PMMA
*9.04 mm + 8.9 mm, 2 plates
| 016 (2,000,000 events)
|-
| -950 || ST || || Long gate: 100 ns || || 017 (2,000,000 events)
|-
| -950 || ST || I: 158.7 μA || Long gate: 200 ns || 19.94 mm of PMMA
*9.04 mm + 10.9 mm, 2 plates
| 018 (2,000,000 events)
|-
| -950 || ST || || Long gate: 100 ns || || 019 (2,000,000 events)
|-
| -950 || ST || Beam dropped || Long gate: 200 ns || 21.70 mm of PMMA
*9.04 mm + 12.66 mm, 2 plates
| 020 (1,000,000 events)
|-
| -950 || ST || Repeat above run after beam drop || Long gate: 200 ns || || 021 (2,000,000 events)
|-
| -950 || ST || Beam dropped || Long gate: 100 ns || || 022 (900,000 events)
|-
| -950 || ST || Repeat above run after beam drop || Long gate: 100 ns || || 023 (1,500,000 events)
|-
| -950 || ST || || Long gate: 200 ns || 0 mm PMMA || 024 (2,000,000 events)
|-
| -950 || ST || || Long gate: 100 ns || || 025 (2,000,000 events)
|}
===Beam tests===
{|class="wikitable"
!High Voltage (V)
!Trigger
!Rate + Additional Information
!Electronics Setup
!Collimator Size + Additional Information
!Run Number
|-
| -950 || ST ||
*Rate: ~700 kHz
*Amplitude has dropped
| Long gate: 100 ns ||
*1.3 mA current
*0.2 cc/minute gas
*wd: tests
| 026 (1,500,000 events)
|-
| -950 || ST || ||
*Long gate: 50 ns
*Gate offset: 10 ns
| Focusing angle: 69.1° || 027 (2,000,000 events)
|-
| -950 || ST || ||
*Long gate: 30 ns
*Gate offset: 7 ns
| || 028 (2,000,000 events)
|-
| -950 || ST || ||
*Long gate: 200 ns
*Gate offset: 18 ns
| Waveforms (~2 GB) || 030
|-
| -950 || ST ||
*Rate: 600 kHz
*I: 183 μA
|
*Long gate: 60 ns
*Gate offset: 10 ns
|
*1.01 mA
*4.9 dose rate
*315 pA
| 031 (2,000,000 events)
|-
| -950 || ST || ||
*Long gate: 60 ns
*Gate offset: 10 ns
| Waveforms (~3.5 GB) || 033
|-
| -950 || ST ||
*Rate: ~700 kHz (from extrapolation)
*"Standard" run
|
*Long gate: 100 ns
*Gate offset: 18 ns
|
*Arc current: 104 mA
*Scattering: 0.3 nA
*Dose rate: 0.6
*Angle: 64.1°
| 034 (2,000,000 events)
|-
| -950 || ST || Rate: >500 kHz (extrapolated) || ||
*Angle: 65.4°
*Dose: 3.5
*Scattering: 0.22 nA
| 036 (2,000,000 events)
|-
| -950 || ST ||
*Rate: ~400 kHz (extrapolated)
*Higher amplitude
| ||
*Angle: 66.5°
*Dose rate: 2.1
*Scattering: 0.14 nA
| 036 (2,000,000 events)
|-
| -950 || ST ||
*Rate: ~200 kHz (extrapolated)
*Rate: ~200 kHz (measured on scope)
| ||
*Angle: 67.6°
*Dose rate: 1.0
*Scattering: 0.065 nA
| 037 (2,000,000 events)
|-
| -950 || ST || Rate: ~100 kHz (extrapolated) || ||
*Angle: 68.5°
*Dose: 0.5
*Scattering: 0.034 nA
| 038 (2,000,000 events)
|-
| -950 || ST || || ||
*Angle: 69.5°
*Dose: 0.2
*Scattering: 0.014 na
| 039 (2,000,000 events)
|-
| -950 || ST || || ||
*Angle: 71.5°
*Dose: 0-0.1
*Scattering: 0.0033 nA
| 040 (1,000,000 events)
|-
| -950 || ST || Higher amplitude / ADC || ||
*Angle: 73.5°
*Dose: 0-0.1
*Scattering: 0.0006 nA
| 041 (300,000 events)
|-
| -950 || ST || Back to 71.5° to see where ADC peak is - back to normal || Rate (scope):
*~88 ms
*~20 kHz
| Angle: 71.5° || 042 (300,000 events
|-
| -950 || ST || Shifted again || || Angle: 73.5° || 043 (100,000 events)
|-
| -950 || ST || || || Angle: 66.5° || 044 (1,000,000 events)
|-
| -950 || ST || || || Angle: 65.4° || 045 (1,000,000 events)
|-
| -950 || ST || Scope rate: 730 ms
*~1.2 kHz
| || Angle: 73.5° || 047 (50,000 events)
|-
| -950 || ST || Scan across arc current/dose rate to see if there are any shifts in ADC peak || || Beginning:
*Dose: 0.4-0.5
*Scattering: 0.029 nA
Finishing dose: 3
| 048 (scan)
|-
| -950 || ST || Rate: 500 kHz
*2 ms at dose: 3
| At ~600 kHz we get a shift due to frequency cutoff ||
*Left peak: 7109
*Right peak: 8894
| 049 (scan)
|-
| -950 || ST || Rate: ~1.2 MHz (extrapolated) || Long gate: 100 ns ||
*Dose: 13
*Arc current: 1.7 mA
*Waveforms (2.7 GB)
| 051
|}

Latest revision as of 16:33, 26 July 2017

Include all details of experimental runs from Anastasia's PDFs.

PAGE UNDER CONSTRUCTION

Day 1 (December 21st)

Notes

  • Current at -1200 V: 200.9 μA (beam off)
  • Some double peaks and very large signals present - noise of some sort?
  • Background rate (-1200 V, beam off, 3 mm collimator): ~4 kHz
    • Threshold: -19.5 mV
    • Rise time: 2.5 ns (without averaging), 2.9 ns over 10 sweeps
    • Amplitude: ~75 mV over 10 sweeps
  • Having the corridor lights on does NOT create enough light to trip the PMT and is acceptable! (Current goes up by 0.1 μA)
  • Peak current estimate at -1100 V: 44 mA
    • Amplitude: 2.2 V
    • Width: ~20 ns

Hamamatsu Base + Enhanced Cuboid Tests

High Voltage (V) Trigger Rate + Additional Information Electronics Setup Collimator Size + Additional Information Run Number
-1200 ST
  • Rate: ~4 kH
  • Rec. Length: 456 ns
  • BL Mean: 32
  • Charge sens.: 20 pC/LSB
  • Threshold: 40 LSB
  • Gate offset: 20 ns
  • Short gate: 50 ns
  • Long gate: 100 ns
  • DC Offset: 0
  • Pre-trigger: 48 ns
3 mm
  • Background measurement (beam off)
001 (4,462 events)
-1200 ST I: Tripped 3mm
-800 ST
  • Amplitude: 300 mV (-194 mV threshold)
  • I: 152.5 μA
  • Rate: ~500 kHz
-800 ST
  • I: 133.4 μA
  • Rate: ~4 kHz
  • Unstable beam
Long gate: 100 ns
  • Gas off (leakage)
  • 0 arc current (±0.1 μA)
  • 0.1 dose
002 (845,794 events)
-900 ST I: 150.3 μA Edge on peak 003-004
-1000 ST I: 167.3 μA 005
-1100 ST I: 184.1 μA Peak shifted during running 006
-1100 ST
  • Rate: ~2 kHz - 25 kHz
  • Unstable
Long gate: 200 ns 2 mm
  • gas off
  • 0 arc current
007 (~3,000 events in peaks)
-1100 ST
  • Rate ~19 kHz
  • I: 184.1 μA
  • Stable rate
Long gate: 200 ns
  • 0.2 cc/minute
  • Arc current ~0
008 (6,000 events in peak)
-1100 ST Long gate: 250 ns 009
-1100 ST Long gate: 150 ns 010
-1100 ST Long gate: 100 ns 011
-1100 ST Long gate: 200 ns (irrelevant) Waveforms 012-14
-1100 ST
  • Rate: 14 kHz
  • I: 184 μA
  • Beam dropped due to preset time running out
  • PMMA is in range shifter box, upstream
  • Long gate: 200 ns
2 mm
  • + 5.07 mm of PMMA (1 plate)
015 (600,000 events)
-1100 ST Long gate: 200 ns Repeat above with larger stats 016 (1,000,000 events)
-1100 ST Long gate: 100 ns 017 (1,000,000 events)
-1100 ST Long gate: 100 ns Waveforms (~2 GB) 019
-1100 ST
  • Rate: 1.3 kHz
  • I: 183.99 μA
Long gate: 200 ns 9.8 mm of PMMA (1 plate) 020
-1100 ST Long gate: 100 ns 021
-1100 ST Long gate: 100 ns Waveforms (~2 GB) 023
-1100 ST Rate: 30 kHz Long gate: 200 ns Upped current at 0.1 mA 025 (300,000 events)
-1100 ST Long gate: 100 ns 026 (300,000 events)
-1100 ST Long gate: 100 ns Waveforms (~2 GB) 027
-1100 ST Long gate: 200 ns 17.94 mm of PMMA
  • 9.04 mm + 8.9 mm, 2 plates
028 (400,000 events)
-1100 ST Long gate: 100 ns Waveforms (~2 GB) 031
-1100 ST Long gate: 200 ns 19.94 mm of PMMA
  • 9.04 mm + 10.9 mm, 2 plates
032 (400,000 events)
-1100 ST Long gate: 100 ns 033 (300,000 events)
-1100 ST I: 184 μA Long gate: 100 ns 035 (400,000 events)
-1100 ST
  • Rate: 150 kHz
  • I: 184.9 μA
Long gate: 200 ns No PMMA 036 (2,000,000 events)
-1100 ST Long gate: 100 ns 037 (2,000,000 events)
-1100 ST Long gate: 100 ns Waveforms (~2 GB) 039
-1100 ST Long gate: 50 ns 040 (2,000,000 events)
-1100 ST Long gate: 60 ns 041 (1,500,000 events)
-1100 ST Long gate: 70 ns 042 (1,500,000 events)
-1100 ST Long gate: 80 ns 043 (1,500,000 events)
-1100 ST Long gate: 90 ns 044 (1,500,000 events)
Break for lunch
-1100 ST
  • Rate: 200 kHz
  • I: 185.3 μA
Long gate: 200 ns 045 (2,000,000 events)
-1100 ST Long gate: 100 ns 046 (2,000,000 events)
-1100 ST Long gate: 50 ns 047 (2,000,000 events)
-1100 ST Long gate: 50 ns Waveforms (~2 GB) 049
-1100 ST
  • Rate: 400 kHz
  • I: 209 μA
Long gate: 200 ns 050 (2,000,000 events)
-1100 ST Long gate: 100 ns 051 (2,000,000 evvents)
-1100 ST Long gate: 50 ns 052 (2,000,000 events)
-1100 ST Long gate: 30 ns 053 (3,000,000 events)
-1100 ST
  • Long gate: 20 ns
  • Short gate: 19 ns
054 (2,000,000 events)
-1100 ST Note: Intrinsic spread in energy during running for ~1 s
  • Long gate: 20 ns
  • Short gate: 19 ns
Waveforms (~2 GB) 056
-1100 ST
  • Gate offset: 1 ns
  • Long gate: 30 ns
057 (2,500,000 events)
-1100 ST Long gate: 20 ns 059 (2,000,000 events)
-1100 ST
  • Rate: 450-550 kHz
  • I: 211.6 μA
  • Gate offset: 18 ns
  • Long gate: 100 ns
060 (4,000,000 events)
-1100 ST
  • Gate offset: 10 ns
  • Long gate: 50 ns
061 (4,000,000 events)
-1100 ST
  • Gate offste: 1 ns
  • Long gate: 30 ns
062 (3,000,000 events)
-1100 ST
  • Gate offset: 1 ns
  • Long gate: 20 ns
063 (3,000,000 events)
-1100 ST
  • I: 211.9 μA
  • Rate: 450-500 kHz
  • Gate offset: 18 ns
  • Long gate: 100 ns
Waveforms (~3 GB) 066

Day 2 (December 22nd)

Notes

  • Current at -900 V: 150.2 μA (beam off)
  • Rate at -900 V: 400 kHz (threshold: -194 mV)

Hamamatsu Base + Enhanced Cuboid Tests

High Voltage (V) Trigger Rate + Additional Information Electronics Setup Collimator Size + Additional Information Run Number
-900 ST
  • Rate: 30 kHz
  • Peak ~7000 ADC
  • I: 150.2 μA
  • Long gate: 100 ns
  • Gate offset: 18 ns
002 (1,000,000 events)
-950 ST
  • I: 158.8 μA
  • Rate: 30 kHz
  • Peak ~9000 ADC
Long gate: 100 ns 003 (3,000,000 events)
-950 ST Long gate: 200 ns 004 (2,000,000 events)
-950 ST Long gate: 50 ns 005 (1,000,000 events)
-950 ST Long gate: 200 ns Waveforms (~2 GB) 007
-950 ST
  • Rate: 60-50 kHz
Long gate: 100 ns 008 (2,000,000 events)
-950 ST
  • Rate: 50-40 kHz
  • Current on beam dropped during run, creating shoulder
Long gate: 200 ns 5.07 mm of PMMA (1 plate)
  • wd: scan
009 (2,000,000 events)
-950 ST Rate: 70 kHz Long gate: 200 ns 010 (2,000,000 events)
-950 ST Long gate: 100 ns 011 (2,000,000 events)
-950 ST Long gate: 200 ns 9.8 mm of PMMA (1 plate) 012 (2,000,000 events)
-950 ST Long gate: 100 ns 013 (2,000,000 events)
-950 ST Long gate: 200 ns 14.12 mm of PMMA (1 plate) 014 (2,000,000 events)
-950 ST Long gate: 100 ns 015 (2,000,000 events)
-950 ST Long gate: 200 ns 17.94 mm of PMMA
  • 9.04 mm + 8.9 mm, 2 plates
016 (2,000,000 events)
-950 ST Long gate: 100 ns 017 (2,000,000 events)
-950 ST I: 158.7 μA Long gate: 200 ns 19.94 mm of PMMA
  • 9.04 mm + 10.9 mm, 2 plates
018 (2,000,000 events)
-950 ST Long gate: 100 ns 019 (2,000,000 events)
-950 ST Beam dropped Long gate: 200 ns 21.70 mm of PMMA
  • 9.04 mm + 12.66 mm, 2 plates
020 (1,000,000 events)
-950 ST Repeat above run after beam drop Long gate: 200 ns 021 (2,000,000 events)
-950 ST Beam dropped Long gate: 100 ns 022 (900,000 events)
-950 ST Repeat above run after beam drop Long gate: 100 ns 023 (1,500,000 events)
-950 ST Long gate: 200 ns 0 mm PMMA 024 (2,000,000 events)
-950 ST Long gate: 100 ns 025 (2,000,000 events)

Beam tests

High Voltage (V) Trigger Rate + Additional Information Electronics Setup Collimator Size + Additional Information Run Number
-950 ST
  • Rate: ~700 kHz
  • Amplitude has dropped
Long gate: 100 ns
  • 1.3 mA current
  • 0.2 cc/minute gas
  • wd: tests
026 (1,500,000 events)
-950 ST
  • Long gate: 50 ns
  • Gate offset: 10 ns
Focusing angle: 69.1° 027 (2,000,000 events)
-950 ST
  • Long gate: 30 ns
  • Gate offset: 7 ns
028 (2,000,000 events)
-950 ST
  • Long gate: 200 ns
  • Gate offset: 18 ns
Waveforms (~2 GB) 030
-950 ST
  • Rate: 600 kHz
  • I: 183 μA
  • Long gate: 60 ns
  • Gate offset: 10 ns
  • 1.01 mA
  • 4.9 dose rate
  • 315 pA
031 (2,000,000 events)
-950 ST
  • Long gate: 60 ns
  • Gate offset: 10 ns
Waveforms (~3.5 GB) 033
-950 ST
  • Rate: ~700 kHz (from extrapolation)
  • "Standard" run
  • Long gate: 100 ns
  • Gate offset: 18 ns
  • Arc current: 104 mA
  • Scattering: 0.3 nA
  • Dose rate: 0.6
  • Angle: 64.1°
034 (2,000,000 events)
-950 ST Rate: >500 kHz (extrapolated)
  • Angle: 65.4°
  • Dose: 3.5
  • Scattering: 0.22 nA
036 (2,000,000 events)
-950 ST
  • Rate: ~400 kHz (extrapolated)
  • Higher amplitude
  • Angle: 66.5°
  • Dose rate: 2.1
  • Scattering: 0.14 nA
036 (2,000,000 events)
-950 ST
  • Rate: ~200 kHz (extrapolated)
  • Rate: ~200 kHz (measured on scope)
  • Angle: 67.6°
  • Dose rate: 1.0
  • Scattering: 0.065 nA
037 (2,000,000 events)
-950 ST Rate: ~100 kHz (extrapolated)
  • Angle: 68.5°
  • Dose: 0.5
  • Scattering: 0.034 nA
038 (2,000,000 events)
-950 ST
  • Angle: 69.5°
  • Dose: 0.2
  • Scattering: 0.014 na
039 (2,000,000 events)
-950 ST
  • Angle: 71.5°
  • Dose: 0-0.1
  • Scattering: 0.0033 nA
040 (1,000,000 events)
-950 ST Higher amplitude / ADC
  • Angle: 73.5°
  • Dose: 0-0.1
  • Scattering: 0.0006 nA
041 (300,000 events)
-950 ST Back to 71.5° to see where ADC peak is - back to normal Rate (scope):
  • ~88 ms
  • ~20 kHz
Angle: 71.5° 042 (300,000 events
-950 ST Shifted again Angle: 73.5° 043 (100,000 events)
-950 ST Angle: 66.5° 044 (1,000,000 events)
-950 ST Angle: 65.4° 045 (1,000,000 events)
-950 ST Scope rate: 730 ms
  • ~1.2 kHz
Angle: 73.5° 047 (50,000 events)
-950 ST Scan across arc current/dose rate to see if there are any shifts in ADC peak Beginning:
  • Dose: 0.4-0.5
  • Scattering: 0.029 nA

Finishing dose: 3

048 (scan)
-950 ST Rate: 500 kHz
  • 2 ms at dose: 3
At ~600 kHz we get a shift due to frequency cutoff
  • Left peak: 7109
  • Right peak: 8894
049 (scan)
-950 ST Rate: ~1.2 MHz (extrapolated) Long gate: 100 ns
  • Dose: 13
  • Arc current: 1.7 mA
  • Waveforms (2.7 GB)
051