Proton Calorimetry/Experimental Runs/2018/Nov16-18: Difference between revisions
(2 intermediate revisions by the same user not shown) | |||
Line 424: | Line 424: | ||
|80 || high || 221.06 || 308.1 || 3.2x10^9 || 8.1 || p || 50mm PMMA absorber upstream, shoot-through || - | | |80 || high || 221.06 || 308.1 || 3.2x10^9 || 8.1 || p || 50mm PMMA absorber upstream, shoot-through || - | | ||
|- | |- | ||
|81 || high || 48. | |81 || high || 48.12 || 20.8 || 3.2x10^9 || 32.4 || p || low energy, low intensity || - | | ||
|- | |- | ||
|82 || high || 121.95 || 109.7 || 1.2x10^9 || 13.2 || p || range scan || - | | |82 || high || 121.95 || 109.7 || 1.2x10^9 || 13.2 || p || range scan || - | | ||
Line 434: | Line 434: | ||
|85 || high || 101.9 || 79.8 || 1.2x10^9 || 15.5 || p || range scan || - | | |85 || high || 101.9 || 79.8 || 1.2x10^9 || 15.5 || p || range scan || - | | ||
|- | |- | ||
|86 || high || 94. | |86 || high || 94.54 || 69.9 || 1.2x10^9 || 16.6 || p || range scan || - | | ||
|- | |- | ||
|87 || high || 86.72 || 59.9 || 1.2x10^9 || 18.2 || p || range scan || - | | |87 || high || 86.72 || 59.9 || 1.2x10^9 || 18.2 || p || range scan || - | | ||
Line 462: | Line 462: | ||
|97 || high || 56.49 || 27.8 || 1.2x10^9 || 28.3 || p || range scan 1mm scan || - | | |97 || high || 56.49 || 27.8 || 1.2x10^9 || 28.3 || p || range scan 1mm scan || - | | ||
|- | |- | ||
|98 || high || | |98 || high || 55.34 || 26.8 || 1.2x10^9 || 28.9 || p || range scan || - | | ||
|- | |- | ||
|99 || high || 54.19 || 25.8 || 1.2x10^9 || 29.4 || p || range scan || - | | |99 || high || 54.19 || 25.8 || 1.2x10^9 || 29.4 || p || range scan || - | | ||
Line 500: | Line 500: | ||
|116 || high || 125.06 || 114.7 || 1.2x10^9 || 12.9 || p || range scan || - | | |116 || high || 125.06 || 114.7 || 1.2x10^9 || 12.9 || p || range scan || - | | ||
|- | |- | ||
|117 || high || 128. | |117 || high || 128.11 || 119.7 || 1.2x10^9 || 12.7 || p || range scan || - | | ||
|- | |- | ||
|118 || high || None || - || None || None || - || background || - | | |118 || high || None || - || None || None || - || background || - | |
Latest revision as of 14:21, 1 November 2019
2 night shifts with range calorimeter and single PMT module.
Equipment List
Item | Notes |
---|---|
Network Hub | Set in control room to take output from experimental room ethernet switcher. Control laptops connected via ethernet or 5GHz WiFi. |
Control Laptop x2 | |
DAQ desktop PC | |
Ethernet Switcher | Set in experiment room and connected DAQ desktop PC. Output sent to control room Network Hub. |
Ethernet Cable x 5 | To connect DAQ PC to switcher, switcher to Network Hub (long cable), 2 laptops to Network Hub, if scope: scope to switcher. |
Range Detector Experiment
Item | Notes |
---|---|
Portable Enclosure | Modified Big Zarges Waterproof Wheeled Equipment Case.
Features mount for scintillator and PMT, opening for beam, and ports for SHV, BNC, SMA, Camera Link cables. |
Scintillator stack | 14 x 2 mm, 15 x 2.6 mm and 20 x 3 mm sheets in ascending order from the front/beam of the end of the scintillator. Sheet 21 not used. Sheet numbers from front to end (in beam direction): 30,29,28,27,26,22,19,18,17,15,14,13,10,7,25,24,23,20,16,12,11,9,8,6,5,4,3,2,1,31,32,33,34,35,36,37,38,39,40,41,42,43,44,45,46,47,48,49,50. |
ISDI CMOS sensor | sensor pixel dimension: 1030 x 1536. NO optical grease between scintillator and sensor. Connected to DAQ PC via Camera Link cable. |
DAQ desktop PC | Controls sensor aquisition. |
Gloves | For handling scintillator |
Single Module Experiment
Item | Notes |
---|---|
Portable Enclosure | Modified Big Zarges Waterproof Wheeled Equipment Case.
Features mount for scintillator and PMT, opening for beam, and ports for SHV, BNC, SMA, Camera Link cables. |
Single Module Scintillator Blocks x5 | 1cm/3cm/5cm x 10cm x 10cm standard scintillator block. Optical gel required to couple Scintillator to PMT. 5cm and 3cm with black wrapping. 5cm, 3cm, 1cm mylar wrapping. |
Photomultiplier Tube | Hamamatsu R13089 2" |
Caen DT5751 Digitiser | Records PMT output, connected via SMA-MCX conversion. Used channel 3. |
Caen NDT1470 HV Supply | Supplies HV to PMT |
USB Cable | Caen units to DAQ PC |
DAQ desktop PC | Controls Caen HV supply; records data from Caen digitiser. |
Male-to-Male SMA Cable x3 | Signal from enclosure port to digitiser. Short/Light and Long cables connected via SMA to SMA connector. If scope used: SMA splitter + SMA cable + SMA to BNC adaptor used to split the input signal and send it both to scope and digitiser. |
DC-4200 MHz Signal Splitter | Splits SMA signal to two SMA signals – sent to Oscilloscope and Digitiser. |
Female-to-Female SMA Connector | Connects Long SMA cable to Short/Light SMA cable. |
Female-to-Female BNC to SMA Connector x2 | From case to SMA cable (or splitter); From SMA cable to scope. |
Male SMA to Male MCX Adaptor | Sends output from SMA cable to Digitiser |
Gloves | For handling scintillator |
Optical gel | For coupling scintillator to PMT |
Wipes | For removing optical gel |
Experiment List
16–17th November
Range calorimeter measurements with ISDI sensor + scintillator stack.
- Helium:
- Background measurement.
- Intensity check with 50.57 MeV/u pristine Bragg peak.
- Shoot-through at 220.51 MeV/u for sensor intensity calibration.
- Individual range measurements from 50.57 MeV/u (2.287cm) to 110 mm (=122.3 MeV/u) range:
- 5mm steps from min to max.
- 10 x 1mm steps from 50.57MeV/u upwards. (10 lowest energy steps; 50.57MeV/u, 51.79MeV/u, 52.97,54.14,55.30, 56.44, 57.54,58.64,59.74,60.80 MeV/u = 2.287, 2.388,2.488,2.587,2.587,2.788,2.887,2.987,3.088,3.187 cm)
- 10 x 1mm steps from ~100mm downwards. (the 10 step down from 115.9 MeV: Ranges : 10.09,99.89,9.889,9.789,9.689,9.589,9.489,9.388,9.289,9.189)
- SOBP with single pixel stripe in single acquisition.
- Carbon:
- Background measurement.
- Intensity check with 88.83 MeV/u pristine Bragg peak.
- Shoot-through at 430.10 MeV/u for sensor intensity calibration.
- Individual range measurements from 88.83 MeV/u to 110 mm range:
- 5mm steps from min to max.
- 10 x 1mm steps from 88.83 MeV/u upwards.
- 10 x 1mm steps from ~100mm downwards.
- SOBP with single pixel stripe in single acquisition.
- Proton:
- Background measurement.
- Intensity check with 48.12 MeV pristine Bragg peak.
- Shoot-through at 221.06 MeV/u for sensor intensity calibration.
- Individual range measurements from 48.12 MeV/u to 110 mm range:
- 5mm steps from min to max.
- 10 x 1mm steps from 48.12 MeV/u upwards.
- 10 x 1mm steps from ~100mm downwards.
- SOBP with single pixel stripe in single acquisition.
- Position scan at 100mm with 30mm transverse offset in both positive and negative x.
Oxygen measurements planned but not requested:
- Background measurement.
- Intensity check with 103.77 MeV pristine Bragg peak.
- Shoot-through at 514.82 MeV/u for sensor intensity calibration.
- Individual range measurements from 103.77 MeV/u to 110 mm range:
- 5mm steps from min to max.
- 10 x 1mm steps from 103.77 MeV/u upwards.
- 10 x 1mm steps from ~100mm downwards.
- SOBP with single pixel stripe in single acquisition.
17–18th November
Range calorimeter measurements with ISDI sensor + scintillator stack
- Helium:
- Background measurement.
- Intensity check with 50.57 MeV/u pristine Bragg peak.
- Shoot-through at 220.51 MeV/u for sensor intensity calibration.
- SOBP measurements:
- Correct SOBP measurement with 100mm distal edge using equal particle rate but clinical weighting for each layer with 3x3 spot grid.
- Correct SOBP measurement with 100mm distal edge using clinical weighting with approx. equal time to deliver each each layer with single on-axis spot.
- Faked SOBP measurement with 100mm distal edge using equal particle rate and equal weighting for each spot with single on-axis spot.
- Both measurements SOBP with single pixel stripe in single acquisition: 6x1,536 pixel area; approx 1m50s acquisition time.
- Mixed particle tests:
- Carbon clinical SOBP with 100mm distal edge using equal particle rate but clinical weighting for 15 layers with 2mm range spacing and 3x3 spot grid.
- Carbon faked SOBP with 100mm distal edge using equal particle rate and equal weighting for each layer with single on-axis spot for 10 layers with 3mm layer spacing.
- Helium replica of 100mm Carbon clinical SOBP (260mm distal edge) with 10 layers with 8/9mm layer spacing (closest approximation of 3mm Carbon layer spacing); equal particle rate with clinical weighting and 3x3 spot grid.
- Helium replica of 100mm Carbon faked SOBP (260mm distal edge) using equal particle rate and equal weighting for each layer with single on-axis spot for 10 layers with 8/9mm layer spacing.
- Repeat measurements with 50mm PMMA sheets until Helium SOBP fully contained in detector.
- Add 30mm PMMA sheet and repeat Helium measurements in small transverse steps (tumour motion simulation).
Single module
All tests with Helium.
- 30mm thick scintillator sheet: check max signal height and replace with black wrapped sheet if signal too large. 22mm Helium range so Bragg peak fully contained in scintillator.
- Increase rate from kHz up to MHz to identify maximum safe particle rate. Observe effect of PMT current saturation.
- Increase Helium energy from 22mm up to ~260mm in 10mm range steps.
- 1mm range steps from 22mm to 32mm.
- 1mm range steps from 190mm to 200mm.
- Investigate effect of upstream degrader material.
Experiment Log
PMMA blocks are 50mm thick. Water Equivalent Thickness is 57.81mm.
Beam Tests
16–17th November
Beam focus ranges from step 1 (smallest) to step 4 (of 6).
Helium ion beam
full sensor area
Run number | Full Well Mode | Beam Energy (MeV/u) | Range (estimated) | N particles | Spot size (mm) | Test type | Comments | Thumbnail |
---|---|---|---|---|---|---|---|---|
N | well mode | Energy MeV/u | Range | particles | Spot FWHM | test | comments | |
00 | None | None | - | None | None | - | background | |
01 | low | 50.57 | 22.9 | 10^8 | 18.6 | He | low energy, low intensity | |
02 | high | 50.57 | 22.9 | 8x10^8 | 18.6 | He | low energy, max clinical intensity | |
03 | high | 220.51 | 308.3 | 8x10^8 | 4.9 | He | 50mm PMMA absorber upstream, shoot-through | |
04 | high | 220.51 | 308.3 | 8x10^8 | 4.9 | He | (same as 3) 50mm PMMA absorber upstream, shoot-through | |
05 | high | 220.51 | 308.3 | 8x10^8 | 4.9 | He | (same as 3) 50mm PMMA absorber upstream, shoot-through | |
high | 122.3 | 110.9 | 8x10^8 | 8.0 | He | range measurement was saturated, not saved | ||
06 | high | 122.3 | 110.9 | 5x10^8 | 8.0 | He | range measurement, max range | |
07 | high | 122.3 | 110.9 | 5x10^8 | 8.0 | He | (same as 6) range measurement, max range | |
08 | high | 119.15 | 105.9 | 5x10^8 | 12.0 | He | range measurement 5mm step | |
09 | high | 119.15 | 105.9 | 5x10^8 | 12.0 | He | (same as 8) range measurement 10mm step | |
10 | high | 115.93 | 100.9 | 5x10^8 | 12.2 | He | range measurement 10mm step | |
11 | high | 109.28 | 90.9 | 5x10^8 | 12.5 | He | range measurement 10mm step | |
12 | high | 102.34 | 80.9 | 5x10^8 | 12.8 | He | range measurement 10mm step | |
13 | high | 95.02 | 70.9 | 5x10^8 | 13.3 | He | range measurement 10mm step | |
14 | high | 87.25 | 60.9 | 5x10^8 | 13.9 | He | range measurement 10mm step | |
15 | high | 78.92 | 50.9 | 3x10^8 | 14.7 | He | range measurement 10mm step | |
16 | high | 69.84 | 40.9 | 3x10^8 | 13.3 | He | range measurement 10mm step | |
17 | high | 60.8 | 31.9 | 3x10^8 | 15.4 | He | range measurement 10mm step | |
18 | high | 60.8 | 31.9 | 5x10^8 | 15.4 | He | range measurement 10mm step | |
19 | high | 59.74 | 30.9 | 5x10^8 | 15.7 | He | range measurement 1mm step | |
20 | high | 58.7 | 29.9 | 5x10^8 | 16.0 | He | range measurement 1mm step | |
21 | high | 57.55 | 28.9 | 5x10^8 | 16.3 | He | range measurement 1mm step | |
22 | high | 56.44 | 27.9 | 5x10^8 | 16.7 | He | range measurement 1mm step | |
23 | high | 55.30 | 26.9 | 5x10^8 | 17.0 | He | range measurement 1mm step | |
24 | high | 55.30 | 26.9 | 8x10^8 | 17.0 | He | (same as 23) range measurement 1mm step | |
25 | high | 54.15 | 25.9 | 8x10^8 | 17.4 | He | range measurement 1mm step | |
26 | high | 52.98 | 24.9 | 8x10^8 | 17.8 | He | range measurement 1mm step | |
27 | high | 51.79 | 23.9 | 8x10^8 | 18.2 | He | range measurement 1mm step | |
28 | high | 50.97 | 22.9 | 8x10^8 | 18.6 | He | range measurement 1mm step | |
29 | high | 122.3 | 110.9 | 5x10^8 | 8.0 | He | high energy range measurement, 1mm step | |
30 | high | 121.68 | 109.9 | 5x10^8 | 8.0 | He | high energy range measurement, 1mm step | |
31 | high | 121.05 | 108.9 | 5x10^8 | 8.1 | He | high energy range measurement, 1mm step | |
32 | high | 120.42 | 107.9 | 5x10^8 | 8.1 | He | high energy range measurement, 1mm step | |
33 | high | 119.78 | 106.9 | 5x10^8 | 8.1 | He | high energy range measurement, 1mm step | |
34 | high | 119.15 | 105.9 | 5x10^8 | 8.2 | He | high energy range measurement, 1mm step | |
35 | high | 118.51 | 104.9 | 5x10^8 | 8.2 | He | high energy range measurement, 1mm step | |
36 | high | 117.87 | 103.9 | 5x10^8 | 8.2 | He | high energy range measurement, 1mm step | |
37 | high | 117.23 | 102.9 | 5x10^8 | 8.3 | He | high energy range measurement, 1mm step | |
38 | high | 116.58 | 101.9 | 5x10^8 | 8.3 | He | high energy range measurement, 1mm step | |
39 | high | 115.93 | 100.9 | 5x10^8 | 8.3 | He | high energy range measurement, 1mm step | |
40 | high | 50.57 | 100.9 | 8x10^8 | 8.3 | He | min range repetition | |
41 | high | 122.3 | 110.9 | 5x10^8 | 8.0 | He | max range | |
high | 96.52 - 115.93 | 72.9 - 100.9 | 3x10^8 | 9.7 - 8.3 | He | SOBP, 29 x 1mm steps. Observed, not saved | ||
42 | None | None | - | None | None | - | background 6 pixel rows (512 offset) | |
43 | high | 96.52 - 115.93 | 72.9 - 100.9 | 3x10^8 | 9.7 - 8.3 | He | SOBP, 29 x 1mm steps 6 pixel rows (512 offset) |
Carbon ion beam
full sensor
Run number | Full Well Mode | Beam Energy (MeV/u) | Range (estimated) | N particles | Spot size (mm) | Test type | Comments | Thumbnail |
---|---|---|---|---|---|---|---|---|
N | well mode | Energy MeV/u | Range | particles | Spot FWHM | test | comments | |
44 | high | None | - | None | None | - | background | |
45 | low | None | - | None | None | - | background | |
Beam off for 30 min | - | - | - | - | C | - | ||
46 | low | 88.83 | 21.1 | 3x10^7 | 9.8 | C | low energy, low intensity | |
47 | low | 430.1 | 309.4 | 3x10^7 | 3.4 | C | 50mm PMMA absorber upstream, shoot-through | |
48 | low | 217.25 | 100.4 | 3x10^7 | 4.8 | C | range scan | |
49 | low | 204.27 | 90.4 | 3x10^7 | 5.0 | C | range scan | |
50 | low | 190.75 | 80.4 | 3x10^7 | 5.3 | C | range scan | |
51 | low | 176.56 | 70.4 | 3x10^7 | 5.7 | C | range scan | |
52 | low | 161.55 | 60.4 | 3x10^7 | 6.1 | C | range scan | |
53 | low | 145.47 | 50.3 | 3x10^7 | 6.7 | C | range scan, Saturated | |
54 | low | 145.47 | 50.3 | 2x10^7 | 6.7 | C | range scan | |
55 | low | 127.96 | 40.2 | 2x10^7 | 7.5 | C | range scan | |
56 | low | 108.53 | 30.1 | 2x10^7 | 8.5 | C | range scan | |
57 | low | 106.45 | 29.0 | 3x10^7 | 8.6 | C | range scan, 10 steps | |
58 | low | 104.35 | 28.0 | 3x10^7 | 8.8 | C | range scan, 10 steps | |
59 | low | 102.22 | 27.0 | 3x10^7 | 8.9 | C | range scan, 10 steps | |
60 | low | 100.07 | 26.0 | 3x10^7 | 9.0 | C | range scan, 10 steps | |
61 | low | 97.88 | 25.0 | 3x10^7 | 9.2 | C | range scan, 10 steps | |
62 | low | 95.67 | 24.0 | 3x10^7 | 9.3 | C | range scan, 10 steps | |
63 | low | 91.42 | 23.0 | 3x10^7 | 9.5 | C | range scan, 10 steps | |
64 | low | 91.14 | 22.1 | 3x10^7 | 9.6 | C | range scan, 10 steps | |
65 | low | 88.83 | 21.1 | 3x10^7 | 9.8 | C | range scan, 10 steps | |
66 | low | 204.27 | 90.4 | 3x10^7 | 5.0 | C | range scan 10 steps | |
67 | low | 205.60 | 91.4 | 3x10^7 | 5.0 | C | range scan 10 steps | |
68 | low | 206.91 | 92.4 | 3x10^7 | 5.0 | C | range scan 10 steps | |
69 | low | 208.22 | 93.4 | 3x10^7 | 4.9 | C | range scan 10 steps | |
70 | low | 209.53 | 94.4 | 3x10^7 | 4.9 | C | range scan 10 steps | |
71 | low | 210.83 | 95.4 | 3x10^7 | 4.9 | C | range scan 10 steps | |
72 | low | 212.12 | 96.4 | 3x10^7 | 4.9 | C | range scan 10 steps | |
73 | low | 213.41 | 97.4 | 3x10^7 | 4.8 | C | range scan 10 steps | |
74 | low | 214.70 | 98.4 | 3x10^7 | 4.8 | C | range scan 10 steps | |
75 | low | 215.98 | 99.5 | 3x10^7 | 4.8 | C | range scan 10 steps | |
76 | low | 217.25 | 100.4 | 3x10^7 | 4.8 | C | range scan 10 steps | |
77 | low | 88.83 | 21.1 | 3x10^7 | 9.8 | C | range scan 10 steps | |
78 | low | None | - | None | None | - | background |
Proton beam
full sensor
Run number | Full Well Mode | Beam Energy (MeV/u) | Range (estimated) | N particles | Spot size (mm) | Test type | Comments | Thumbnail |
---|---|---|---|---|---|---|---|---|
N | well mode | Energy MeV/u | Range | particles | Spot FWHM | test | comments | |
79 | high | None | - | None | None | - | background | |
80 | high | 221.06 | 308.1 | 3.2x10^9 | 8.1 | p | 50mm PMMA absorber upstream, shoot-through | |
81 | high | 48.12 | 20.8 | 3.2x10^9 | 32.4 | p | low energy, low intensity | |
82 | high | 121.95 | 109.7 | 1.2x10^9 | 13.2 | p | range scan | |
83 | high | 115.55 | 99.8 | 1.2x10^9 | 13.8 | p | range scan | |
84 | high | 108.88 | 89.8 | 1.2x10^9 | 14.5 | p | range scan | |
85 | high | 101.9 | 79.8 | 1.2x10^9 | 15.5 | p | range scan | |
86 | high | 94.54 | 69.9 | 1.2x10^9 | 16.6 | p | range scan | |
87 | high | 86.72 | 59.9 | 1.2x10^9 | 18.2 | p | range scan | |
Range to be checked | - | - | - | - | - | p | range | |
88 | high | 78.3 | 49.9 | 1.2x10^9 | 20.2 | p | range scan | |
89 | high | 78.3 | 49.9 | 8x10^8 | 20.2 | p | range scan | |
90 | high | 69.06 | 39.8 | 8x10^8 | 23.9 | p | range scan | |
91 | high | 58.72 | 29.8 | 8x10^8 | 27.3 | p | range scan | |
Range to be checked | - | - | - | - | - | p | range | |
92 | high | 78.3 | 49.9 | 8x10^8 | 20.2 | p | range scan | |
93 | high | 69.06 | 39.8 | 8x10^8 | 23.9 | p | range scan | |
94 | high | 69.06 | 39.8 | 1.2x10^9 | 23.9 | p | range scan | |
95 | high | 58.72 | 29.8 | 1.2x10^9 | 27.3 | p | range scan | |
96 | high | 57.61 | 28.8 | 1.2x10^9 | 27.8 | p | range scan | |
97 | high | 56.49 | 27.8 | 1.2x10^9 | 28.3 | p | range scan 1mm scan | |
98 | high | 55.34 | 26.8 | 1.2x10^9 | 28.9 | p | range scan | |
99 | high | 54.19 | 25.8 | 1.2x10^9 | 29.4 | p | range scan | |
100 | high | 53.09 | 24.8 | 2x10^9 | 30.0 | p | range scan | |
101 | high | 51.82 | 23.8 | 3.2x10^9 | 30.7 | p | range scan | |
102 | high | 50.60 | 22.8 | 3.2x10^9 | 31.3 | p | range scan | |
103 | high | 49.37 | 21.8 | 3.2x10^9 | 31.9 | p | range scan | |
104 | high | 48.12 | 20.8 | 3.2x10^9 | 32.4 | p | range scan | |
105 | high | 115.55 | 99.8 | 1.2x10^9 | 13.8 | p | range scan | |
106 | high | 116.20 | 100.8 | 1.2x10^9 | 13.7 | p | range scan | |
107 | high | 116.85 | 101.8 | 1.2x10^9 | 13.7 | p | range scan | |
108 | high | 117.50 | 102.8 | 1.2x10^9 | 13.6 | p | range scan | |
109 | high | 118.14 | 103.7 | 1.2x10^9 | 13.6 | p | range scan | |
110 | high | 118.78 | 104.7 | 1.2x10^9 | 13.5 | p | range scan | |
111 | high | 119.42 | 105.7 | 1.2x10^9 | 13.4 | p | range scan | |
112 | high | 120.05 | 106.7 | 1.2x10^9 | 13.4 | p | range scan | |
113 | high | 120.69 | 107.7 | 1.2x10^9 | 13.3 | p | range scan | |
114 | high | 121.32 | 108.7 | 1.2x10^9 | 13.3 | p | range scan | |
115 | high | 121.95 | 109.7 | 1.2x10^9 | 13.2 | p | range scan | |
116 | high | 125.06 | 114.7 | 1.2x10^9 | 12.9 | p | range scan | |
117 | high | 128.11 | 119.7 | 1.2x10^9 | 12.7 | p | range scan | |
118 | high | None | - | None | None | - | background | |
119 | high | 115.55 | 99.8 | 1.2x10^9 | 13.8 | p | spot scan on axis | |
120 | high | 115.55 | 99.8 | 1.2x10^9 | 13.8 | p | spot scan (2cm shift on the left) | |
121 | high | 115.55 | 99.8 | 1.2x10^9 | 13.8 | p | spot scan (2cm shift on the right) |
17–18th November
Beam focus ranges from step 1 (smallest) to step 4 (of 6).
Helium ion beam
full sensor area
Run number | Full Well Mode | Beam Energy (MeV/u) | Range (estimated) | N ions /s | Spot size (mm) | Test type | Comments | Thumbnail |
---|---|---|---|---|---|---|---|---|
N | well mode | Energy MeV/u | Range | ions/s | Spot FWHM | test | comments | |
00 | high | None | - | None | None | - | background | |
01 | high | 50.57 | 22.9 | 8x10^8 | 18.6 | He | low energy, max clinical intensity | |
- | - | - | - | - | - | New alignment | ||
02 | high | 220.51 | 308.3 | 8x10^8 | 4.9 | He | 50mm PMMA absorber upstream, shoot-through | |
03 | high | None | - | None | None | - | background | |
04 | high | 50.57 | 22.9 | 8x10^8 | 18.6 | He | low energy, max clinical intensity | |
05 | high | 96.52 - 115.93 | 72.9 - 100.9 | 3x10^8 | 9.7 - 8.3 | He | SOBP, 6 pixel rows (512 offset), 100mm distal edge using equal particle rate but clinical weighting for each layer with 3x3 spot grid | |
06 | high | None | - | None | None | - | background 6 pixel rows (512 offset) | |
07 | high | 96.52 - 115.93 | 72.9 - 100.9 | 3x10^8 | 9.7 - 8.3 | He | Clinical SOBP, 6 pixel rows (512 offset), 100mm distal edge using clinical weighting with approx. equal time to deliver each layer with single on-axis spot; (Saturated) | |
08 | high | 96.52 - 115.93 | 72.9 - 100.9 | 5x10^8 | 9.7 - 8.3 | He | Faked SOBP, 4 pixel rows (513 offset), 100mm distal edge using equal particle rate and equal weighting for each spot with single on-axis spot. |
Mixed Helium and Carbon ions beam
full sensor
Run number | Full Well Mode | Beam Energy (MeV/u) | Range (estimated) | N particles | Spot size (mm) | Test type | Comments | Thumbnail |
---|---|---|---|---|---|---|---|---|
N | well mode | Energy MeV/u | Range | particles | Spot FWHM | test | comments | |
09 | high | 182.32 - 222.31 | 74.4 - 104.5 | - | 5.5 - 4.7 | C | Real Clinical SOBP, 6 pixel rows (512 offset), 100mm distal edge using clinical weighting with approx. equal time to deliver each layer with single on-axis spot. | |
10 | high | 182.32 - 222.31 | 74.4 - 104.5 | 8x10^7 | 5.5 - 4.7 | C | Clinical SOBP, 10 pixel rows (510 offset), 100mm distal edge using equal particle rate but clinical weighting for 15 layers with 2mm range spacing and 3x3 spot grid. | |
11 | high | None | - | None | None | - | background, 10 pixel rows (510 offset) | |
12 | high | 182.32 - 222.31 | 74.4 - 104.5 | 8x10^7 | 5.5 - 4.7 | C | Faked SOBP, 10 pixel rows (510 offset), 100mm distal edge using equal particle rate and equal weighting for each spot with single on-axis spot for 10 layers with 3mm layer spacing. | |
13 | high | 188.07 - 220.51 | 78.8 - 308.3 | 3x10^8 | 5.3 - 4.9 | He | Clinical SOBP, 10 pixel rows (510 offset), 100mm distal edge using equal particle rate but clinical weighting for 15 layers with 2mm range spacing and 3x3 spot grid. | |
14 | high | None | - | None | None | - | background, 10 pixel rows (510 offset) | |
15 | high | 188.07 - 220.51 | 78.8 - 308.3 | 3x10^8 | 5.3 - 4.9 | He | PROBLEMATIC RUN! Faked SOBP, 10 pixel rows (510 offset), 100mm distal edge using equal particle rate and equal weighting for each spot with single on-axis spot for 10 layers with 3mm layer spacing. | |
16 | high | 188.07 - 220.51 | 78.8 - 308.3 | 3x10^8 | 5.3 - 4.9 | He | Faked SOBP, 10 pixel rows (510 offset), 100mm distal edge using equal particle rate and equal weighting for each spot with single on-axis spot for 10 layers with 3mm layer spacing. | |
17 | high | 188.07 - 220.51 | 78.8 - 308.3 | 3x10^8 | 5.3 - 4.9 | He | Faked SOBP, 10 pixel rows (510 offset), 100mm distal edge using equal particle rate and equal weighting for each spot with single on-axis spot for 10 layers with 3mm layer spacing. 50mm PMMA absorber | |
18 | high | 188.07 - 220.51 | 78.8 - 308.3 | 3x10^8 | 5.3 - 4.9 | He | Clinical SOBP, 10 pixel rows (510 offset), 100mm distal edge using equal particle rate but clinical weighting for 15 layers with 2mm range spacing and 3x3 spot grid. 50mm PMMA absorber | |
19 | high | 182.32 - 222.31 | 74.4 - 104.5 | 8x10^7 | 5.5 - 4.7 | C | Faked SOBP, 10 pixel rows (510 offset), 100mm distal edge using equal particle rate and equal weighting for each spot with single on-axis spot for 10 layers with 3mm layer spacing. 50mm PMMA absorber | |
20 | high | 182.32 - 222.31 | 74.4 - 104.5 | 8x10^7 | 5.5 - 4.7 | C | Clinical SOBP, 10 pixel rows (510 offset), 100mm distal edge using equal particle rate but clinical weighting for 15 layers with 2mm range spacing and 3x3 spot grid. 50mm PMMA absorber | |
21 | high | 430.1 | 309.4 | 8x10^7 | 3.4 | C | Full sensor, 50mm PMMA absorber upstream, shoot-through | |
22 | high | 220.51 | 308.3 | 8x10^8 | 4.9 | He | Full sensor, 50mm PMMA absorber upstream, shoot-through | |
23 | high | 182.32 - 222.31 | 74.4 - 104.5 | 8x10^7 | 5.5 - 4.7 | C | Clinical SOBP, 10 pixel rows (510 offset), 100mm distal edge using equal particle rate but clinical weighting for 15 layers with 2mm range spacing and 3x3 spot grid. 2x50mm PMMA absorber | |
24 | high | 182.32 - 222.31 | 74.4 - 104.5 | 8x10^7 | 5.5 - 4.7 | C | Faked SOBP, 10 pixel rows (510 offset), 100mm distal edge using equal particle rate and equal weighting for each spot with single on-axis spot for 10 layers with 3mm layer spacing. 2x50mm PMMA absorber | |
25 | high | 188.07 - 220.51 | 78.8 - 308.3 | 3x10^8 | 5.3 - 4.9 | He | Faked SOBP, 10 pixel rows (510 offset), 100mm distal edge using equal particle rate and equal weighting for each spot with single on-axis spot for 10 layers with 3mm layer spacing. 2x50mm PMMA absorber | |
26 | high | 188.07 - 220.51 | 78.8 - 308.3 | 3x10^8 | 5.3 - 4.9 | He | Clinical SOBP, 10 pixel rows (510 offset), 100mm distal edge using equal particle rate but clinical weighting for 15 layers with 2mm range spacing and 3x3 spot grid. 2x50mm PMMA absorber | |
27 | high | 188.07 - 220.51 | 78.8 - 308.3 | 3x10^8 | 5.3 - 4.9 | He | Clinical SOBP, 10 pixel rows (510 offset), 100mm distal edge using equal particle rate but clinical weighting for 15 layers with 2mm range spacing and 3x3 spot grid. 3x50mm PMMA absorber | |
28 | high | 188.07 - 220.51 | 78.8 - 308.3 | 3x10^8 | 5.3 - 4.9 | He | Faked SOBP, 10 pixel rows (510 offset), 100mm distal edge using equal particle rate and equal weighting for each spot with single on-axis spot for 10 layers with 3mm layer spacing. 3x50mm PMMA absorber | |
29 | high | 188.07 - 220.51 | 78.8 - 308.3 | 3x10^8 | 5.3 - 4.9 | He | Faked SOBP, 10 pixel rows (510 offset), 100mm distal edge using equal particle rate and equal weighting for each spot with single on-axis spot for 10 layers with 3mm layer spacing. 3x50mm + 1x27mm PMMA absorber | |
30 | high | 188.07 - 220.51 | 78.8 - 308.3 | 3x10^8 | 5.3 - 4.9 | He | Clinical SOBP, 10 pixel rows (510 offset), 100mm distal edge using equal particle rate but clinical weighting for 15 layers with 2mm range spacing and 3x3 spot grid. 3x50mm + 1x27mm PMMA absorber | |
31 | high | 188.07 - 220.51 | 78.8 - 308.3 | 3x10^8 | 5.3 - 4.9 | He | Clinical SOBP, 10 pixel rows (510 offset), 100mm distal edge using equal particle rate but clinical weighting for 15 layers with 2mm range spacing and 3x3 spot grid. From the box 2x50mm + 1x27mm + 1x50mm + 1x30mm PMMA absorber | |
32 | high | 188.07 - 220.51 | 78.8 - 308.3 | 3x10^8 | 5.3 - 4.9 | He | Faked SOBP, 10 pixel rows (510 offset), 100mm distal edge using equal particle rate and equal weighting for each spot with single on-axis spot for 10 layers with 3mm layer spacing. From the box 2x50mm + 1x27mm + 1x50mm + 1x30mm PMMA absorber | |
33 | high | 188.07 - 220.51 | 78.8 - 308.3 | 3x10^8 | 5.3 - 4.9 | He | Faked SOBP, 10 pixel rows (510 offset), 100mm distal edge using equal particle rate and equal weighting for each spot with single on-axis spot for 10 layers with 3mm layer spacing. From the box 2x50mm + 1x27mm (shifted of half block on the -x axis (+z axis is the beam direction) 3mm further in rispect to the central axis of the beam (+3mm respect to the central axis)) + 1x50mm + 1x30mm PMMA absorber | |
34 | high | 188.07 - 220.51 | 78.8 - 308.3 | 3x10^8 | 5.3 - 4.9 | He | Clinical SOBP, 10 pixel rows (510 offset), 100mm distal edge using equal particle rate but clinical weighting for 15 layers with 2mm range spacing and 3x3 spot grid. From the box 2x50mm + 1x27mm (shifted of half block to the -x axis (+z axis is the beam direction) 3mm further in rispect to the central axis of the beam (+3mm respect to the central axis)) + 1x50mm + 1x30mm PMMA absorber | |
35 | high | 188.07 - 220.51 | 78.8 - 308.3 | 3x10^8 | 5.3 - 4.9 | He | Clinical SOBP, 10 pixel rows (510 offset), 100mm distal edge using equal particle rate but clinical weighting for 15 layers with 2mm range spacing and 3x3 spot grid. From the box 2x50mm + 1x27mm (shifted of half block to the -x axis (+z axis is the beam direction) rispect to the central axis of the beam (0.0mm respect to the central axis))) + 1x50mm + 1x30mm PMMA absorber | |
36 | high | 188.07 - 220.51 | 78.8 - 308.3 | 3x10^8 | 5.3 - 4.9 | He | Faked SOBP, 10 pixel rows (510 offset), 100mm distal edge using equal particle rate and equal weighting for each spot with single on-axis spot for 10 layers with 3mm layer spacing. From the box 2x50mm + 1x27mm (shifted of half block to the -x axis (+z axis is the beam direction) rispect to the central axis of the beam (0.0mm respect to the central axis)) + 1x50mm + 1x30mm PMMA absorber | |
37 | high | 220.51 | 308.3 | 3x10^8 | 4.9 | He | Trasverse position scan. 3mm Right. Single energy layer 3x3, full sensor. From the box 2x50mm + 1x27mm (shifted of half block to the -x axis (+z axis is the beam direction) rispect to the central axis of the beam) + 1x50mm + 1x30mm PMMA absorber | |
38 | high | 220.51 | 308.3 | 3x10^8 | 4.9 | He | Trasverse position scan. Middle. Single energy layer 3x3, full sensor. From the box 2x50mm + 1x27mm (shifted of half block to the -x axis (+z axis is the beam direction) rispect to the central axis of the beam) + 1x50mm + 1x30mm PMMA absorber | |
39 | high | 220.51 | 308.3 | 3x10^8 | 4.9 | He | Trasverse position scan. 3mm Left. Single energy layer 3x3, full sensor. From the box 2x50mm + 1x27mm (shifted of half block to the -x axis (+z axis is the beam direction) rispect to the central axis of the beam) + 1x50mm + 1x30mm PMMA absorber | |
40 | high | 188.07 - 220.51 | 78.8 - 308.3 | 3x10^8 | 5.3 - 4.9 | He | Clinical SOBP, 10 pixel rows (510 offset), 100mm distal edge using equal particle rate but clinical weighting for 15 layers with 2mm range spacing and 3x3 spot grid. From the box 2x50mm + 1x27mm (shifted of half block to the -x axis (+z axis is the beam direction) 3mm further out rispect to the central axis of the beam (-3mm respect to the central axis)) + 1x50mm + 1x30mm PMMA absorber | |
41 | high | 188.07 - 220.51 | 78.8 - 308.3 | 3x10^8 | 5.3 - 4.9 | He | Faked SOBP, 10 pixel rows (510 offset), 100mm distal edge using equal particle rate and equal weighting for each spot with single on-axis spot for 10 layers with 3mm layer spacing. From the box 2x50mm + 1x27mm (shifted of half block to the -x axis (+z axis is the beam direction) 3mm further out rispect to the central axis of the beam (-3mm respect to the central axis)) + 1x50mm + 1x30mm PMMA absorber | |
42 | high | 188.07 - 220.51 | 78.8 - 308.3 | 3x10^8 | 5.3 - 4.9 | He | Faked SOBP, 10 pixel rows (510 offset), 100mm distal edge using equal particle rate and equal weighting for each spot with single on-axis spot for 10 layers with 3mm layer spacing. From the box 2x50mm PMMA + 1x27mm AIR + 1x50mm PMMA + 1x30mm PMMA absorber. 27mm absorber removed | |
43 | high | 188.07 - 220.51 | 78.8 - 308.3 | 3x10^8 | 5.3 - 4.9 | He | Clinical SOBP, 10 pixel rows (510 offset), 100mm distal edge using equal particle rate but clinical weighting for 15 layers with 2mm range spacing and 3x3 spot grid. From the box From the box 2x50mm PMMA + 1x27mm AIR + 1x50mm PMMA + 1x30mm PMMA absorber. 27mm absorber removed | |
44 | high | None | - | None | None | - | background, 10 pixel rows (510 offset) |
Single Module
- He beam with focus F4
- PMT Current at HV supply -900 V, beam off: 0 μA, channel_0 used
- The digitiser always operated in Self Trigger mode. channel_3 used
- controlling the acquisition remotely.
- Working directory where digitiser data is stored:
/unix/pbt/data/hit/nov18/20181117_PMT
- 10x10x3 cm scintillator, mylar wrapping
Run number | High Voltage, V | PMT Current, uA | Beam Energy (MeV/u) | Beam Intensity | Rate (Measured on Digitiser) | Digitiser treshold LSB | Digitiser Settings and Notes |
---|---|---|---|---|---|---|---|
run | HV | current | beam E | beam Intensity | Rate | Treshold. | Notes |
N/A | -900 | 150.1 (beam off) | N/A | N/A | N/A | 40 | DC Offset: 0, Rec. Length: 3996 ns, Threshold: 40 LSB, Gate Offset: 40 ns, Long Gate: 320 ns, Short Gate: 28 ns, BL Mean: 8 ns, Charge. Sensitivty: 20 fC/LSB, Pre-trigger: 40 ns |
001 | -900 | 150.1 | 50.56 | 16 kHz | 40 | test | |
002 | -900 | 150.1 | 50.56 | 5.7 kHz | 40 | note | |
003 | -900 | 150.1 | 50.56 | 50 kHz | 40 | note | |
004 | -900 | 150.1 | 50.56 | 60 kHz | 40 | note | |
005 | -900 | 150.1 | 50.56 | 60 kHz | 40 | note | |
006 | -900 | 150.1 | 50.56 | 80 kHz | 40 | note | |
007 | -900 | 150.1 | 50.56 | 95 kHz | 40 | note | |
008 | -900 | 150.1 | 50.56 | 220 kHz | 40 | note | |
009 | -900 | 150.1 | 50.56 | 300 kHz | 40 | note | |
010 | -900 | 150.1 | 50.56 | 170 kHz | 40 | note | |
011 | -900 | 150.1 | 50.56 | 290 kHz | 40 | note | |
012 | -900 | 150.1 | 50.56 | 320 kHz | 40 | note | |
013 | -900 | 150.1 | 50.56 | 440 kHz | 40 | note | |
014 | -900 | 150.1 | 50.56 | 770 kHz | 40 | high intensity degradation | |
015 | -900 | 150.1 | 50.56 | 500 kHz | 40 | note | |
016 | -900 | 150.1 | 50.56 | 450 kHz | 40 | note | |
017 | -900 | 150.1 | 50.56 | 200 kHz | 40 | optimal Rate | |
018 | -900 | 150.1 | 51.79 | 170 kHz | 40 | energy scan | |
019 | -900 | 150.1 | 52.98 | 230 kHz | 40 | energy scan | |
020 | -900 | 150.1 | 54.15 | 170 kHz | 40 | energy scan | |
021 | -900 | 150.1 | 55.30 | 170 kHz | 40 | energy scan | |
022 | -900 | 150.1 | 56.44 | 200 kHz | 40 | energy scan | |
023 | -900 | 150.1 | 57.55 | 200 kHz | 40 | energy scan | |
024 | -900 | 150.1 | 58.65 | 260 kHz | 40 | energy scan | |
025 | -900 | 150.1 | 59.74 | 200 kHz | 40 | energy scan | |
026 | -900 | 150.1 | 60.80 | 200 kHz | 40 | energy scan | |
027 | -900 | 150.1 | 61.86 | 250 kHz | 40 | energy scan | |
028 | -900 | 150.1 | 62.90 | 250 kHz | 40 | energy scan | |
029 | -900 | 150.1 | 63.93 | 280 kHz | 40 | energy scan | |
030 | -900 | 150.1 | 64.94 | 230 kHz | 40 | energy scan | |
031 | -900 | 150.1 | 65.94 | 230 kHz | 40 | energy scan | |
032 | -900 | 150.1 | 66.93 | 250 kHz | 40 | energy scan | |
033 | -900 | 150.1 | 67.91 | 230 kHz | 40 | energy scan | |
034 | -900 | 150.1 | 68.88 | 200 kHz | 40 | energy scan | |
035 | -900 | 150.1 | 69.84 | 280 kHz | 40 | energy scan | |
036 | -900 | 150.1 | 70.79 | 280 kHz | 40 | energy scan | |
037 | -900 | 150.1 | 72.65 | 350 kHz | 40 | energy scan | |
038 | -900 | 150.1 | 74.48 | 200 kHz | 40 | energy scan | |
039 | -900 | 150.1 | 79.78 | 250 kHz | 40 | energy scan | |
040 | -900 | 150.1 | 88.05 | 200 kHz | 40 | energy scan | |
041 | -900 | 150.1 | 95.78 | 250 kHz | 40 | energy scan | |
042 | -900 | 150.1 | 103.05 | 230 kHz | 40 | energy scan | |
043 | -900 | 150.1 | 109.97 | 250 kHz | 40 | energy scan | |
044 | -900 | 150.1 | 116.58 | 300 kHz | 40 | energy scan | |
045 | -900 | 150.1 | 129.05 | 250 kHz | 40 | energy scan | |
046 | -900 | 150.1 | 140.71 | 250 kHz | 40 | energy scan | |
047 | -900 | 150.1 | 151.73 | 250 kHz | 40 | energy scan | |
048 | -900 | 150.1 | 162.22 | 260 kHz | 40 | energy scan | |
049 | -900 | 150.1 | 172.28 | 250 kHz | 40 | energy scan | |
050 | -900 | 150.1 | 183.85 | 250 kHz | 40 | energy scan | |
051 | -900 | 150.1 | 197.68 | 230 kHz | 40 | energy scan | |
052 | -900 | 150.1 | 210.92 | 260 kHz | 40 | energy scan | |
053 | -900 | 150.1 | 220.51 | 200 kHz | 40 | energy scan | |
054 | -900 | 150.1 | 50.57 | 200 kHz | 40 | 10mm PMMA absorber, 11.2mm Water Equivalent thickness | |
055 | -900 | 150.1 | 50.57 | 2 kHz | 40 | 2x10mm PMMA absorber, 2x11.2mm Water Equivalent thickness, 4 spills | |
056 | -900 | 150.1 | 95.78 | 17 kHz | 40 | 60mm PMMA absorber, 6x11.2mm Water Equivalent thickness | |
057 | -900 | 150.1 | 99.46 | 200 kHz | 40 | 60mm PMMA absorber, 6x11.2mm Water Equivalent thickness | |
058 | -900 | 150.1 | 103.05 | 150 kHz | 40 | 60mm PMMA absorber, 6x11.2mm Water Equivalent thickness | |
059 | -900 | 150.1 | 106.55 | 200 kHz | 40 | 60mm PMMA absorber, 6x11.2mm Water Equivalent thickness | |
060 | -900 | 150.1 | 109.97 | 250 kHz | 40 | 60mm PMMA absorber, 11.2mm Water Equivalent thickness | |
061 | -900 | 150.1 | 113.31 | 200 kHz | 40 | 60mm PMMA absorber, 6x11.2mm Water Equivalent thickness | |
062 | -900 | 150.1 | 116.58 | 250 kHz | 40 | 60mm PMMA absorber, 6x11.2mm Water Equivalent thickness | |
063 | -900 | 150.1 | 119.78 | 350 kHz | 40 | 60mm PMMA absorber, 6x11.2mm Water Equivalent thickness |