Line 113: | Line 113: | ||
# Tested trigger output of Tracker into Oscilloscope: square wave of period ~15 ns with amplitude ~400 mV. Pulse example recorded on Oscilloscope disk in folder: SShaikh/Tracker Trigger | # Tested trigger output of Tracker into Oscilloscope: square wave of period ~15 ns with amplitude ~400 mV. Pulse example recorded on Oscilloscope disk in folder: SShaikh/Tracker Trigger | ||
# Placed signal splitter for trigger to send to both Digitiser and Oscilloscope. Amplitude reduced to ~50%. Pulse example recorded on Oscilloscope disk in folder: SShaikh/Tracker Trigger (Split) | # Placed signal splitter for trigger to send to both Digitiser and Oscilloscope. Amplitude reduced to ~50%. Pulse example recorded on Oscilloscope disk in folder: SShaikh/Tracker Trigger (Split) | ||
# Split trigger | # Split trigger found to be too small for Digitiser (accepts trigger of order ~V), hence Digitiser set to self-trigger mode. | ||
# Aligned calorimeter to tracker, with centre of mylar window and tracker window aligned. | # Aligned calorimeter to tracker, with centre of mylar window and tracker window aligned approximately by eye. | ||
# | # Performed beam test (had approx 30 mins beam time), details provided below. | ||
# Was found that the tracker was not completely set centre to the beam. | |||
# Error in data taking: Trigger trace was recorded instead of PMT trace. | |||
'''26/10/18''' | |||
# Tracker was realigned to be centred to the beam. | |||
# Using Tipp-Ex markers and lable stickers, window to scintillator was aligned more accurately to window of tracker. | |||
# Test of tracker/PMT synchronisation (check tracker trigger pulse/PMT pulse sync and numbers of events in files) with scope vertical scale 50mV/div. Window center corresponds to right edge vertical line intersecting upper edge horizontal line, looking at the window. | |||
# Test of tracker/PMT synchronisation with scope vertical scale 100mV/div, 2mm collimator centered. | |||
# Test of tracker/PMT synchronisation with lower current, 2mm collimator centered. | |||
# Degrade beam energy and measure response with 0.95mm, 2.01mm, 2.96mm, 3.95mm, 4.90mm, 5.70mm PMMA before calorimeter, 2mm collimator centered. | |||
# Offset 2mm collimator (by 6mm in +ve) and measure response with 5.70 mm PMMA absorber. | |||
# Double 2mm collimator (12mm separated in x direction) and measure response with 5.70 mm PMMA absorber. | |||
# Offset 2mm collimator (by 6mm in +ve), no absorber. | |||
# Double 2mm collimator (12mm separated in x direction), no absorber. | |||
# Double 2mm collimator (12mm separated in x direction) repetition with lower current, no absorber. | |||
# Double 2mm collimator (12mm separated in x direction) with left-hand hole covered by 3.95 mm PMMA. | |||
== Beam Tests == | == Beam Tests == | ||
=== Thursday 25th October === | |||
PMT Current at HV supply -900 V, beam off: 150 μA | PMT Current at HV supply -900 V, beam off: 150 μA | ||
28MeV beam. Scope triggered by tracker. Cyclotron issues meant short run (~90 mins). Test of tracker triggering/scope synchronisation. | |||
Scope timebase: 20ns/div; trigger delay 80ns; C1 vertical scale 50mV/div. | |||
Tracker data indicates beam horizontal offset of 12mm; vertical offset of 1.3mm. | |||
Data stored in directory ProCal_Birm_25-10-18 | |||
{| class="wikitable" | {| class="wikitable" | ||
! style="text-align: center;" | Run Number | ! style="text-align: center;" | Run Number | ||
! style="text-align: center;" | High Voltage (V) | ! style="text-align: center;" | High Voltage (V) | ||
! style="text-align: center;" | | ! style="text-align: center;" | Trigger Level (mV) | ||
! style="text-align: center;" | | ! style="text-align: center;" | Trigger Source | ||
! style="text-align: center;" | Beam Current / Rate | |||
! style="text-align: center;" | Collimator | ! style="text-align: center;" | Collimator | ||
! style="text-align: center;" | | ! style="text-align: center;" | Filename | ||
! style="text-align: center;" | Notes | ! style="text-align: center;" | Notes | ||
|- | |- | ||
| style="text-align: center;" | 00 | | style="text-align: center;" | 00 | ||
| style="text-align: center;" | -900 | | style="text-align: center;" | -900 | ||
| style="text-align: center;" | - | | style="text-align: center;" | 70, +ve | ||
| style="text-align: center;" | ~10 kHz | | style="text-align: center;" | Tracker (C2) | ||
| style="text-align: center;" | 2 mm | | style="text-align: center;" | 170pA/~10 kHz | ||
| style="text-align: center;" | | | style="text-align: center;" | 2 mm centred | ||
| style="text-align: center;" | -- | | style="text-align: center;" | run00_2mmCCol_160pA_TrackTrig | ||
| style="text-align: center;" | | | style="text-align: center;" | First test of tracker/PMT synchronisation | ||
|- | |||
| style="text-align: center;" | 01 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 160pA/~10 kHz | |||
| style="text-align: center;" | 2 mm centred | |||
| style="text-align: center;" | run01_2mmCCol_160pA_TrackTrig | |||
| style="text-align: center;" | Also took Caen ADC data | |||
|- | |||
| style="text-align: center;" | 02 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 160pA/~10 kHz | |||
| style="text-align: center;" | 2 mm centred | |||
| style="text-align: center;" | run02_2mmCCol_160pA_TrackTrig | |||
| style="text-align: center;" | Caen ADC run 01 data also includes run 02 (long run) | |||
|- | |||
|} | |||
=== Friday 26th October === | |||
PMT Current at HV supply -900 V (actual 899.8V), beam off: 150.15 μA. | |||
36MeV beam: scattering foil inserted to increase beam size; 2 mm on-axis collimator. | |||
Scope triggered by tracker. Scope timebase: 20ns/div; trigger delay 80ns; C1 vertical scale 50mV/div. | |||
Data stored in directory ProCal_Birm_26-10-18 | |||
{| class="wikitable" | |||
! style="text-align: center;" | Run Number | |||
! style="text-align: center;" | High Voltage (V) | |||
! style="text-align: center;" | Trigger Level (mV) | |||
! style="text-align: center;" | Trigger Source | |||
! style="text-align: center;" | Beam Current / Rate | |||
! style="text-align: center;" | Collimator | |||
! style="text-align: center;" | Filename | |||
! style="text-align: center;" | Notes | |||
|- | |||
| style="text-align: center;" | 00 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 160-200pA/~10 kHz | |||
| style="text-align: center;" | 2 mm centred | |||
| style="text-align: center;" | run00_2mmCCol_160pA_TrackTrig | |||
| Test of tracker/PMT synchronisation with 5s tracker run, 50,000 scope segments, all traces saved; check tracker trigger pulse/PMT pulse sync and numbers of events in files. | |||
|- | |||
| style="text-align: center;" | 01 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 160-170pA/~10 kHz | |||
| style="text-align: center;" | 2 mm centred | |||
| style="text-align: center;" | run01_2mmCCol_160pA_TrackTrig | |||
| Repeat of previous test; scope vertical trace increased to 100mV/div. | |||
|- | |||
| style="text-align: center;" | 02 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 100pA/~10 kHz | |||
| style="text-align: center;" | 2 mm centred | |||
| style="text-align: center;" | run02_2mmCCol_100pA_TrackTrig | |||
| Lower current check. | |||
|- | |||
| style="text-align: center;" | 03 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 80pA/~10 kHz | |||
| style="text-align: center;" | 2 mm centred | |||
| style="text-align: center;" | run03_2mmCCol_80pA_TrackTrig | |||
| Repeat of run 2, no ADC data. | |||
|- | |||
| style="text-align: center;" | 04 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 90pA/~10 kHz | |||
| style="text-align: center;" | 2 mm centred, 0.95mm PMMA | |||
| style="text-align: center;" | run04_2mmCCol_90pA_1mmPMMA | |||
| Tracker triggered, 0.95mm PMMA. | |||
|- | |||
| style="text-align: center;" | 05 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 70pA/~10 kHz | |||
| style="text-align: center;" | 2 mm centred, 0.95mm PMMA | |||
| style="text-align: center;" | run05_2mmCCol_70pA_1mmPMMA | |||
| Repeat of run 4 (ADC data for runs 4, 5 and 6). | |||
|- | |||
| style="text-align: center;" | 06 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 80pA/~10 kHz | |||
| style="text-align: center;" | 2 mm centred, 0.95mm PMMA | |||
| style="text-align: center;" | run06_2mmCCol_80pA_1mmPMMA | |||
| Repeat of run 4 (ADC data for runs 4, 5 and 6). | |||
|- | |||
| style="text-align: center;" | 07 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 60pA/~10 kHz | |||
| style="text-align: center;" | 2 mm centred, 2.01mm PMMA | |||
| style="text-align: center;" | run07_2mmCCol_60pA_2mmPMMA | |||
| Tracker triggered, 2.01mm PMMA. | |||
|- | |||
| style="text-align: center;" | 08 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 80pA/~10 kHz | |||
| style="text-align: center;" | 2 mm centred, 2.01mm PMMA | |||
| style="text-align: center;" | run08_2mmCCol_80pA_2mmPMMA | |||
| Repeat of run 7 (ADC data for runs 7, 8 and 9). | |||
|- | |||
| style="text-align: center;" | 09 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 90pA/~10 kHz | |||
| style="text-align: center;" | 2 mm centred, 2.01mm PMMA | |||
| style="text-align: center;" | run09_2mmCCol_90pA_2mmPMMA | |||
| Repeat of run 7 (ADC data for runs 7, 8 and 9). | |||
|- | |||
| style="text-align: center;" | 10 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 80pA/~10 kHz | |||
| style="text-align: center;" | 2 mm centred, 2.96mm PMMA | |||
| style="text-align: center;" | run10_2mmCCol_80pA_3mmPMMA | |||
| Tracker triggered, 2.96mm PMMA. | |||
|- | |||
| style="text-align: center;" | 11 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 85pA/~10 kHz | |||
| style="text-align: center;" | 2 mm centred, 2.96mm PMMA | |||
| style="text-align: center;" | run11_2mmCCol_85pA_3mmPMMA | |||
| Repeat of run 10 (ADC data for runs 10, 11 and 12). | |||
|- | |||
| style="text-align: center;" | 12 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 90pA/~10 kHz | |||
| style="text-align: center;" | 2 mm centred, 2.96mm PMMA | |||
| style="text-align: center;" | run12_2mmCCol_90pA_3mmPMMA | |||
| Repeat of run 10 (ADC data for runs 10, 11 and 12). | |||
|- | |||
| style="text-align: center;" | 13 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 50pA/~10 kHz | |||
| style="text-align: center;" | 2 mm centred, 3.95mm PMMA | |||
| style="text-align: center;" | run13_2mmCCol_50pA_4mmPMMA | |||
| Tracker triggered, 3.95mm PMMA. | |||
|- | |||
| style="text-align: center;" | 14 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 77pA/~10 kHz | |||
| style="text-align: center;" | 2 mm centred, 3.95mm PMMA | |||
| style="text-align: center;" | run14_2mmCCol_77pA_4mmPMMA | |||
| Repeat of run 13 (ADC data for runs 13, 14 and 15). | |||
|- | |||
| style="text-align: center;" | 15 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 90pA/~10 kHz | |||
| style="text-align: center;" | 2 mm centred, 3.95mm PMMA | |||
| style="text-align: center;" | run15_2mmCCol_90pA_4mmPMMA | |||
| Repeat of run 13 (ADC data for runs 13, 14 and 15). | |||
|- | |||
| style="text-align: center;" | 16 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 70pA/~10 kHz | |||
| style="text-align: center;" | 2 mm centred, 4.90mm PMMA | |||
| style="text-align: center;" | run16_2mmCCol_70pA_5mmPMMA | |||
| Tracker triggered, 4.90mm PMMA. Large current fluctuations: check shape of ADC spectrum. | |||
|- | |||
| style="text-align: center;" | 17 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 65pA/~10 kHz | |||
| style="text-align: center;" | 2 mm centred, 4.90mm PMMA | |||
| style="text-align: center;" | run17_2mmCCol_65pA_5mmPMMA | |||
| Repeat of run 16 (ADC data for runs 16, 17 and 18). | |||
|- | |||
| style="text-align: center;" | 18 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 60pA/~10 kHz | |||
| style="text-align: center;" | 2 mm centred, 4.90mm PMMA | |||
| style="text-align: center;" | run18_2mmCCol_60pA_5mmPMMA | |||
| Repeat of run 16 (ADC data for runs 16, 17 and 18). | |||
|- | |||
| style="text-align: center;" | 19 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 50pA/~10 kHz | |||
| style="text-align: center;" | 2 mm centred, 5.70mm PMMA | |||
| style="text-align: center;" | run19_2mmCCol_50pA_6mmPMMA | |||
| Tracker triggered, 5.70mm PMMA. | |||
|- | |||
| style="text-align: center;" | 20 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 50pA/~10 kHz | |||
| style="text-align: center;" | 2 mm centred, 5.70mm PMMA | |||
| style="text-align: center;" | run20_2mmCCol_50pA_6mmPMMA | |||
| Repeat of run 19 (ADC data for runs 19, 20 and 21). | |||
|- | |||
| style="text-align: center;" | 21 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 60pA/~10 kHz | |||
| style="text-align: center;" | 2 mm centred, 5.70mm PMMA | |||
| style="text-align: center;" | run21_2mmCCol_60pA_6mmPMMA | |||
| Repeat of run 19 (ADC data for runs 19, 20 and 21). | |||
|- | |||
| style="text-align: center;" | 22 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 90-140pA/~10 kHz | |||
| style="text-align: center;" | 2 mm with 6mm offset in +ve x, 5.70mm PMMA. | |||
| style="text-align: center;" | run22_2mmOSCol_100pA_6mmPMMA | |||
| Tracker triggered, 2 mm collimator with 6 mm offset in +ve x direction; 5.70 mm PMMA absorber. | |||
|- | |||
| style="text-align: center;" | 23 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 85-120pA/~10 kHz | |||
| style="text-align: center;" | 2 mm with 6mm offset, 5.70mm PMMA | |||
| style="text-align: center;" | run23_2mmOSCol_100pA_6mmPMMA | |||
| Repeat of run 22 (ADC data for runs 22, 23 and 24). | |||
|- | |||
| style="text-align: center;" | 24 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 10-75pA/~10 kHz | |||
| style="text-align: center;" | 2 mm with 6mm offset, 5.70mm PMMA | |||
| style="text-align: center;" | run24_2mmOSCol_70pA_6mmPMMA | |||
| Repeat of run 22 (ADC data for runs 22, 23 and 24). | |||
|- | |||
| style="text-align: center;" | 25 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 40-50pA/~10 kHz | |||
| style="text-align: center;" | 2 mm pair, centres 12mm apart in x, 5.70mm PMMA | |||
| style="text-align: center;" | run25_2mmSBCol_40pA_6mmPMMA | |||
| Tracker triggered, 2 mm hole pair collimator with centres 12 mm apart in x direction, 5.70 mm PMMA absorber. | |||
|- | |||
| style="text-align: center;" | 26 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 45pA/~10 kHz | |||
| style="text-align: center;" | 2 mm pair, centres 12mm apart, 5.70mm PMMA | |||
| style="text-align: center;" | run26_2mmSBCol_45pA_6mmPMMA | |||
| Repeat of run 25 (ADC data for runs 25, 26 and 27). | |||
|- | |||
| style="text-align: center;" | 27 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 45-55pA/~10 kHz | |||
| style="text-align: center;" | 2 mm pair, centres 12mm apart, 5.70mm PMMA | |||
| style="text-align: center;" | run27_2mmSBCol_50pA_6mmPMMA | |||
| Repeat of run 25 (ADC data for runs 25, 26 and 27). | |||
|- | |||
| style="text-align: center;" | 28 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 70-90pA/~10 kHz | |||
| style="text-align: center;" | 2 mm with 6mm offset in +ve x. | |||
| style="text-align: center;" | run28_2mmOSCol_70pA_TrackTrig | |||
| Tracker triggered, 2 mm collimator with 6 mm offset in +ve x direction. | |||
|- | |||
| style="text-align: center;" | 29 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 60pA/~10 kHz | |||
| style="text-align: center;" | 2 mm with 6mm offset | |||
| style="text-align: center;" | run29_2mmOSCol_60pA_TrackTrig | |||
| Repeat of run 28 (ADC data for runs 28, 29 and 30). | |||
|- | |||
| style="text-align: center;" | 30 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 60pA/~10 kHz | |||
| style="text-align: center;" | 2 mm with 6mm offset | |||
| style="text-align: center;" | run30_2mmOSCol_60pA_TrackTrig | |||
| Repeat of run 28 (ADC data for runs 28, 29 and 30). | |||
|- | |||
| style="text-align: center;" | 31 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 40-50pA/~10 kHz | |||
| style="text-align: center;" | 2 mm pair, centres 12mm apart in x. | |||
| style="text-align: center;" | run31_2mmSBCol_46pA_TrackTrig | |||
| Tracker triggered, 2 mm hole pair collimator with centres 12 mm apart in x direction. | |||
|- | |||
| style="text-align: center;" | 32 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 40-50pA/~10 kHz | |||
| style="text-align: center;" | 2 mm pair, centres 12mm apart | |||
| style="text-align: center;" | run32_2mmSBCol_45pA_TrackTrig | |||
| Repeat of run 31 (ADC data for runs 31, 32 and 33). | |||
|- | |||
| style="text-align: center;" | 33 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 35-40pA/~10 kHz | |||
| style="text-align: center;" | 2 mm pair, centres 12mm apart | |||
| style="text-align: center;" | run33_2mmSBCol_40pA_TrackTrig | |||
| Repeat of run 31 (ADC data for runs 31, 32 and 33); large current dropout. | |||
|- | |||
| style="text-align: center;" | 34 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 10pA/~10 kHz | |||
| style="text-align: center;" | 2 mm pair, centres 12mm apart in x. | |||
| style="text-align: center;" | run34_2mmSBCol_10pA_TrackTrig | |||
| Tracker triggered, 2 mm hole pair collimator with centres 12 mm apart in x direction; lower current repeat of 2 hole collimator. | |||
|- | |||
| style="text-align: center;" | 35 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 10pA/~10 kHz | |||
| style="text-align: center;" | 2 mm pair, centres 12mm apart | |||
| style="text-align: center;" | run35_2mmSBCol_10pA_TrackTrig | |||
| Repeat of run 34 (ADC data for runs 34, 35 and 36). | |||
|- | |||
| style="text-align: center;" | 36 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 8pA/~10 kHz | |||
| style="text-align: center;" | 2 mm pair, centres 12mm apart | |||
| style="text-align: center;" | run36_2mmSBCol_8pA_TrackTrig | |||
| Repeat of run 34 (ADC data for runs 34, 35 and 36). | |||
|- | |||
| style="text-align: center;" | 37 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 20-24pA/~10 kHz | |||
| style="text-align: center;" | 2 mm pair, centres 12mm apart with 3.95mm PMMA over 1 hole. | |||
| style="text-align: center;" | run37_2mmEPCol_20pA_4mmPMMA | |||
| Tracker triggered, 2 mm hole pair collimator with centres 12 mm apart in x direction; left-hand hole covered by 3.95&mm PMMA. | |||
|- | |||
| style="text-align: center;" | 38 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 11pA/~10 kHz | |||
| style="text-align: center;" | 2 mm pair, centres 12mm apart with 3.95mm PMMA over 1 hole. | |||
| style="text-align: center;" | run38_2mmEPCol_11pA_4mmPMMA | |||
| Repeat of run 37 (ADC data for runs 37, 38 and 39). | |||
|- | |||
| style="text-align: center;" | 39 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 9pA/~10 kHz | |||
| style="text-align: center;" | 2 mm pair, centres 12mm apart with 3.95mm PMMA over 1 hole. | |||
| style="text-align: center;" | run39_2mmEPCol_9pA_4mmPMMA | |||
| Repeat of run 37 (ADC data for runs 37, 38 and 39). | |||
|- | |||
| style="text-align: center;" | 40 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 20-70pA/~10 kHz | |||
| style="text-align: center;" | 2 mm pair, centres 12mm apart with 3.95mm PMMA over 1 hole. | |||
| style="text-align: center;" | run40_2mmEPCol_40pA_4mmPMMA | |||
| Tracker triggered, 2 mm hole pair collimator with centres 12 mm apart in x direction; left-hand hole covered by 3.95&mm PMMA. | |||
|- | |||
| style="text-align: center;" | 41 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 40-120pA/~10 kHz | |||
| style="text-align: center;" | 2 mm pair, centres 12mm apart with 3.95mm PMMA over 1 hole. | |||
| style="text-align: center;" | run41_2mmEPCol_40pA_4mmPMMA | |||
| Repeat of run 40 (ADC data for runs 40, 41 and 42); large current fluctuations. | |||
|- | |||
| style="text-align: center;" | 42 | |||
| style="text-align: center;" | -900 | |||
| style="text-align: center;" | 70, +ve | |||
| style="text-align: center;" | Tracker (C2) | |||
| style="text-align: center;" | 15-45pA/~10 kHz | |||
| style="text-align: center;" | 2 mm pair, centres 12mm apart with 3.95mm PMMA over 1 hole. | |||
| style="text-align: center;" | run42_2mmEPCol_30pA_4mmPMMA | |||
| Repeat of run 40 (ADC data for runs 40, 41 and 42). | |||
|- | |- | ||
|} | |} |
Latest revision as of 10:30, 2 November 2018
Beam tests of the single module with PRaVDA Silicon strip tracker with trigger pulse output at the Birmingham Cyclotron with a 36 MeV beam.
Equipment List
Item | Notes |
---|---|
Single Module Scintillator Block | 3cm x 3cm x 5cm standard scintillator block. Optical gel required to couple Scintillator to PMT |
Photomultiplier Tube | Hamamatsu R13089 2" |
Portable Enclosure | Modified Peli 1510 Waterproof Wheeled Equipment Case. 230 x 555 x 350mm |
LeCroy HDO6104 Oscilloscope | Records PMT output, connected via SMA-BNC conversion. |
Caen DT5751 Digitiser | Records PMT output, connected via SMA-MCX conversion. Used channel 3. |
DAQ Laptop | Controls Caen HV supply; records data from Caen digitiser. |
Caen NDT1470 HV Supply | Supplies HV to PMT |
PRaVDA Strip Tracker | PRaVDA Unit A, 3x150um Si strip sensors with 90.4um strip pitch. Firmware version update to yield square pulse out when 2 or 3 layers have a strip over threshold per read out cycle |
Control Laptop x 3 | Two used for Remote Desktop control of LeCroy Oscilloscope and DAQ Laptop. Third used for taking down notes. |
Ethernet Switcher | Set in experiment room and connected LeCroy Oscilloscope and DAQ laptop. Output sent to control room Network Hub. |
Network Hub | Set in control room to take output from ethernet switcher. Control laptops connected via ethernet or 5GHz WiFi. |
Ethernet Cable x 3 | To connect 1 control laptop, the LeCroy HDO6104 and DAQ Laptop to their respective network hubs. |
Ethernet Cable | To connect control Network Hub to Ethernet Switcher. |
Male-to-Male SMA Cable | Signal from enclosure port to Splitter |
DC-4200 MHz Signal Splitter | Splits SMA signal to two SMA signals – sent to Oscilloscope and Digitiser. |
Male SMA to Male BNC cable | Sends 1 output from splitter to Oscilloscope. |
Male SMA to Male MCX cable | Sends 1 output from splitter to Digitiser |
USB Cable | Caen HV unit to DAQ Laptop |
Gloves | For handling scintillator |
Optical gel | For coupling scintillator to PMT |
Wipes | For removing optical gel |
Experiment List
28 MeV beam Scattering Foil in 2mm diameter collimator 10kHz rate ~= 160pA onto Transmission Chamber before 2mm collimation
- Bragg Peak Measurements
- Combined scope / digitizer measurements with pristine bragg peak, 100mV trigger threshold
- Combined scope / digitizer measurements with pristine bragg peak, varying trigger threshold
- Degrade beam energy and measure response with 1-6mm PMMA before calorimeter
- PRaVDA box with 2mm collimator, trigger from scope
- PRaVDA box with 2mm collimator for trigger output
36 MeV beam
Scattering Foil in
2mm diameter collimator
10kHz rate ~= 160pA onto Transmission Chamber before 2mm collimation
- Bragg Peak Measurements
- Combined scope / digitizer measurements with pristine bragg peak, 100mV trigger threshold
- Combined scope / digitizer measurements with pristine bragg peak, varying trigger threshold
- Degrade beam energy and measure response with 1-6mm PMMA before calorimeter
- Offset 2mm collimator (by 6mm) for optimal trigger threshold
- Double 2mm collimator (12mm separated) for optimal trigger threshold
- PRaVDA box with 2mm collimator, trigger from scope
- PRaVDA box with 2mm collimator for trigger output
- Degrade beam with 1-6mm PMMA with tracker
Experiment Log
25/10/18
- Set up apparatus as described in equipment list.
- Beam experienced difficulties, delaying data collection.
- Tracker operates on a 26MHz clock, and a trigger is sent whenever 2/3 silicon layers in tracker record events in a timestamp. Trigger tests were conducted with noise, hence a random rate in tests below.
- Tested trigger output of Tracker into Oscilloscope: square wave of period ~15 ns with amplitude ~400 mV. Pulse example recorded on Oscilloscope disk in folder: SShaikh/Tracker Trigger
- Placed signal splitter for trigger to send to both Digitiser and Oscilloscope. Amplitude reduced to ~50%. Pulse example recorded on Oscilloscope disk in folder: SShaikh/Tracker Trigger (Split)
- Split trigger found to be too small for Digitiser (accepts trigger of order ~V), hence Digitiser set to self-trigger mode.
- Aligned calorimeter to tracker, with centre of mylar window and tracker window aligned approximately by eye.
- Performed beam test (had approx 30 mins beam time), details provided below.
- Was found that the tracker was not completely set centre to the beam.
- Error in data taking: Trigger trace was recorded instead of PMT trace.
26/10/18
- Tracker was realigned to be centred to the beam.
- Using Tipp-Ex markers and lable stickers, window to scintillator was aligned more accurately to window of tracker.
- Test of tracker/PMT synchronisation (check tracker trigger pulse/PMT pulse sync and numbers of events in files) with scope vertical scale 50mV/div. Window center corresponds to right edge vertical line intersecting upper edge horizontal line, looking at the window.
- Test of tracker/PMT synchronisation with scope vertical scale 100mV/div, 2mm collimator centered.
- Test of tracker/PMT synchronisation with lower current, 2mm collimator centered.
- Degrade beam energy and measure response with 0.95mm, 2.01mm, 2.96mm, 3.95mm, 4.90mm, 5.70mm PMMA before calorimeter, 2mm collimator centered.
- Offset 2mm collimator (by 6mm in +ve) and measure response with 5.70 mm PMMA absorber.
- Double 2mm collimator (12mm separated in x direction) and measure response with 5.70 mm PMMA absorber.
- Offset 2mm collimator (by 6mm in +ve), no absorber.
- Double 2mm collimator (12mm separated in x direction), no absorber.
- Double 2mm collimator (12mm separated in x direction) repetition with lower current, no absorber.
- Double 2mm collimator (12mm separated in x direction) with left-hand hole covered by 3.95 mm PMMA.
Beam Tests
Thursday 25th October
PMT Current at HV supply -900 V, beam off: 150 μA
28MeV beam. Scope triggered by tracker. Cyclotron issues meant short run (~90 mins). Test of tracker triggering/scope synchronisation.
Scope timebase: 20ns/div; trigger delay 80ns; C1 vertical scale 50mV/div.
Tracker data indicates beam horizontal offset of 12mm; vertical offset of 1.3mm.
Data stored in directory ProCal_Birm_25-10-18
Run Number | High Voltage (V) | Trigger Level (mV) | Trigger Source | Beam Current / Rate | Collimator | Filename | Notes |
---|---|---|---|---|---|---|---|
00 | -900 | 70, +ve | Tracker (C2) | 170pA/~10 kHz | 2 mm centred | run00_2mmCCol_160pA_TrackTrig | First test of tracker/PMT synchronisation |
01 | -900 | 70, +ve | Tracker (C2) | 160pA/~10 kHz | 2 mm centred | run01_2mmCCol_160pA_TrackTrig | Also took Caen ADC data |
02 | -900 | 70, +ve | Tracker (C2) | 160pA/~10 kHz | 2 mm centred | run02_2mmCCol_160pA_TrackTrig | Caen ADC run 01 data also includes run 02 (long run) |
Friday 26th October
PMT Current at HV supply -900 V (actual 899.8V), beam off: 150.15 μA.
36MeV beam: scattering foil inserted to increase beam size; 2 mm on-axis collimator.
Scope triggered by tracker. Scope timebase: 20ns/div; trigger delay 80ns; C1 vertical scale 50mV/div.
Data stored in directory ProCal_Birm_26-10-18
Run Number | High Voltage (V) | Trigger Level (mV) | Trigger Source | Beam Current / Rate | Collimator | Filename | Notes |
---|---|---|---|---|---|---|---|
00 | -900 | 70, +ve | Tracker (C2) | 160-200pA/~10 kHz | 2 mm centred | run00_2mmCCol_160pA_TrackTrig | Test of tracker/PMT synchronisation with 5s tracker run, 50,000 scope segments, all traces saved; check tracker trigger pulse/PMT pulse sync and numbers of events in files. |
01 | -900 | 70, +ve | Tracker (C2) | 160-170pA/~10 kHz | 2 mm centred | run01_2mmCCol_160pA_TrackTrig | Repeat of previous test; scope vertical trace increased to 100mV/div. |
02 | -900 | 70, +ve | Tracker (C2) | 100pA/~10 kHz | 2 mm centred | run02_2mmCCol_100pA_TrackTrig | Lower current check. |
03 | -900 | 70, +ve | Tracker (C2) | 80pA/~10 kHz | 2 mm centred | run03_2mmCCol_80pA_TrackTrig | Repeat of run 2, no ADC data. |
04 | -900 | 70, +ve | Tracker (C2) | 90pA/~10 kHz | 2 mm centred, 0.95mm PMMA | run04_2mmCCol_90pA_1mmPMMA | Tracker triggered, 0.95mm PMMA. |
05 | -900 | 70, +ve | Tracker (C2) | 70pA/~10 kHz | 2 mm centred, 0.95mm PMMA | run05_2mmCCol_70pA_1mmPMMA | Repeat of run 4 (ADC data for runs 4, 5 and 6). |
06 | -900 | 70, +ve | Tracker (C2) | 80pA/~10 kHz | 2 mm centred, 0.95mm PMMA | run06_2mmCCol_80pA_1mmPMMA | Repeat of run 4 (ADC data for runs 4, 5 and 6). |
07 | -900 | 70, +ve | Tracker (C2) | 60pA/~10 kHz | 2 mm centred, 2.01mm PMMA | run07_2mmCCol_60pA_2mmPMMA | Tracker triggered, 2.01mm PMMA. |
08 | -900 | 70, +ve | Tracker (C2) | 80pA/~10 kHz | 2 mm centred, 2.01mm PMMA | run08_2mmCCol_80pA_2mmPMMA | Repeat of run 7 (ADC data for runs 7, 8 and 9). |
09 | -900 | 70, +ve | Tracker (C2) | 90pA/~10 kHz | 2 mm centred, 2.01mm PMMA | run09_2mmCCol_90pA_2mmPMMA | Repeat of run 7 (ADC data for runs 7, 8 and 9). |
10 | -900 | 70, +ve | Tracker (C2) | 80pA/~10 kHz | 2 mm centred, 2.96mm PMMA | run10_2mmCCol_80pA_3mmPMMA | Tracker triggered, 2.96mm PMMA. |
11 | -900 | 70, +ve | Tracker (C2) | 85pA/~10 kHz | 2 mm centred, 2.96mm PMMA | run11_2mmCCol_85pA_3mmPMMA | Repeat of run 10 (ADC data for runs 10, 11 and 12). |
12 | -900 | 70, +ve | Tracker (C2) | 90pA/~10 kHz | 2 mm centred, 2.96mm PMMA | run12_2mmCCol_90pA_3mmPMMA | Repeat of run 10 (ADC data for runs 10, 11 and 12). |
13 | -900 | 70, +ve | Tracker (C2) | 50pA/~10 kHz | 2 mm centred, 3.95mm PMMA | run13_2mmCCol_50pA_4mmPMMA | Tracker triggered, 3.95mm PMMA. |
14 | -900 | 70, +ve | Tracker (C2) | 77pA/~10 kHz | 2 mm centred, 3.95mm PMMA | run14_2mmCCol_77pA_4mmPMMA | Repeat of run 13 (ADC data for runs 13, 14 and 15). |
15 | -900 | 70, +ve | Tracker (C2) | 90pA/~10 kHz | 2 mm centred, 3.95mm PMMA | run15_2mmCCol_90pA_4mmPMMA | Repeat of run 13 (ADC data for runs 13, 14 and 15). |
16 | -900 | 70, +ve | Tracker (C2) | 70pA/~10 kHz | 2 mm centred, 4.90mm PMMA | run16_2mmCCol_70pA_5mmPMMA | Tracker triggered, 4.90mm PMMA. Large current fluctuations: check shape of ADC spectrum. |
17 | -900 | 70, +ve | Tracker (C2) | 65pA/~10 kHz | 2 mm centred, 4.90mm PMMA | run17_2mmCCol_65pA_5mmPMMA | Repeat of run 16 (ADC data for runs 16, 17 and 18). |
18 | -900 | 70, +ve | Tracker (C2) | 60pA/~10 kHz | 2 mm centred, 4.90mm PMMA | run18_2mmCCol_60pA_5mmPMMA | Repeat of run 16 (ADC data for runs 16, 17 and 18). |
19 | -900 | 70, +ve | Tracker (C2) | 50pA/~10 kHz | 2 mm centred, 5.70mm PMMA | run19_2mmCCol_50pA_6mmPMMA | Tracker triggered, 5.70mm PMMA. |
20 | -900 | 70, +ve | Tracker (C2) | 50pA/~10 kHz | 2 mm centred, 5.70mm PMMA | run20_2mmCCol_50pA_6mmPMMA | Repeat of run 19 (ADC data for runs 19, 20 and 21). |
21 | -900 | 70, +ve | Tracker (C2) | 60pA/~10 kHz | 2 mm centred, 5.70mm PMMA | run21_2mmCCol_60pA_6mmPMMA | Repeat of run 19 (ADC data for runs 19, 20 and 21). |
22 | -900 | 70, +ve | Tracker (C2) | 90-140pA/~10 kHz | 2 mm with 6mm offset in +ve x, 5.70mm PMMA. | run22_2mmOSCol_100pA_6mmPMMA | Tracker triggered, 2 mm collimator with 6 mm offset in +ve x direction; 5.70 mm PMMA absorber. |
23 | -900 | 70, +ve | Tracker (C2) | 85-120pA/~10 kHz | 2 mm with 6mm offset, 5.70mm PMMA | run23_2mmOSCol_100pA_6mmPMMA | Repeat of run 22 (ADC data for runs 22, 23 and 24). |
24 | -900 | 70, +ve | Tracker (C2) | 10-75pA/~10 kHz | 2 mm with 6mm offset, 5.70mm PMMA | run24_2mmOSCol_70pA_6mmPMMA | Repeat of run 22 (ADC data for runs 22, 23 and 24). |
25 | -900 | 70, +ve | Tracker (C2) | 40-50pA/~10 kHz | 2 mm pair, centres 12mm apart in x, 5.70mm PMMA | run25_2mmSBCol_40pA_6mmPMMA | Tracker triggered, 2 mm hole pair collimator with centres 12 mm apart in x direction, 5.70 mm PMMA absorber. |
26 | -900 | 70, +ve | Tracker (C2) | 45pA/~10 kHz | 2 mm pair, centres 12mm apart, 5.70mm PMMA | run26_2mmSBCol_45pA_6mmPMMA | Repeat of run 25 (ADC data for runs 25, 26 and 27). |
27 | -900 | 70, +ve | Tracker (C2) | 45-55pA/~10 kHz | 2 mm pair, centres 12mm apart, 5.70mm PMMA | run27_2mmSBCol_50pA_6mmPMMA | Repeat of run 25 (ADC data for runs 25, 26 and 27). |
28 | -900 | 70, +ve | Tracker (C2) | 70-90pA/~10 kHz | 2 mm with 6mm offset in +ve x. | run28_2mmOSCol_70pA_TrackTrig | Tracker triggered, 2 mm collimator with 6 mm offset in +ve x direction. |
29 | -900 | 70, +ve | Tracker (C2) | 60pA/~10 kHz | 2 mm with 6mm offset | run29_2mmOSCol_60pA_TrackTrig | Repeat of run 28 (ADC data for runs 28, 29 and 30). |
30 | -900 | 70, +ve | Tracker (C2) | 60pA/~10 kHz | 2 mm with 6mm offset | run30_2mmOSCol_60pA_TrackTrig | Repeat of run 28 (ADC data for runs 28, 29 and 30). |
31 | -900 | 70, +ve | Tracker (C2) | 40-50pA/~10 kHz | 2 mm pair, centres 12mm apart in x. | run31_2mmSBCol_46pA_TrackTrig | Tracker triggered, 2 mm hole pair collimator with centres 12 mm apart in x direction. |
32 | -900 | 70, +ve | Tracker (C2) | 40-50pA/~10 kHz | 2 mm pair, centres 12mm apart | run32_2mmSBCol_45pA_TrackTrig | Repeat of run 31 (ADC data for runs 31, 32 and 33). |
33 | -900 | 70, +ve | Tracker (C2) | 35-40pA/~10 kHz | 2 mm pair, centres 12mm apart | run33_2mmSBCol_40pA_TrackTrig | Repeat of run 31 (ADC data for runs 31, 32 and 33); large current dropout. |
34 | -900 | 70, +ve | Tracker (C2) | 10pA/~10 kHz | 2 mm pair, centres 12mm apart in x. | run34_2mmSBCol_10pA_TrackTrig | Tracker triggered, 2 mm hole pair collimator with centres 12 mm apart in x direction; lower current repeat of 2 hole collimator. |
35 | -900 | 70, +ve | Tracker (C2) | 10pA/~10 kHz | 2 mm pair, centres 12mm apart | run35_2mmSBCol_10pA_TrackTrig | Repeat of run 34 (ADC data for runs 34, 35 and 36). |
36 | -900 | 70, +ve | Tracker (C2) | 8pA/~10 kHz | 2 mm pair, centres 12mm apart | run36_2mmSBCol_8pA_TrackTrig | Repeat of run 34 (ADC data for runs 34, 35 and 36). |
37 | -900 | 70, +ve | Tracker (C2) | 20-24pA/~10 kHz | 2 mm pair, centres 12mm apart with 3.95mm PMMA over 1 hole. | run37_2mmEPCol_20pA_4mmPMMA | Tracker triggered, 2 mm hole pair collimator with centres 12 mm apart in x direction; left-hand hole covered by 3.95&mm PMMA. |
38 | -900 | 70, +ve | Tracker (C2) | 11pA/~10 kHz | 2 mm pair, centres 12mm apart with 3.95mm PMMA over 1 hole. | run38_2mmEPCol_11pA_4mmPMMA | Repeat of run 37 (ADC data for runs 37, 38 and 39). |
39 | -900 | 70, +ve | Tracker (C2) | 9pA/~10 kHz | 2 mm pair, centres 12mm apart with 3.95mm PMMA over 1 hole. | run39_2mmEPCol_9pA_4mmPMMA | Repeat of run 37 (ADC data for runs 37, 38 and 39). |
40 | -900 | 70, +ve | Tracker (C2) | 20-70pA/~10 kHz | 2 mm pair, centres 12mm apart with 3.95mm PMMA over 1 hole. | run40_2mmEPCol_40pA_4mmPMMA | Tracker triggered, 2 mm hole pair collimator with centres 12 mm apart in x direction; left-hand hole covered by 3.95&mm PMMA. |
41 | -900 | 70, +ve | Tracker (C2) | 40-120pA/~10 kHz | 2 mm pair, centres 12mm apart with 3.95mm PMMA over 1 hole. | run41_2mmEPCol_40pA_4mmPMMA | Repeat of run 40 (ADC data for runs 40, 41 and 42); large current fluctuations. |
42 | -900 | 70, +ve | Tracker (C2) | 15-45pA/~10 kHz | 2 mm pair, centres 12mm apart with 3.95mm PMMA over 1 hole. | run42_2mmEPCol_30pA_4mmPMMA | Repeat of run 40 (ADC data for runs 40, 41 and 42). |