== Friday 23rd - Saturday 24th of March 2018 ==
== Friday 23rd - Saturday 24th of March 2018 ==
=== Pre-data Taking Information and Notes ===
=== Pre-data Taking Information and Notes ===
* The tests were carried out with an energy of '''252.7 MeV''' with a '''5 second beam spill''' and a range of intensities.
* The tests were carried out with an energy of '''252.7 MeV''' with a '''5 second beam spill''' and six different beam intensities.
* No collimators were used at any point.
* No collimators were used at any point.
* The setup consisted of:
* The setup consisted of:
Line 26: Line 26:
!PMT Current, uA
!PMT Current, uA
!Beam Intensity
!Beam Intensity
!Rate (Measured on Digitiser) and Notes
!Rate (Measured on Digitiser)
!Digitiser Settings
!Digitiser Settings and Notes
|-
|-


Line 50: Line 50:
| 006 || -500 || 82.6 || Lowest || 0.3 - 5 kHz || Threshold: 10 LSB. Too close to the threshold, but cannot lower more due to noise floor.
| 006 || -500 || 82.6 || Lowest || 0.3 - 5 kHz || Threshold: 10 LSB. Too close to the threshold, but cannot lower more due to noise floor.
|-
|-
| 007 || -600 ||  || Increased by 10 - 15 times of original || 5 - 30 kHz || Threshold: 20 LSB
|-
| 008 || -700 || 116.3  || Increased by 10 - 15 times of original || 5 - 30 kHz || As above
|-
| 009 || -800 || 133.2 || Increased by 10 - 15 times of original || 5 - 30 kHz || As above
|-
| 010 || -900 || 150.1 || Increased by 10 - 15 times of original || 5 - 30 kHz || As above
|-
| 011 || -900 || || Increased by factor of 3 from 2nd intensity || 150 - 200 kHz || As above
|-
| 012 || -800 || 133.2 || Increased by factor of 3 from 2nd intensity || 150 - 200 kHz || As above
|-
| 013 || -700 || 116.55 || Increased by factor of 3 from 2nd intensity || 150 - 200 kHz || As above
|-
| 014 || -600 || 99.45 || Increased by factor of 3 from 2nd intensity || 150 - 200 kHz || As above
|-
| 015 || -600 || || Increased by factor of 2 from 3rd intensity || 300 - 340 kHz || As above
|-
| 016 || -700 || 116.3 || Increased by factor of 2 from 3rd intensity || 300 - 340 kHz || As above
|-
| 017 || -800 || || Increased by factor of 2 from 3rd intensity || 300 - 340 kHz || As above
|-
| 018 || -900 || 150.1 || Increased by factor of 2 from 3rd intensity || 300 - 340 kHz || As above
|-
| 019 || -900 || || Increased by factor of 2 from 4th intensity || 600 - 800 kHz || As above. No sensible peak at this HV.
|-
| 020 || -800 || 133.2 || Increased by factor of 2 from 4th intensity || 600 - 800 kHz || As above
|-
| 021 || -700 || 116.35 || Increased by factor of 2 from 4th intensity || 600 - 800 kHz || As above
|-
| 022 || -600 || 99.35 || Increased by factor of 2 from 4th intensity || 600 - 800 kHz || As above
|-
| 023 || -600 || || Increased by factor of 2 from 5th intensity || 1.2 MHz || As above. Pile up is becoming more apparent.
|-
| 024 || -550 || 91.25 || Increased by factor of 2 from 5th intensity || 1.2 MHz || As above
|-
| 025 || -550 || || Increased by factor of 2 from 5th intensity || 1.2 MHz || Threshold: 10 LSB
|-
| 026 || -580 || 96.1 || Increased by factor of 2 from 5th intensity || 1.2 MHz || Threshold: 10 LSB
|-
| 027 || -580 || || Increased by factor of 2 from 5th intensity || 1.2 MHz || Threshold: 5 LSB
|}
----
<center>'''End of test beam.'''</center>

Latest revision as of 11:00, 28 March 2018

Beam tests of the UCL single module at medAustron.

Friday 23rd - Saturday 24th of March 2018

Pre-data Taking Information and Notes

  • The tests were carried out with an energy of 252.7 MeV with a 5 second beam spill and six different beam intensities.
  • No collimators were used at any point.
  • The setup consisted of:
    • 2" Hamamatsu R13089-100-11 PMT with negative HV active divider base (made by Hamamatsu)
    • coupled with Eljen optical gel to a
    • 3 cm x 3 cm x 45 cm cuboid ENVINET standard scintillator wrapped in two layers 6 um (each) of Mylar
  • The "pensioner" 5 dynode stage PMT was also briefly tested, but no sensible data was able to be acquired.
  • The digitiser always operated in Self Trigger mode.
  • Differences since last medAustron test beam:
    • Short cables used for signals and SMA instead of BNC, controlling the acquisition remotely.
    • Tests carried out at a range of HVs.
  • Working directory where digitiser data is stored:

C:\Users\abf\Desktop\digitiser_data\medAustron\mar18\ham

Data Taking

Run number High Voltage, V PMT Current, uA Beam Intensity Rate (Measured on Digitiser) Digitiser Settings and Notes
N/A -900 150.1 (beam off) N/A N/A DC Offset: 0, Rec. Length: 396 ns, Threshold: 40 LSB, Gate Offset: 1 ns, Long Gate: 396 ns, BL Mean: 8 ns, Charge. Sensitivty: 20 fC/LSB, Pre-trigger: 9 ns
001 -900 150.1 Lowest 0.3 - 5 kHz DC Offset: -30, Threshold: 40 LSB, Gate Offset: 20 ns, Long Gate: 150 ns
002 -800 133.2 Lowest 0.3 - 5 kHz As above
003 -700 116.35 Lowest 0.3 - 5 kHz As above
004 -600 99.45 Lowest 0.3 - 5 kHz Beam dropped during running!
005 -600 99.45 Lowest 0.3 - 5 kHz Repeat run with beam back.
006 -500 82.6 Lowest 0.3 - 5 kHz Threshold: 10 LSB. Too close to the threshold, but cannot lower more due to noise floor.
007 -600 Increased by 10 - 15 times of original 5 - 30 kHz Threshold: 20 LSB
008 -700 116.3 Increased by 10 - 15 times of original 5 - 30 kHz As above
009 -800 133.2 Increased by 10 - 15 times of original 5 - 30 kHz As above
010 -900 150.1 Increased by 10 - 15 times of original 5 - 30 kHz As above
011 -900 Increased by factor of 3 from 2nd intensity 150 - 200 kHz As above
012 -800 133.2 Increased by factor of 3 from 2nd intensity 150 - 200 kHz As above
013 -700 116.55 Increased by factor of 3 from 2nd intensity 150 - 200 kHz As above
014 -600 99.45 Increased by factor of 3 from 2nd intensity 150 - 200 kHz As above
015 -600 Increased by factor of 2 from 3rd intensity 300 - 340 kHz As above
016 -700 116.3 Increased by factor of 2 from 3rd intensity 300 - 340 kHz As above
017 -800 Increased by factor of 2 from 3rd intensity 300 - 340 kHz As above
018 -900 150.1 Increased by factor of 2 from 3rd intensity 300 - 340 kHz As above
019 -900 Increased by factor of 2 from 4th intensity 600 - 800 kHz As above. No sensible peak at this HV.
020 -800 133.2 Increased by factor of 2 from 4th intensity 600 - 800 kHz As above
021 -700 116.35 Increased by factor of 2 from 4th intensity 600 - 800 kHz As above
022 -600 99.35 Increased by factor of 2 from 4th intensity 600 - 800 kHz As above
023 -600 Increased by factor of 2 from 5th intensity 1.2 MHz As above. Pile up is becoming more apparent.
024 -550 91.25 Increased by factor of 2 from 5th intensity 1.2 MHz As above
025 -550 Increased by factor of 2 from 5th intensity 1.2 MHz Threshold: 10 LSB
026 -580 96.1 Increased by factor of 2 from 5th intensity 1.2 MHz Threshold: 10 LSB
027 -580 Increased by factor of 2 from 5th intensity 1.2 MHz Threshold: 5 LSB

End of test beam.