Quality Assurance: Difference between revisions

From PBTWiki
Jump to navigation Jump to search
No edit summary
No edit summary
 
(11 intermediate revisions by the same user not shown)
Line 12: Line 12:
=== Commercially Available Devices ===
=== Commercially Available Devices ===
====IBA Dosimetry====
====IBA Dosimetry====
* [https://www.iba-dosimetry.com/product/lynx-pt Lynx/Sphinx]
** Usage: Daily QA, Monthly QA, Annual QA, Commissioning
** For performing spot size and position measurements and numerous energies.
** 0.4 mm thick and 300 x 300 mm^2 Ga- based plastic scintillator screen, with mirror and CCD camera.
**Technical specs:
*** Pixel size: 4.65 x 4.65 um²
*** Effective spacial resolution: 0.5 mm
*** Video camera: 1024 x 1024, 12-bit resolution
*** Sampling time: 7 images / s
*** Dynamic range can be modified through CCD camera gain settings and lens aperture.
*** Total dimensions 360 x 370 x 600 mm
** Sphinx phantom consists of wedges for range consistency checks for 4 energies (no direct range measurement) and absolute dose measurements with [https://www.iba-dosimetry.com/product/ppc05-plane-parallel-chamber PPC05 Plane Parallel Chamber].
** Used with myQA machine software: myQA Machines Software
* [https://www.iba-dosimetry.com/product/sphinx-compact Sphinx Compact]
** Usage: Daily QA, Monthly QA, Annual QA, Commissioning
** Similar to Lynx detector and Sphinx phantom, but is instead an Silicon flat-panel detector.
** Advertised as being able to complete daily QA measurements in <10 mins.
** Technical specs:
*** 200 x 200 mm^2 flat-panel detector.
*** 0.2 mm spatial resolution
*** 16 bit depth
*** Six different gain values are available: 0.25, 0.5, 1, 2, 4, and 8 pF
** Interval time can be adjusted in 4 steps (100, 200, 400, and 800 ms) - equivalent to dose rates of 0.4 - 4000 Gy/min
*** Approx 11 kg weight.
** Phantom consists of wedges for range consistency checks for 3 energies (no direct range measurement) and absolute dose measurements with [https://www.iba-dosimetry.com/product/ppc05-plane-parallel-chamber PPC05 Plane Parallel Chamber].
** Used with myQA machine software: myQA Machines Software
** https://www.iba-dosimetry.com/fileadmin/user_upload/products/03_proton_therapy/Sphinx_Compact/J_Applied_Clin_Med_Phys_-_2019_-_Rana_-_Development_and_long%E2%80%90term_stability_of_a_comprehensive_daily_QA_program_for_a__1_.pdf
* [https://www.iba-dosimetry.com/product/ppc05-plane-parallel-chamber PPC05 Plane Parallel Chamber]
** Usage: Daily QA, Monthly QA, Annual QA, Commissioning
** Used with reference class electrometer [https://www.iba-dosimetry.com/product/dose-x Dose-X]
* [https://www.iba-dosimetry.com/product/myqa-phoenix myQA Phoenix]
** Usage: Monthly QA, Annual QA, Commissioning
**
* [https://www.iba-dosimetry.com/product/myqa-matrixx-air myQA MatriXX AiR]
** Usage: PSQA
** Wireless ionisation chamber array used for planar verification of of patient treatment plans.
** WiFi and battery operation mode available.
** Technical specs:
*** 1521 air-vented ICs
*** 25 × 25 cm² area
*** 6.5 mm center-to-center resolution
*** Minimal sampling time of 20 ms (50 Hz frame rate)
** Reads out 2D dose distribution at different depths using a
* [https://www.iba-dosimetry.com/product/giraffe Giraffe]
** Usage: Monthly QA, Annual QA, Commissioning
* [https://www.iba-dosimetry.com/product/zebra Zebra]
** Usage: Monthly QA, Annual QA, Commissioning
* [https://www.iba-dosimetry.com/product/stingray StingRay]
** Usage: Monthly QA, Annual QA, Commissioning
====PTW====
====PTW====
* [https://www.ptwdosimetry.com/en/products/octavius-detector-1500xdr Octavius Detector 1500XDR]
* [https://www.ptwdosimetry.com/en/products/octavius-detector-1600xdr Octavius Detector 1600XDR]
* [https://www.ptwdosimetry.com/en/products/bragg-peak-chamber-34070 Bragg Peak Chamber 34070]
* [https://www.ptwdosimetry.com/en/products/bragg-peak-chamber-34080 Bragg Peak Chamber 34080]
* [https://www.ptwdosimetry.com/en/products/bragg-peak-150 Bragg Peak 150]
====Sun Nuclear====
====Sun Nuclear====
* [https://www.sunnuclear.com/products/daily-qa-3 Daily QA 3]
* [https://www.sunnuclear.com/products/ic-profiler IC Profiler]
====DET.TEC.TOR====
====DET.TEC.TOR====
====Logos Systems====
====Logos Systems====
====MedScint====
====MedScint====

Latest revision as of 16:42, 10 August 2026

A reading list for QA devices

Articles on Detectors

  • "Dosimetry for ion beam therapy" (2010)

This paper looks at current methods and standards for PBT dosimetry. It also gives advantages and disadvantages of each method. This paper was particularly useful for looking at film dosimetry and ionization chambers. https://iopscience.iop.org/article/10.1088/0031-9155/55/21/R01/meta

  • "A critical review of the practices of proton daily quality assurance programs" (2021)

https://dx.doi.org/10.21037/TRO-21-11

Commercially Available Devices

IBA Dosimetry

  • Lynx/Sphinx
    • Usage: Daily QA, Monthly QA, Annual QA, Commissioning
    • For performing spot size and position measurements and numerous energies.
    • 0.4 mm thick and 300 x 300 mm^2 Ga- based plastic scintillator screen, with mirror and CCD camera.
    • Technical specs:
      • Pixel size: 4.65 x 4.65 um²
      • Effective spacial resolution: 0.5 mm
      • Video camera: 1024 x 1024, 12-bit resolution
      • Sampling time: 7 images / s
      • Dynamic range can be modified through CCD camera gain settings and lens aperture.
      • Total dimensions 360 x 370 x 600 mm
    • Sphinx phantom consists of wedges for range consistency checks for 4 energies (no direct range measurement) and absolute dose measurements with PPC05 Plane Parallel Chamber.
    • Used with myQA machine software: myQA Machines Software





  • myQA MatriXX AiR
    • Usage: PSQA
    • Wireless ionisation chamber array used for planar verification of of patient treatment plans.
    • WiFi and battery operation mode available.
    • Technical specs:
      • 1521 air-vented ICs
      • 25 × 25 cm² area
      • 6.5 mm center-to-center resolution
      • Minimal sampling time of 20 ms (50 Hz frame rate)
    • Reads out 2D dose distribution at different depths using a



  • Giraffe
    • Usage: Monthly QA, Annual QA, Commissioning
  • Zebra
    • Usage: Monthly QA, Annual QA, Commissioning
  • StingRay
    • Usage: Monthly QA, Annual QA, Commissioning

PTW

Sun Nuclear

DET.TEC.TOR

Logos Systems

MedScint