Proton Calorimetry/Meetings/2021/11/17: Difference between revisions
SaadShaikh (talk | contribs) (Created page with "== Minutes for UCL Proton Calorimetry Meetings, 17th November == === Present === '''Simon Jolly''', '''Saad Shaikh''', '''Raffaella Radogna''' === [[ELogs/SaadShaikh|Saad Shai...") |
SaadShaikh (talk | contribs) No edit summary |
||
(2 intermediate revisions by the same user not shown) | |||
Line 6: | Line 6: | ||
=== [[ELogs/SaadShaikh|Saad Shaikh]] === | === [[ELogs/SaadShaikh|Saad Shaikh]] === | ||
* | *Further analysis of November test beam data, discussed in [http://www.hep.ucl.ac.uk/pbt/wikiData/presentations/2021/SS211110_UCLH.pptx this presentation]. | ||
**Compared November photodiode data to April CMOS data: | |||
***Observed consistent discrepancy in build-up to Bragg peak causing poor fits with photodiodes. | |||
***WET seems to be consistent across data sets. | |||
**Fitting data around peak does not improve things. | |||
**Reverse engineered fit using fixed "correct" values of range and sigma, and varying only Birks' and phi. | |||
***Observe large differences in Birks' constant across data sets, photodiode fits remain worse. | |||
*Will perform fits with CMOS data, taking only central stripe. | |||
**Simon to provide WET calibration value for fitting Nikon data from November. | |||
*Discussed potential causes of poor fits with Laurent: | |||
**Could be that scintillator sheets are damaged after stripping and repainting. | |||
**Overarching problem of detector is non-uniform light output, dependant on beam position and light sensor position/size. Photodiodes could be missing dose. | |||
**Doing a test with mylar foil wrapping could be useful – especially with new sheets from Nuvia and Rev. C DDC232 boards. | |||
*FPGA programming: | |||
**Nexys Video design ready for testing with Rev. C hardware. | |||
**Ported PMOD design to USB104, will test and debug. | |||
**Will debug Matt's fix for signals crossing clock domains. | |||
*First draft of FRPT presentation, small changes to videos/images required. |
Latest revision as of 17:40, 9 December 2021
Minutes for UCL Proton Calorimetry Meetings, 17th November
Present
Simon Jolly, Saad Shaikh, Raffaella Radogna
Saad Shaikh
- Further analysis of November test beam data, discussed in this presentation.
- Compared November photodiode data to April CMOS data:
- Observed consistent discrepancy in build-up to Bragg peak causing poor fits with photodiodes.
- WET seems to be consistent across data sets.
- Fitting data around peak does not improve things.
- Reverse engineered fit using fixed "correct" values of range and sigma, and varying only Birks' and phi.
- Observe large differences in Birks' constant across data sets, photodiode fits remain worse.
- Compared November photodiode data to April CMOS data:
- Will perform fits with CMOS data, taking only central stripe.
- Simon to provide WET calibration value for fitting Nikon data from November.
- Discussed potential causes of poor fits with Laurent:
- Could be that scintillator sheets are damaged after stripping and repainting.
- Overarching problem of detector is non-uniform light output, dependant on beam position and light sensor position/size. Photodiodes could be missing dose.
- Doing a test with mylar foil wrapping could be useful – especially with new sheets from Nuvia and Rev. C DDC232 boards.
- FPGA programming:
- Nexys Video design ready for testing with Rev. C hardware.
- Ported PMOD design to USB104, will test and debug.
- Will debug Matt's fix for signals crossing clock domains.
- First draft of FRPT presentation, small changes to videos/images required.