list of runs in this file unix/pbt/data/ucl/20201127/20201127_LEDandADCBoard.xlsx | list of runs in this file unix/pbt/data/ucl/20201127/20201127_LEDandADCBoard.xlsx | ||
data saved in unix/pbt/data/ucl/ | data saved in unix/pbt/data/ucl/20201127/ | ||
{|class="wikitable" | {|class="wikitable" | ||
Line 37: | Line 37: | ||
|- | |- | ||
|9|| 170us || 350pC FSR || PDs: S12915-16R|| Raffy's office D106: DDC232 board with PD placed OUTSIDE the box. UV torch. | |9|| 170us || 350pC FSR || PDs: S12915-16R|| Raffy's office D106: DDC232 board with PD placed OUTSIDE the box. UV torch. | ||
|- | |||
|10|| 170us || 350pC FSR || PDs: S12915-16R|| D109, Dark Measurement, PELI Case (15/12/20). | |||
|- | |||
|11|| 170us || 350pC FSR || PDs: S12915-16R|| D27, Dark Measurement, PELI Case (15/12/20). | |||
|- | |- |
Latest revision as of 14:35, 20 April 2021
27 Nov 2020
Dark noise measurements in different rooms at UCL and with different conditions
Setup: - PD array exposed to dark - some runs have noise level recorded with PD unplugged
Aim: Measure the dark noise with different conditions. The spikes visible for all runs recorded inside D27 might come from noise
list of runs in this file unix/pbt/data/ucl/20201127/20201127_LEDandADCBoard.xlsx data saved in unix/pbt/data/ucl/20201127/
Run | DDC int time | DDC Full scale | Photodiodes | Notes |
---|---|---|---|---|
1 | 170us | 350pC FSR | PDs: S1337-16BR | Lab D27: dark room. DDC232 board with PD placed INSIDE the box, close to the patch panel holes (probably getting some light). |
2 | 170us | 350pC FSR | PDs: S12915-16R | Lab D27: dark room. DDC232 board with PD placed face down INSIDE the box. |
3 | 170us | 350pC FSR | PDs: S12915-16R | Saad's office D109: dark room. DDC232 board with PD placed face down INSIDE the box. |
4 | 170us | 350pC FSR | PDs: S12915-16R | Saad's office D109: dark room. DDC232 board with PD placed and covered OUTSIDE the box. |
5 | 170us | 350pC FSR | PDs: S12915-16R | Saad's office D109: dark room. DDC232 board with PD UNPLUGGED placed OUTSIDE the box. |
6 | 170us | 350pC FSR | PDs: S12915-16R | Raffy's office D106: dark room (light off, but blinds open). DDC232 board with PD UNPLUGGED placed OUTSIDE the box. |
7 | 170us | 350pC FSR | PDs: S12915-16R | Raffy's office D106: dark room (blinds closed). DDC232 board with PD UNPLUGGED placed OUTSIDE the box. |
8 | 170us | 350pC FSR | PDs: S12915-16R | Raffy's office D106: dark room (blinds closed). DDC232 board with PD placed and covered OUTSIDE the box. |
9 | 170us | 350pC FSR | PDs: S12915-16R | Raffy's office D106: DDC232 board with PD placed OUTSIDE the box. UV torch. |
10 | 170us | 350pC FSR | PDs: S12915-16R | D109, Dark Measurement, PELI Case (15/12/20). |
11 | 170us | 350pC FSR | PDs: S12915-16R | D27, Dark Measurement, PELI Case (15/12/20). |