Proton Calorimetry/Experimental Runs/2020/Oct7: Difference between revisions
(Created page with "07 Oct 2020 == LED injected light measurements == Setup: - scintillator sheets number from bottom to top in sensor image: 51, 31, 50, 52, 34, 44, 53, 47, 33, 54, 46, 42, 55, ...") |
No edit summary |
||
(4 intermediate revisions by 2 users not shown) | |||
Line 13: | Line 13: | ||
Plan is to insert light into sheet n. 8 and measure the light output with the CMOS sensor and different photodiodes arrays. | Plan is to insert light into sheet n. 8 and measure the light output with the CMOS sensor and different photodiodes arrays. | ||
list of runs in this file unix/pbt/data/ucl/ | list of runs in this file unix/pbt/data/ucl/20201007/20201007_LEDandADCBoard.xlsx | ||
data saved in unix/pbt/data/ucl/20201007/ | data saved in unix/pbt/data/ucl/20201007/ | ||
Line 28: | Line 28: | ||
!Notes | !Notes | ||
|- | |- | ||
|1|| 4.5 || 1 || 300 || high || 8 || 170us || 350pC FSR || | |1|| 4.5 || 1 || 300 || high || 8 || 170us || 350pC FSR || Saad's (PDs: S1337-16BR)|| PD placed opposite to the LED fibre | ||
|- | |- | ||
|2|| 4.5 || 1 || 300 || high || 8 || 170us || 350pC FSR || | |2|| 4.5 || 1 || 300 || high || 8 || 170us || 350pC FSR || Saad's (PDs: S1337-16BR)|| Sensor removed and PD placed 90deg to the fibre, in the spot of highes intensity seen by the sensor | ||
|- | |- | ||
|3|| 4.5 || 1 || 300 || high || 8 || | |3|| 4.5 || 1 || 300 || high || 8 || 2000us || 350pC FSR || Saad's (PDs: S1337-16BR) | ||
|- | |- | ||
|4|| 4.5 || 1 || 300 || high || 8 || | |4|| 4.5 || 1 || 300 || high || 8 || 2000us || 12.5pC FSR || Saad's (PDs: S1337-16BR) | ||
|- | |- | ||
| | |5|| 4.5 || 1 || 300 || high || 8 || 20000us || 100pC FSR || Saad's (PDs: S1337-16BR) | ||
|- | |- | ||
| | |6|| 4.5 || 1 || 300 || high || 8 || 170us || 12.5pC FSR || Saad's (PDs: S1337-16BR) | ||
|- | |- | ||
| | |7|| 4.5 || 1 || 300 || high || 8 || 170us || 12.5pC FSR || Raffy's (PDs: S12915-16R) | ||
|- | |- | ||
| | |8|| 4.5 || 1 || 300 || high || 8 || 170us || 12.5pC FSR || Raffy's (PDs: S12915-16R)|| large statistics for noise analysis | ||
|- | |- | ||
| | |9|| 4.5 || 1 || 300 || high || 8 || 170us || 350pC FSR || Saad's (PDs: S1337-16BR) || large statistics for noise analysis | ||
|- | |- | ||
|11|| 4.5 || 1 || 300 || high || 8 || | |10|| 4.5 || 1 || 300 || high || 8 || 2000us || 12.5pC FSR || Raffy's (PDs: S12915-16R)|| large statistics for noise analysis | ||
|- | |||
|11|| 4.5 || 1 || 300 || high || 8 || 2000us || 100pC FSR || Raffy's (PDs: S12915-16R)|| large statistics for noise analysis | |||
|- | |- |
Latest revision as of 08:39, 26 October 2020
07 Oct 2020
LED injected light measurements
Setup: - scintillator sheets number from bottom to top in sensor image: 51, 31, 50, 52, 34, 44, 53, 47, 33, 54, 46, 42, 55, 45, 38, 36, 43, 32, 48, 41, 8, 39, 40, 37, 35, 49, 9 + other sheet as fillers - CMOS ISDI Image sensor. Sensor was allowed to heat up for an hour before measurements - 16 Photodiodes covering the sheets sequence 42, 55, 45, 38, 36, 43, 32, 48, 41, 8, 39, 40, 37, 35, 49, 9. - Optical fibre used with optical gel to couple LED to scintillator sheet n. 8
Aim: Compare light output from sensor and photodiodes. Plan is to insert light into sheet n. 8 and measure the light output with the CMOS sensor and different photodiodes arrays.
list of runs in this file unix/pbt/data/ucl/20201007/20201007_LEDandADCBoard.xlsx data saved in unix/pbt/data/ucl/20201007/
Run | LED Voltage (V) | LED Period (us) | Width (ns) | Sensor Full Well Mode | Illuminated sheet | DDC int time | DDC Full scale | DDC Board and photodiodes | Notes |
---|---|---|---|---|---|---|---|---|---|
1 | 4.5 | 1 | 300 | high | 8 | 170us | 350pC FSR | Saad's (PDs: S1337-16BR) | PD placed opposite to the LED fibre |
2 | 4.5 | 1 | 300 | high | 8 | 170us | 350pC FSR | Saad's (PDs: S1337-16BR) | Sensor removed and PD placed 90deg to the fibre, in the spot of highes intensity seen by the sensor |
3 | 4.5 | 1 | 300 | high | 8 | 2000us | 350pC FSR | Saad's (PDs: S1337-16BR) | |
4 | 4.5 | 1 | 300 | high | 8 | 2000us | 12.5pC FSR | Saad's (PDs: S1337-16BR) | |
5 | 4.5 | 1 | 300 | high | 8 | 20000us | 100pC FSR | Saad's (PDs: S1337-16BR) | |
6 | 4.5 | 1 | 300 | high | 8 | 170us | 12.5pC FSR | Saad's (PDs: S1337-16BR) | |
7 | 4.5 | 1 | 300 | high | 8 | 170us | 12.5pC FSR | Raffy's (PDs: S12915-16R) | |
8 | 4.5 | 1 | 300 | high | 8 | 170us | 12.5pC FSR | Raffy's (PDs: S12915-16R) | large statistics for noise analysis |
9 | 4.5 | 1 | 300 | high | 8 | 170us | 350pC FSR | Saad's (PDs: S1337-16BR) | large statistics for noise analysis |
10 | 4.5 | 1 | 300 | high | 8 | 2000us | 12.5pC FSR | Raffy's (PDs: S12915-16R) | large statistics for noise analysis |
11 | 4.5 | 1 | 300 | high | 8 | 2000us | 100pC FSR | Raffy's (PDs: S12915-16R) | large statistics for noise analysis |